Electrogustometry: validation of bipolar electrode stimulation.

IF 2.8 4区 心理学 Q1 BEHAVIORAL SCIENCES
Toshi Matsuda, Pavana Mysore Ganesh, Robert Brown, Vince Grosso, Richard L Doty
{"title":"Electrogustometry: validation of bipolar electrode stimulation.","authors":"Toshi Matsuda,&nbsp;Pavana Mysore Ganesh,&nbsp;Robert Brown,&nbsp;Vince Grosso,&nbsp;Richard L Doty","doi":"10.1093/chemse/bjad009","DOIUrl":null,"url":null,"abstract":"<p><p>Electrogustometry (EGM) is a practical way to test taste. It is typically performed using unipolar electrodes, with the anode on the tongue and the cathode on the hand, forearm, or neck. This results in electric current passing through nontaste tissues and adds a level of impracticality to its clinical application. We compared, using a repeated measures counterbalanced design, anodal thresholds from a unipolar electrode to those of a unique bipolar electrode in which the anode and cathode are contiguously located. Both sides of the anterior tongue were assessed in 70 subjects, as were the effects of age and sex. Nonparametric analyses were performed. The median threshold of the bipolar electrode's central disk (2.49 µA) did not differ from that of the unipolar electrode (2.96 µA) (P = 0.84). On average, older persons exhibited higher thresholds. No significant sex or tongue side effects were evident. Interestingly, when the annular (donut-shaped) bipolar electrode served as the anode, the threshold was higher than that of the other electrodes (5.19 µA; Ps < 0.001). This conceivably reflected lessened summation of activity among adjacent afferents and partial sampling of tongue regions with fewer taste buds. Correlations among all EGM thresholds were nominally higher for women than for men, ranging from 0.83 to 0.85 for women and 0.54 to 0.67 for men; all Ps < 0.001. This study validates the use of a bipolar electrode for assessing taste function, averting movement of current through nontaste-related tissues and making such testing safer and more practical.</p>","PeriodicalId":9771,"journal":{"name":"Chemical Senses","volume":"48 ","pages":""},"PeriodicalIF":2.8000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Chemical Senses","FirstCategoryId":"102","ListUrlMain":"https://doi.org/10.1093/chemse/bjad009","RegionNum":4,"RegionCategory":"心理学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"BEHAVIORAL SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

Electrogustometry (EGM) is a practical way to test taste. It is typically performed using unipolar electrodes, with the anode on the tongue and the cathode on the hand, forearm, or neck. This results in electric current passing through nontaste tissues and adds a level of impracticality to its clinical application. We compared, using a repeated measures counterbalanced design, anodal thresholds from a unipolar electrode to those of a unique bipolar electrode in which the anode and cathode are contiguously located. Both sides of the anterior tongue were assessed in 70 subjects, as were the effects of age and sex. Nonparametric analyses were performed. The median threshold of the bipolar electrode's central disk (2.49 µA) did not differ from that of the unipolar electrode (2.96 µA) (P = 0.84). On average, older persons exhibited higher thresholds. No significant sex or tongue side effects were evident. Interestingly, when the annular (donut-shaped) bipolar electrode served as the anode, the threshold was higher than that of the other electrodes (5.19 µA; Ps < 0.001). This conceivably reflected lessened summation of activity among adjacent afferents and partial sampling of tongue regions with fewer taste buds. Correlations among all EGM thresholds were nominally higher for women than for men, ranging from 0.83 to 0.85 for women and 0.54 to 0.67 for men; all Ps < 0.001. This study validates the use of a bipolar electrode for assessing taste function, averting movement of current through nontaste-related tissues and making such testing safer and more practical.

电测法:双极电极刺激的验证。
电味觉法(EGM)是一种实用的味觉测试方法。它通常使用单极电极,阳极在舌头上,阴极在手、前臂或脖子上。这导致电流通过非味觉组织,增加了其临床应用的不实用性。我们使用重复测量平衡设计,比较了从单极电极到阳极和阴极相邻的独特双极电极的阳极阈值。对70名受试者的舌头前两侧以及年龄和性别的影响进行了评估。进行非参数分析。双极电极中央圆盘的中位阈值(2.49µA)与单极电极的中位阈值(2.96µA)无显著差异(P = 0.84)。平均而言,老年人的阈值较高。没有明显的性或舌头副作用。有趣的是,当环形(甜甜圈形)双极电极作为阳极时,阈值高于其他电极(5.19µA;p < 0.001)。可以想象,这反映了相邻传入事件之间活动总和的减少,以及味蕾较少的舌头区域的部分采样。所有EGM阈值之间的相关性在名义上高于男性,女性为0.83至0.85,男性为0.54至0.67;p < 0.001。这项研究验证了使用双极电极来评估味觉功能,避免电流通过与味觉无关的组织,使这种测试更安全、更实用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Chemical Senses
Chemical Senses 医学-行为科学
CiteScore
8.60
自引率
2.90%
发文量
25
审稿时长
1 months
期刊介绍: Chemical Senses publishes original research and review papers on all aspects of chemoreception in both humans and animals. An important part of the journal''s coverage is devoted to techniques and the development and application of new methods for investigating chemoreception and chemosensory structures.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信