{"title":"根据ISO 81060-2:2018/AMD 1:2020协议,在普通人群中验证U60EH腕戴式电子血压计。","authors":"Sicong Mu, Min Zhu, Liping Ma, Lixiu Fang","doi":"10.1097/MBP.0000000000000659","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To evaluate the accuracy of the U60EH Wrist Electronic Blood Pressure Monitor in general population according to the Universal Standard (ISO 81060-2:2018/AMD 1:2020).</p><p><strong>Methods: </strong>Subjects were recruited to fulfill the age, gender, blood pressure (BP) and cuff distribution criteria of the Universal Standard in a general population using the same arm sequential BP measurement method. A single cuff for wrist sizes 13.5-21.5 cm was used on this test device.</p><p><strong>Results: </strong>According to Criterion 1, the mean difference of SBP between the test device and the reference device was 1.51 mmHg, with an SD of 6.48 mmHg. The mean difference of DBP was -0.44 mmHg, with an SD of 5.98 mmHg. The mean difference of both SBP and DBP was less than 5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to Criterion 2, the mean difference of SBP between the test device and the reference device was 1.51 mmHg, and the SD was 5.88 mmHg, which was less than 6.78 mmHg and met the requirements. The mean difference of DBP was -0.44 mmHg, and the SD was 5.22 mmHg, which was less than 6.93 mmHg and met the requirements.</p><p><strong>Conclusion: </strong>All results passed the Standard (ISO 81060-2:2018/AMD 1:2020) requirements. The U60EH Wrist Electronic Blood Pressure Monitor can be recommended for home and clinical use.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":"28 5","pages":"280-283"},"PeriodicalIF":1.2000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/46/13/bpmj-28-280.PMC10484183.pdf","citationCount":"0","resultStr":"{\"title\":\"Validation of the U60EH Wrist Electronic Blood Pressure Monitor in general population according to the ISO 81060-2:2018/AMD 1:2020 Protocol.\",\"authors\":\"Sicong Mu, Min Zhu, Liping Ma, Lixiu Fang\",\"doi\":\"10.1097/MBP.0000000000000659\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To evaluate the accuracy of the U60EH Wrist Electronic Blood Pressure Monitor in general population according to the Universal Standard (ISO 81060-2:2018/AMD 1:2020).</p><p><strong>Methods: </strong>Subjects were recruited to fulfill the age, gender, blood pressure (BP) and cuff distribution criteria of the Universal Standard in a general population using the same arm sequential BP measurement method. A single cuff for wrist sizes 13.5-21.5 cm was used on this test device.</p><p><strong>Results: </strong>According to Criterion 1, the mean difference of SBP between the test device and the reference device was 1.51 mmHg, with an SD of 6.48 mmHg. The mean difference of DBP was -0.44 mmHg, with an SD of 5.98 mmHg. The mean difference of both SBP and DBP was less than 5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to Criterion 2, the mean difference of SBP between the test device and the reference device was 1.51 mmHg, and the SD was 5.88 mmHg, which was less than 6.78 mmHg and met the requirements. The mean difference of DBP was -0.44 mmHg, and the SD was 5.22 mmHg, which was less than 6.93 mmHg and met the requirements.</p><p><strong>Conclusion: </strong>All results passed the Standard (ISO 81060-2:2018/AMD 1:2020) requirements. The U60EH Wrist Electronic Blood Pressure Monitor can be recommended for home and clinical use.</p>\",\"PeriodicalId\":8950,\"journal\":{\"name\":\"Blood Pressure Monitoring\",\"volume\":\"28 5\",\"pages\":\"280-283\"},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2023-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/46/13/bpmj-28-280.PMC10484183.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Blood Pressure Monitoring\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MBP.0000000000000659\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/7/19 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"PERIPHERAL VASCULAR DISEASE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000659","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/7/19 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

摘要

目的:根据通用标准(ISO 81060-2:2018/AMD 1:2020)评估U60EH腕戴式电子血压计在普通人群中的准确性。适用于手腕尺寸13.5-21.5的单袖带 cm用于该测试装置。结果:根据标准1,试验装置和参考装置之间的SBP平均差为1.51 mmHg,SD为6.48 mmHg。DBP的平均差异为-0.44 mmHg,SD为5.98 mmHg。SBP和DBP的平均差异均小于5 mmHg,SD小于8 mmHg,符合要求。根据标准2,试验装置和参考装置之间的SBP平均差为1.51 mmHg,SD为5.88 mmHg,小于6.78 mmHg,符合要求。DBP的平均差异为-0.44 mmHg,SD为5.22 mmHg,小于6.93 mmHg,符合要求。结论:所有结果均符合标准(ISO 81060-2:2018/AMD 1:2020)要求。U60EH腕戴式电子血压计可推荐用于家庭和临床。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Validation of the U60EH Wrist Electronic Blood Pressure Monitor in general population according to the ISO 81060-2:2018/AMD 1:2020 Protocol.

Validation of the U60EH Wrist Electronic Blood Pressure Monitor in general population according to the ISO 81060-2:2018/AMD 1:2020 Protocol.

Validation of the U60EH Wrist Electronic Blood Pressure Monitor in general population according to the ISO 81060-2:2018/AMD 1:2020 Protocol.

Objective: To evaluate the accuracy of the U60EH Wrist Electronic Blood Pressure Monitor in general population according to the Universal Standard (ISO 81060-2:2018/AMD 1:2020).

Methods: Subjects were recruited to fulfill the age, gender, blood pressure (BP) and cuff distribution criteria of the Universal Standard in a general population using the same arm sequential BP measurement method. A single cuff for wrist sizes 13.5-21.5 cm was used on this test device.

Results: According to Criterion 1, the mean difference of SBP between the test device and the reference device was 1.51 mmHg, with an SD of 6.48 mmHg. The mean difference of DBP was -0.44 mmHg, with an SD of 5.98 mmHg. The mean difference of both SBP and DBP was less than 5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to Criterion 2, the mean difference of SBP between the test device and the reference device was 1.51 mmHg, and the SD was 5.88 mmHg, which was less than 6.78 mmHg and met the requirements. The mean difference of DBP was -0.44 mmHg, and the SD was 5.22 mmHg, which was less than 6.93 mmHg and met the requirements.

Conclusion: All results passed the Standard (ISO 81060-2:2018/AMD 1:2020) requirements. The U60EH Wrist Electronic Blood Pressure Monitor can be recommended for home and clinical use.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Blood Pressure Monitoring
Blood Pressure Monitoring 医学-外周血管病
CiteScore
2.00
自引率
7.70%
发文量
110
审稿时长
>12 weeks
期刊介绍: Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research. This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine. Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信