P. Busca, C. Fiorini, A. D. Butt, M. Occhipinti, R. Quaglia, P. Trigilio, K. Nagy, T. Bukki, M. Czeller, Z. Nyitrai, C. Piemonte, F. Acerbi, A. Ferri, J. Rieger
{"title":"开发一种基于sipm的Anger相机,用于INSERT,这是一种用于临床前和临床成像的新型多模态SPECT/MRI系统","authors":"P. Busca, C. Fiorini, A. D. Butt, M. Occhipinti, R. Quaglia, P. Trigilio, K. Nagy, T. Bukki, M. Czeller, Z. Nyitrai, C. Piemonte, F. Acerbi, A. Ferri, J. Rieger","doi":"10.1109/NSSMIC.2014.7430774","DOIUrl":null,"url":null,"abstract":"A new multi-modality imaging tool is under development in the framework of the INSERT (INtegrated SPECT/MRI for Enhanced Stratification in Radio-chemoTherapy) project, supported by the European Community. A custom SPECT apparatus, used as an insert for commercially available MRI systems, could enhance the treatment of brain tumors (primarily glioma) by offering more effective and earlier diagnosis with potentially better outcome in survival. In this work we briefly describe the architecture of the gamma-ray photo-detection module with focus on the Silicon PhotoMultiplier (SiPM) detectors and their performance with low-energy gamma-rays (57Co - 122 keV). The camera has a 50.40 mm × 51.70 mm photo-detection area, comprising of 144 SiPMs with 36 readout channels, each one composed of four short-circuited SiPMs for an 8 mm × 8 mm active area. We have developed an Anger camera to characterize and qualify the array when coupled to a CsI:Tl scintillator (8 mm thick). We have measured an intrinsic spatial resolution better than 1.35 mm and an energy resolution below 14% using a 57Co source, when the tile is cooled down to -17°C to reduce the impact of the dark count rate on the measurements.","PeriodicalId":144711,"journal":{"name":"2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","volume":"34 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2014-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"8","resultStr":"{\"title\":\"Development of a SiPM-based Anger camera for INSERT, a new multi-modality SPECT/MRI system for preclinical and clinical imaging\",\"authors\":\"P. Busca, C. Fiorini, A. D. Butt, M. Occhipinti, R. Quaglia, P. Trigilio, K. Nagy, T. Bukki, M. Czeller, Z. Nyitrai, C. Piemonte, F. Acerbi, A. Ferri, J. Rieger\",\"doi\":\"10.1109/NSSMIC.2014.7430774\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"A new multi-modality imaging tool is under development in the framework of the INSERT (INtegrated SPECT/MRI for Enhanced Stratification in Radio-chemoTherapy) project, supported by the European Community. A custom SPECT apparatus, used as an insert for commercially available MRI systems, could enhance the treatment of brain tumors (primarily glioma) by offering more effective and earlier diagnosis with potentially better outcome in survival. In this work we briefly describe the architecture of the gamma-ray photo-detection module with focus on the Silicon PhotoMultiplier (SiPM) detectors and their performance with low-energy gamma-rays (57Co - 122 keV). The camera has a 50.40 mm × 51.70 mm photo-detection area, comprising of 144 SiPMs with 36 readout channels, each one composed of four short-circuited SiPMs for an 8 mm × 8 mm active area. We have developed an Anger camera to characterize and qualify the array when coupled to a CsI:Tl scintillator (8 mm thick). We have measured an intrinsic spatial resolution better than 1.35 mm and an energy resolution below 14% using a 57Co source, when the tile is cooled down to -17°C to reduce the impact of the dark count rate on the measurements.\",\"PeriodicalId\":144711,\"journal\":{\"name\":\"2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)\",\"volume\":\"34 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2014-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"8\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/NSSMIC.2014.7430774\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/NSSMIC.2014.7430774","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 8
摘要
在欧洲共同体的支持下,一种新的多模态成像工具正在INSERT(集成SPECT/MRI用于放化疗强化分层)项目的框架下开发。一种定制的SPECT设备,作为商业上可用的MRI系统的插入物,可以通过提供更有效和更早的诊断和潜在的更好的生存结果来增强脑肿瘤(主要是胶质瘤)的治疗。本文简要介绍了伽玛射线光探测模块的结构,重点介绍了硅光电倍增管(SiPM)探测器及其在低能伽玛射线(57Co - 122 keV)下的性能。相机具有50.40 mm × 51.70 mm的光探测区域,由144个sipm和36个读出通道组成,每个通道由4个短路sipm组成,用于8 mm × 8 mm的有源区域。我们开发了一种Anger相机,当与CsI:Tl闪烁体(8毫米厚)耦合时,可以表征和鉴定该阵列。当瓷砖冷却至-17°C以减少暗计数率对测量的影响时,我们使用57Co源测量了优于1.35 mm的固有空间分辨率和低于14%的能量分辨率。
Development of a SiPM-based Anger camera for INSERT, a new multi-modality SPECT/MRI system for preclinical and clinical imaging
A new multi-modality imaging tool is under development in the framework of the INSERT (INtegrated SPECT/MRI for Enhanced Stratification in Radio-chemoTherapy) project, supported by the European Community. A custom SPECT apparatus, used as an insert for commercially available MRI systems, could enhance the treatment of brain tumors (primarily glioma) by offering more effective and earlier diagnosis with potentially better outcome in survival. In this work we briefly describe the architecture of the gamma-ray photo-detection module with focus on the Silicon PhotoMultiplier (SiPM) detectors and their performance with low-energy gamma-rays (57Co - 122 keV). The camera has a 50.40 mm × 51.70 mm photo-detection area, comprising of 144 SiPMs with 36 readout channels, each one composed of four short-circuited SiPMs for an 8 mm × 8 mm active area. We have developed an Anger camera to characterize and qualify the array when coupled to a CsI:Tl scintillator (8 mm thick). We have measured an intrinsic spatial resolution better than 1.35 mm and an energy resolution below 14% using a 57Co source, when the tile is cooled down to -17°C to reduce the impact of the dark count rate on the measurements.