Maria M Aspradakis , Thierry Buchillier , Götz Kohler , Christian Kottler , Jérôme Krayenbühl
{"title":"SSRMP Recommendations No 9: Reference dosimetry in low and medium energy x-ray beams","authors":"Maria M Aspradakis , Thierry Buchillier , Götz Kohler , Christian Kottler , Jérôme Krayenbühl","doi":"10.1016/j.zemedi.2022.12.003","DOIUrl":null,"url":null,"abstract":"<div><p>The SSRMP recommendations on reference dosimetry in kilovolt beams as used in radiation therapy were revised to establish current practice in Switzerland.</p><p>The recommendations specify the dosimetry formalism, reference class dosimeter systems and conditions used for the calibration of low and medium energy x-ray beams. Practical guidance is provided on the determination of the beam quality specifier and all corrections required for converting instrument readings to absorbed dose to water. Guidance is also provided on the determination of relative dose under non-reference conditions and on the cross calibration of instruments.</p><p>The effect of lack of electron equilibrium and influence of contaminant electrons when using thin window plane parallel chambers at x-ray tube potentials higher than 50<!--> <!-->kV is elaborated in an appendix. In Switzerland the calibration of the reference system used for dosimetry is regulated by law. METAS and IRA are the authorities providing this calibration service to the radiotherapy departments. The last appendix of these recommendations summarise this calibration chain.</p></div>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":null,"pages":null},"PeriodicalIF":4.6000,"publicationDate":"2023-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S0939388922001362/pdfft?md5=7010f5938db47cdcfb3586cb6766ddc7&pid=1-s2.0-S0939388922001362-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0939388922001362","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
引用次数: 0
Abstract
The SSRMP recommendations on reference dosimetry in kilovolt beams as used in radiation therapy were revised to establish current practice in Switzerland.
The recommendations specify the dosimetry formalism, reference class dosimeter systems and conditions used for the calibration of low and medium energy x-ray beams. Practical guidance is provided on the determination of the beam quality specifier and all corrections required for converting instrument readings to absorbed dose to water. Guidance is also provided on the determination of relative dose under non-reference conditions and on the cross calibration of instruments.
The effect of lack of electron equilibrium and influence of contaminant electrons when using thin window plane parallel chambers at x-ray tube potentials higher than 50 kV is elaborated in an appendix. In Switzerland the calibration of the reference system used for dosimetry is regulated by law. METAS and IRA are the authorities providing this calibration service to the radiotherapy departments. The last appendix of these recommendations summarise this calibration chain.
SSRMP 对放射治疗中使用的千伏光束参考剂量测定建议进行了修订,以确立瑞士的现行做法。该建议规定了剂量测定形式、参考类剂量计系统以及用于校准低能量和中等能量 X 射线光束的条件。该建议书提供了实用指南,指导如何确定光束质量规格,以及将仪器读数转换为水吸收剂量所需的所有修正。附录中详细阐述了在 X 射线管电位高于 50 千伏的情况下使用薄窗平面平行腔时缺乏电子平衡的影响和杂质电子的影响。在瑞士,用于剂量测定的参考系统的校准受法律约束。METAS 和 IRA 是为放射治疗部门提供校准服务的机构。本建议书的最后一个附录总结了这一校准链。