一种新型便携式心电监护仪的设计

Jun Liu, Yaqi Zhou
{"title":"一种新型便携式心电监护仪的设计","authors":"Jun Liu, Yaqi Zhou","doi":"10.1109/ISCID.2013.178","DOIUrl":null,"url":null,"abstract":"Using the ECG analog front-end and ARM Cortex-M3 processor to develop a portable ECG monitor. The STM32 as the core unit, the ADS1292 as the acquisition analog front-end, it also includes a touch screen display module, an SD card storage module and a voltage conversion module. Automatic ECG analysis algorithms including QRS complex detection, QRS width detection and ST segment detection. ECG can be divided into four kinds of heart beat and eight kinds of arrhythmia rhythm using the extracted ECG parameters. The results have been evaluated on the MIT-BIH Arrhythmia Database, the sensitivity of QRS complex detection was 99% and the sensitivity of heart beat classification was above 95%. The monitor can display the real-time ECG waveform and the current heart rate, to make recommendations for the subjects, and it stored the abnormal ECG waveform that provided to physicians for further analysis and diagnosis. Ultimately the monitor gives a composite score based on heart rate, arrhythmia and ST segment to facilitate subjects for heart health.","PeriodicalId":297027,"journal":{"name":"2013 Sixth International Symposium on Computational Intelligence and Design","volume":"75 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2013-10-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"11","resultStr":"{\"title\":\"Design of a Novel Portable ECG Monitor for Heart Health\",\"authors\":\"Jun Liu, Yaqi Zhou\",\"doi\":\"10.1109/ISCID.2013.178\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Using the ECG analog front-end and ARM Cortex-M3 processor to develop a portable ECG monitor. The STM32 as the core unit, the ADS1292 as the acquisition analog front-end, it also includes a touch screen display module, an SD card storage module and a voltage conversion module. Automatic ECG analysis algorithms including QRS complex detection, QRS width detection and ST segment detection. ECG can be divided into four kinds of heart beat and eight kinds of arrhythmia rhythm using the extracted ECG parameters. The results have been evaluated on the MIT-BIH Arrhythmia Database, the sensitivity of QRS complex detection was 99% and the sensitivity of heart beat classification was above 95%. The monitor can display the real-time ECG waveform and the current heart rate, to make recommendations for the subjects, and it stored the abnormal ECG waveform that provided to physicians for further analysis and diagnosis. Ultimately the monitor gives a composite score based on heart rate, arrhythmia and ST segment to facilitate subjects for heart health.\",\"PeriodicalId\":297027,\"journal\":{\"name\":\"2013 Sixth International Symposium on Computational Intelligence and Design\",\"volume\":\"75 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2013-10-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"11\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2013 Sixth International Symposium on Computational Intelligence and Design\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/ISCID.2013.178\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2013 Sixth International Symposium on Computational Intelligence and Design","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ISCID.2013.178","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 11

摘要

采用心电模拟前端和ARM Cortex-M3处理器开发了一款便携式心电监护仪。该系统以STM32为核心单元,ADS1292为采集模拟前端,还包括触摸屏显示模块、SD卡存储模块和电压转换模块。自动心电分析算法包括QRS复合体检测、QRS宽度检测和ST段检测。利用提取的心电参数可将心电分为4种心跳和8种心律失常节律。结果在MIT-BIH心律失常数据库中进行评价,QRS复合体检测灵敏度为99%,心跳分类灵敏度在95%以上。监护仪可以实时显示心电波形和当前心率,为受试者提供建议,并存储异常心电波形,提供给医生进一步分析诊断。最终,监测器根据心率、心律失常和ST段给出一个综合评分,以促进受试者的心脏健康。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Design of a Novel Portable ECG Monitor for Heart Health
Using the ECG analog front-end and ARM Cortex-M3 processor to develop a portable ECG monitor. The STM32 as the core unit, the ADS1292 as the acquisition analog front-end, it also includes a touch screen display module, an SD card storage module and a voltage conversion module. Automatic ECG analysis algorithms including QRS complex detection, QRS width detection and ST segment detection. ECG can be divided into four kinds of heart beat and eight kinds of arrhythmia rhythm using the extracted ECG parameters. The results have been evaluated on the MIT-BIH Arrhythmia Database, the sensitivity of QRS complex detection was 99% and the sensitivity of heart beat classification was above 95%. The monitor can display the real-time ECG waveform and the current heart rate, to make recommendations for the subjects, and it stored the abnormal ECG waveform that provided to physicians for further analysis and diagnosis. Ultimately the monitor gives a composite score based on heart rate, arrhythmia and ST segment to facilitate subjects for heart health.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信