Optimization of biochip assay for illegal substances on drug abusers' whole blood: Randox Evidence vs LC-MS/MS.

Q2 Health Professions
Pinar Efeoglu Ozseker, Sevinc Puren Yucel, Nebile Daglioglu
{"title":"Optimization of biochip assay for illegal substances on drug abusers' whole blood: Randox Evidence vs LC-MS/MS.","authors":"Pinar Efeoglu Ozseker,&nbsp;Sevinc Puren Yucel,&nbsp;Nebile Daglioglu","doi":"10.1080/15321819.2023.2189451","DOIUrl":null,"url":null,"abstract":"<p><p>Immunoassay tests are used in clinical and forensic toxicology laboratories to determine illicit drug use in biological samples. Therefore, this study aims to optimize the cutoff concentrations of DOA I Plus in the blood and compare the LC-MS/MS results. 680 authentic forensic whole blood specimens with Randox Evidence DOA I Plus array were screened for drug of abuse and confirmed by LC-MS/MS. Regarding the manufacturer recommended threshold values, 139 out of 680 authentic blood samples were positive for one or more analytes, while 541 were negative. Nearly all of the 139 positive blood samples confirmed by LC-MS/MS were true positive for one or more analytes, while 522 of the 541 negative blood samples were true negative. The overall sensitivity and the specificity were 87.8% and 99.6%, respectively. THC was considered in detail, and a receiver operator characteristic curve analysis was performed to determine the optimum cutoff for THC, as it accounts for 78% of all positive results according to the manufacturer's recommended thresholds. The optimal threshold value for THC was determined at a concentration of 23 ng/mL, while these values for other parameters were defined as recommended by the manufacturer.</p>","PeriodicalId":15990,"journal":{"name":"Journal of immunoassay & immunochemistry","volume":"44 4","pages":"313-325"},"PeriodicalIF":0.0000,"publicationDate":"2023-07-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of immunoassay & immunochemistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/15321819.2023.2189451","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/3/15 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 0

Abstract

Immunoassay tests are used in clinical and forensic toxicology laboratories to determine illicit drug use in biological samples. Therefore, this study aims to optimize the cutoff concentrations of DOA I Plus in the blood and compare the LC-MS/MS results. 680 authentic forensic whole blood specimens with Randox Evidence DOA I Plus array were screened for drug of abuse and confirmed by LC-MS/MS. Regarding the manufacturer recommended threshold values, 139 out of 680 authentic blood samples were positive for one or more analytes, while 541 were negative. Nearly all of the 139 positive blood samples confirmed by LC-MS/MS were true positive for one or more analytes, while 522 of the 541 negative blood samples were true negative. The overall sensitivity and the specificity were 87.8% and 99.6%, respectively. THC was considered in detail, and a receiver operator characteristic curve analysis was performed to determine the optimum cutoff for THC, as it accounts for 78% of all positive results according to the manufacturer's recommended thresholds. The optimal threshold value for THC was determined at a concentration of 23 ng/mL, while these values for other parameters were defined as recommended by the manufacturer.

药物滥用者全血中非法物质生物芯片检测的优化:Randox证据与LC-MS/MS。
免疫分析测试用于临床和法医毒理学实验室,以确定生物样本中的非法药物使用。因此,本研究旨在优化血液中DOA I Plus的截止浓度,并比较LC-MS/MS结果。使用Randox Evidence DOA I Plus阵列对680份真实的法医全血样本进行滥用药物筛选,并通过LC-MS/MS进行确认。关于制造商建议的阈值,680份真实血液样本中有139份对一种或多种分析物呈阳性,541份呈阴性。LC-MS/MS确认的139份阳性血液样本中,几乎所有的一种或多种分析物均为真阳性,541份阴性血液样本中有522份为真阴性。总的敏感性和特异性分别为87.8%和99.6%。详细考虑了四氢大麻酚,并进行了受试者-操作员特性曲线分析,以确定四氢大麻醚的最佳截止值,因为根据制造商的建议阈值,它占所有阳性结果的78%。THC的最佳阈值是在浓度为23时确定的 ng/mL,而其他参数的这些值是按照制造商的建议定义的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.50
自引率
0.00%
发文量
38
审稿时长
>12 weeks
期刊介绍: The Journal of Immunoassay & Immunochemistry is an international forum for rapid dissemination of research results and methodologies dealing with all aspects of immunoassay and immunochemistry, as well as selected aspects of immunology. They include receptor assay, enzyme-linked immunosorbent assay (ELISA) in all of its embodiments, ligand-based assays, biological markers of ligand-receptor interaction, in vivo and in vitro diagnostic reagents and techniques, diagnosis of AIDS, point-of-care testing, clinical immunology, antibody isolation and purification, and others.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信