使用快速成像协议合成磁共振成像评估正常大脑的 T1、T2 弛豫和质子密度

IF 4.6 Q2 MATERIALS SCIENCE, BIOMATERIALS
ACS Applied Bio Materials Pub Date : 2024-10-01 Epub Date: 2023-09-08 DOI:10.2463/mrms.tn.2022-0161
Zuofeng Zheng, Yawen Liu, Hongxia Yin, Pengling Ren, Tingting Zhang, Jiafei Yang, Zhenchang Wang
{"title":"使用快速成像协议合成磁共振成像评估正常大脑的 T1、T2 弛豫和质子密度","authors":"Zuofeng Zheng, Yawen Liu, Hongxia Yin, Pengling Ren, Tingting Zhang, Jiafei Yang, Zhenchang Wang","doi":"10.2463/mrms.tn.2022-0161","DOIUrl":null,"url":null,"abstract":"<p><p>Synthetic MRI is being increasingly used for the quantification of brain longitudinal relaxation time (T1), transverse relaxation time (T2), and proton density (PD) values. However, the effect of fast imaging protocols on these quantitative values has not been fully estimated. The purpose of this study was to investigate the effect of fast scan parameters on T1, T2, and PD measured with a multi-dynamic multi-echo (MDME) sequence of normal brain at 3.0T. Thirty-four volunteers were scanned using 3 MDME sequences with different scan times (named Fast, 2 min, 29 sec; Routine, 4 min, 07 sec; and Research, 7 min, 46 sec, respectively). The measured T1, T2, and PD in 18 volumes of interest (VOI) of brain were compared between the 3 sequences using rank sum test, t test, coefficients of variation (CVs) analysis, correlation analysis, and Bland-Altman analysis. We found that even though T1, T2, and PD were significantly different between the 3 sequences in most of the brain regions, the intersequence CVs were relatively low and linear correlation were high. Bland-Altman plots showed that most of the values fall within the 95% prediction limits. We concluded that fast imaging protocols of MDME sequence used in our study can potentially be used for quantitative evaluation of brain tissues. Since changing scan parameters can affect the measured T1, T2, and PD values, it is necessary to use consistent scan parameter for comparing or following up cases quantitatively.</p>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":" ","pages":"514-524"},"PeriodicalIF":4.6000,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11447464/pdf/","citationCount":"0","resultStr":"{\"title\":\"Evaluating T1, T2 Relaxation, and Proton Density in Normal Brain Using Synthetic MRI with Fast Imaging Protocol.\",\"authors\":\"Zuofeng Zheng, Yawen Liu, Hongxia Yin, Pengling Ren, Tingting Zhang, Jiafei Yang, Zhenchang Wang\",\"doi\":\"10.2463/mrms.tn.2022-0161\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Synthetic MRI is being increasingly used for the quantification of brain longitudinal relaxation time (T1), transverse relaxation time (T2), and proton density (PD) values. However, the effect of fast imaging protocols on these quantitative values has not been fully estimated. The purpose of this study was to investigate the effect of fast scan parameters on T1, T2, and PD measured with a multi-dynamic multi-echo (MDME) sequence of normal brain at 3.0T. Thirty-four volunteers were scanned using 3 MDME sequences with different scan times (named Fast, 2 min, 29 sec; Routine, 4 min, 07 sec; and Research, 7 min, 46 sec, respectively). The measured T1, T2, and PD in 18 volumes of interest (VOI) of brain were compared between the 3 sequences using rank sum test, t test, coefficients of variation (CVs) analysis, correlation analysis, and Bland-Altman analysis. We found that even though T1, T2, and PD were significantly different between the 3 sequences in most of the brain regions, the intersequence CVs were relatively low and linear correlation were high. Bland-Altman plots showed that most of the values fall within the 95% prediction limits. We concluded that fast imaging protocols of MDME sequence used in our study can potentially be used for quantitative evaluation of brain tissues. Since changing scan parameters can affect the measured T1, T2, and PD values, it is necessary to use consistent scan parameter for comparing or following up cases quantitatively.</p>\",\"PeriodicalId\":2,\"journal\":{\"name\":\"ACS Applied Bio Materials\",\"volume\":\" \",\"pages\":\"514-524\"},\"PeriodicalIF\":4.6000,\"publicationDate\":\"2024-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11447464/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ACS Applied Bio Materials\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.2463/mrms.tn.2022-0161\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/9/8 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"MATERIALS SCIENCE, BIOMATERIALS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2463/mrms.tn.2022-0161","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/9/8 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
引用次数: 0

摘要

合成 MRI 越来越多地被用于量化大脑纵向弛豫时间(T1)、横向弛豫时间(T2)和质子密度(PD)值。然而,快速成像方案对这些定量值的影响尚未得到充分估计。本研究旨在研究快速扫描参数对 3.0T 下正常大脑多动态多回波(MDME)序列测量的 T1、T2 和 PD 的影响。研究人员使用 3 个扫描时间不同的 MDME 序列(分别为快速,2 分 29 秒;常规,4 分 07 秒;研究,7 分 46 秒)对 34 名志愿者进行了扫描。使用秩和检验、t检验、变异系数(CVs)分析、相关分析和布兰-阿尔特曼(Bland-Altman)分析比较了三种序列在大脑18个感兴趣容积(VOI)中测得的T1、T2和PD。我们发现,尽管 3 个序列的 T1、T2 和 PD 在大多数脑区有显著差异,但序列间变异系数相对较低,线性相关性较高。Bland-Altman图显示,大多数数值都在95%的预测范围内。我们的结论是,研究中使用的 MDME 序列快速成像方案可用于脑组织的定量评估。由于改变扫描参数会影响测量的 T1、T2 和 PD 值,因此有必要使用一致的扫描参数对病例进行定量比较或随访。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Evaluating T1, T2 Relaxation, and Proton Density in Normal Brain Using Synthetic MRI with Fast Imaging Protocol.

Synthetic MRI is being increasingly used for the quantification of brain longitudinal relaxation time (T1), transverse relaxation time (T2), and proton density (PD) values. However, the effect of fast imaging protocols on these quantitative values has not been fully estimated. The purpose of this study was to investigate the effect of fast scan parameters on T1, T2, and PD measured with a multi-dynamic multi-echo (MDME) sequence of normal brain at 3.0T. Thirty-four volunteers were scanned using 3 MDME sequences with different scan times (named Fast, 2 min, 29 sec; Routine, 4 min, 07 sec; and Research, 7 min, 46 sec, respectively). The measured T1, T2, and PD in 18 volumes of interest (VOI) of brain were compared between the 3 sequences using rank sum test, t test, coefficients of variation (CVs) analysis, correlation analysis, and Bland-Altman analysis. We found that even though T1, T2, and PD were significantly different between the 3 sequences in most of the brain regions, the intersequence CVs were relatively low and linear correlation were high. Bland-Altman plots showed that most of the values fall within the 95% prediction limits. We concluded that fast imaging protocols of MDME sequence used in our study can potentially be used for quantitative evaluation of brain tissues. Since changing scan parameters can affect the measured T1, T2, and PD values, it is necessary to use consistent scan parameter for comparing or following up cases quantitatively.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
ACS Applied Bio Materials
ACS Applied Bio Materials Chemistry-Chemistry (all)
CiteScore
9.40
自引率
2.10%
发文量
464
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信