{"title":"Gambro Lundia Major 1.36 m2一次性平行平板透析器的评价。","authors":"L J Dziukas,&nbsp;S M Douglas,&nbsp;P M Carter,&nbsp;D C Campbell,&nbsp;J T Andrews,&nbsp;A J d'Apice,&nbsp;P Kincaid-Smith","doi":"10.3109/08860227909063948","DOIUrl":null,"url":null,"abstract":"<p><p>The Gambro Lundia Major 1.36 m2 dialyzer was assessed in vivo in ten hemodialysis patients. Urea and creatinine clearances (measured as whole blood values at 60 minutes with a blood flow rate of 200 ml/minute) were 166 +/- 8 m/minute (mean +/- standard error of the mean, n = 9) and 115 +/- 4 ml/minute (n = 11). The creatinine clearance is 15% lower than in vitro data. There was a marked decrease in urea and creatinine clearance with third use of the dialyzer. The ultrafiltration rate was 490 ml/hour/100 mmHg. The priming volume was 125 ml (at a transmembrane pressure of 100 mmHg) and residual blood volume in the dialyzer was 0.57 +/- 0.11 ml (n = 5). The handling, storage and ease of disposal of the dialyzer is better than previous models and its performance characteristics are clinically acceptable.</p>","PeriodicalId":75998,"journal":{"name":"Journal of dialysis","volume":"3 4","pages":"309-25"},"PeriodicalIF":0.0000,"publicationDate":"1979-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.3109/08860227909063948","citationCount":"0","resultStr":"{\"title\":\"Evaluation of the Gambro Lundia Major 1.36 m2 disposable parallel plate dialyzer.\",\"authors\":\"L J Dziukas,&nbsp;S M Douglas,&nbsp;P M Carter,&nbsp;D C Campbell,&nbsp;J T Andrews,&nbsp;A J d'Apice,&nbsp;P Kincaid-Smith\",\"doi\":\"10.3109/08860227909063948\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The Gambro Lundia Major 1.36 m2 dialyzer was assessed in vivo in ten hemodialysis patients. Urea and creatinine clearances (measured as whole blood values at 60 minutes with a blood flow rate of 200 ml/minute) were 166 +/- 8 m/minute (mean +/- standard error of the mean, n = 9) and 115 +/- 4 ml/minute (n = 11). The creatinine clearance is 15% lower than in vitro data. There was a marked decrease in urea and creatinine clearance with third use of the dialyzer. The ultrafiltration rate was 490 ml/hour/100 mmHg. The priming volume was 125 ml (at a transmembrane pressure of 100 mmHg) and residual blood volume in the dialyzer was 0.57 +/- 0.11 ml (n = 5). The handling, storage and ease of disposal of the dialyzer is better than previous models and its performance characteristics are clinically acceptable.</p>\",\"PeriodicalId\":75998,\"journal\":{\"name\":\"Journal of dialysis\",\"volume\":\"3 4\",\"pages\":\"309-25\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1979-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.3109/08860227909063948\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of dialysis\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.3109/08860227909063948\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of dialysis","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3109/08860227909063948","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

Gambro Lundia Major 1.36 m2透析器在10例血液透析患者体内进行了评估。尿素和肌酐清除率(以血流量为200 ml/min时60分钟的全血值测量)为166 +/- 8 m/min(平均+/-标准误差,n = 9)和115 +/- 4 ml/min (n = 11)。肌酐清除率比体外数据低15%。在第三次使用透析器时,尿素和肌酐清除率明显降低。超滤速率为490 ml/h /100 mmHg。启动量为125 ml(跨膜压力为100 mmHg),透析器残血量为0.57±0.11 ml (n = 5)。该透析器的操作、储存和处置便利性优于以往型号,其性能特点临床可接受。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Evaluation of the Gambro Lundia Major 1.36 m2 disposable parallel plate dialyzer.

The Gambro Lundia Major 1.36 m2 dialyzer was assessed in vivo in ten hemodialysis patients. Urea and creatinine clearances (measured as whole blood values at 60 minutes with a blood flow rate of 200 ml/minute) were 166 +/- 8 m/minute (mean +/- standard error of the mean, n = 9) and 115 +/- 4 ml/minute (n = 11). The creatinine clearance is 15% lower than in vitro data. There was a marked decrease in urea and creatinine clearance with third use of the dialyzer. The ultrafiltration rate was 490 ml/hour/100 mmHg. The priming volume was 125 ml (at a transmembrane pressure of 100 mmHg) and residual blood volume in the dialyzer was 0.57 +/- 0.11 ml (n = 5). The handling, storage and ease of disposal of the dialyzer is better than previous models and its performance characteristics are clinically acceptable.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信