{"title":"三维神经记录阵列","authors":"A. Hoogerwerf, K. Wise","doi":"10.1109/SENSOR.1991.148816","DOIUrl":null,"url":null,"abstract":"The authors describe a three-dimensional recording array for the high-density monitoring of neural activity throughout a volume of cortical tissue. The microassembly techniques used permit multiple multishank planar probes to be precisely configured to form 3D microstructures with probe spacings of 100 mu m or less. The probes are aligned through an orthogonal platform, also formed by silicon micromachining. High-density lead transfers between the probes and the platform are formed by selective electroplating and offer very low values of series resistance and shunt capacitance, with center-to-center lead spacings of 100 mu m or less. A 4*4 shank passive probe array has been found to penetrate pia arachnoid easily and is well accepted by the neural tissue. These assembly techniques also appear to be compatible with a variety of other microelectromechanical systems where 3D structures are required.<<ETX>>","PeriodicalId":273871,"journal":{"name":"TRANSDUCERS '91: 1991 International Conference on Solid-State Sensors and Actuators. Digest of Technical Papers","volume":"298 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1991-06-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"23","resultStr":"{\"title\":\"A three-dimensional neural recording array\",\"authors\":\"A. Hoogerwerf, K. Wise\",\"doi\":\"10.1109/SENSOR.1991.148816\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The authors describe a three-dimensional recording array for the high-density monitoring of neural activity throughout a volume of cortical tissue. The microassembly techniques used permit multiple multishank planar probes to be precisely configured to form 3D microstructures with probe spacings of 100 mu m or less. The probes are aligned through an orthogonal platform, also formed by silicon micromachining. High-density lead transfers between the probes and the platform are formed by selective electroplating and offer very low values of series resistance and shunt capacitance, with center-to-center lead spacings of 100 mu m or less. A 4*4 shank passive probe array has been found to penetrate pia arachnoid easily and is well accepted by the neural tissue. These assembly techniques also appear to be compatible with a variety of other microelectromechanical systems where 3D structures are required.<<ETX>>\",\"PeriodicalId\":273871,\"journal\":{\"name\":\"TRANSDUCERS '91: 1991 International Conference on Solid-State Sensors and Actuators. Digest of Technical Papers\",\"volume\":\"298 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1991-06-24\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"23\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"TRANSDUCERS '91: 1991 International Conference on Solid-State Sensors and Actuators. Digest of Technical Papers\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/SENSOR.1991.148816\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"TRANSDUCERS '91: 1991 International Conference on Solid-State Sensors and Actuators. Digest of Technical Papers","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/SENSOR.1991.148816","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 23

摘要

作者描述了一种三维记录阵列,用于高密度监测整个皮质组织的神经活动。所使用的微组装技术允许多个多柄平面探针精确配置,形成探针间距为100 μ m或更小的3D微结构。探头通过一个正交平台对齐,也是由硅微加工形成的。探头和平台之间的高密度引线传输是通过选择性电镀形成的,并提供非常低的串联电阻和并联电容值,中心到中心的引线间距为100 μ m或更小。一种4*4柄无源探针阵列可以很容易地穿透蛛网膜,并且被神经组织所接受。这些装配技术似乎也与需要3D结构的各种其他微机电系统兼容。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A three-dimensional neural recording array
The authors describe a three-dimensional recording array for the high-density monitoring of neural activity throughout a volume of cortical tissue. The microassembly techniques used permit multiple multishank planar probes to be precisely configured to form 3D microstructures with probe spacings of 100 mu m or less. The probes are aligned through an orthogonal platform, also formed by silicon micromachining. High-density lead transfers between the probes and the platform are formed by selective electroplating and offer very low values of series resistance and shunt capacitance, with center-to-center lead spacings of 100 mu m or less. A 4*4 shank passive probe array has been found to penetrate pia arachnoid easily and is well accepted by the neural tissue. These assembly techniques also appear to be compatible with a variety of other microelectromechanical systems where 3D structures are required.<>
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信