{"title":"电手术过程中生物组织的功率损耗密度分布分析","authors":"C. Knopf, J. Himmel, F. Hochgeschurz, S. Klockner, K. Thelen, O. Kanoun","doi":"10.1109/SSD.2014.6808854","DOIUrl":null,"url":null,"abstract":"Bipolar electrosurgical systems are used for the treatment of benign prostatic hyperplasia (BPH) in urology. In order to analyse electrothermal processes during surgery the power loss density distribution around a bipolar resectoscope is calculated out of the measured potential distribution in isotonic saline solution ex situ. During further analysis power loss density values act as input for the Penne's bioheat equation. To achieve results, which are as realistic as possible, a method to obtain power loss density values, depending on the observed tissue or medium in the operating field, is presented. Applying this method, the power loss density distribution in isotonic saline solution at 25 °C is compared to the distribution calculated for the average conductivity of biological tissue in the region of interest.","PeriodicalId":168063,"journal":{"name":"2014 IEEE 11th International Multi-Conference on Systems, Signals & Devices (SSD14)","volume":"6 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2014-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":"{\"title\":\"Power loss density distribution in biological tissue to analyse processes in electrosurgery\",\"authors\":\"C. Knopf, J. Himmel, F. Hochgeschurz, S. Klockner, K. Thelen, O. Kanoun\",\"doi\":\"10.1109/SSD.2014.6808854\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Bipolar electrosurgical systems are used for the treatment of benign prostatic hyperplasia (BPH) in urology. In order to analyse electrothermal processes during surgery the power loss density distribution around a bipolar resectoscope is calculated out of the measured potential distribution in isotonic saline solution ex situ. During further analysis power loss density values act as input for the Penne's bioheat equation. To achieve results, which are as realistic as possible, a method to obtain power loss density values, depending on the observed tissue or medium in the operating field, is presented. Applying this method, the power loss density distribution in isotonic saline solution at 25 °C is compared to the distribution calculated for the average conductivity of biological tissue in the region of interest.\",\"PeriodicalId\":168063,\"journal\":{\"name\":\"2014 IEEE 11th International Multi-Conference on Systems, Signals & Devices (SSD14)\",\"volume\":\"6 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2014-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"2\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2014 IEEE 11th International Multi-Conference on Systems, Signals & Devices (SSD14)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/SSD.2014.6808854\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2014 IEEE 11th International Multi-Conference on Systems, Signals & Devices (SSD14)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/SSD.2014.6808854","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Power loss density distribution in biological tissue to analyse processes in electrosurgery
Bipolar electrosurgical systems are used for the treatment of benign prostatic hyperplasia (BPH) in urology. In order to analyse electrothermal processes during surgery the power loss density distribution around a bipolar resectoscope is calculated out of the measured potential distribution in isotonic saline solution ex situ. During further analysis power loss density values act as input for the Penne's bioheat equation. To achieve results, which are as realistic as possible, a method to obtain power loss density values, depending on the observed tissue or medium in the operating field, is presented. Applying this method, the power loss density distribution in isotonic saline solution at 25 °C is compared to the distribution calculated for the average conductivity of biological tissue in the region of interest.