How does nano-focus computed tomography impact the quantification of debris within the root canal system?

IF 2.5 4区 医学 Q2 Dentistry
Clarissa Teles Rodrigues, Reinhilde Jacobs, Mostafa Ezeldeen, Karla de Faria Vasconcelos, Paul Lambrechts, Mario Tanomaru Filho, Jader Camilo Pinto, Marco Antonio Hungaro Duarte
{"title":"How does nano-focus computed tomography impact the quantification of debris within the root canal system?","authors":"Clarissa Teles Rodrigues,&nbsp;Reinhilde Jacobs,&nbsp;Mostafa Ezeldeen,&nbsp;Karla de Faria Vasconcelos,&nbsp;Paul Lambrechts,&nbsp;Mario Tanomaru Filho,&nbsp;Jader Camilo Pinto,&nbsp;Marco Antonio Hungaro Duarte","doi":"10.1590/1807-3107bor-2023.vol37.0059","DOIUrl":null,"url":null,"abstract":"<p><p>The aim of this study was to compare the quantification of hard-tissue debris by using micro-computed tomography (micro-CT) and nano-focus computed tomography (nano-CT) after root canal instrumentation. Ten mandibular molars containing an isthmus in the mesial root were scanned in a SkyScan 1172 micro-CT device with a voxel size of 12.8 µm and in a NanoTom nano-CT device with 5.5 µm. The mesial root canals were irrigated with 5 mL of saline solution at the orifice level, instrumented with Reciproc R25 files and a second scanning was performed by micro-CT and nano-CT devices for post-instrumentation images. DataViewer software was used for registering the pre- and post-operative micro-CT and nano-CT images. The root canal and the debris were segmented for quantitative analysis of the volume of the canal and volume of debris using CTAn software. Statistical analysis was performed using the T test for comparison between volume of the canal after instrumentation and volume of debris in both image modalities. The level of significance was set at p < 0.05. Nano-CT images showed higher values of debris when compared with micro-CT (p < 0.05) after root canal instrumentation. No difference was observed between the volume of the root canal after instrumentation in the two imaging methods used (p > 0.05). Nano-CT technology can be recommended as a more precise method for quantitative analysis of hard-tissue debris. Moreover, in Endodontic research it is a promising method, as it is capable of providing higher spatial and contrast resolution, faster scanning and higher image quality.</p>","PeriodicalId":48942,"journal":{"name":"Brazilian Oral Research","volume":null,"pages":null},"PeriodicalIF":2.5000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Brazilian Oral Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1590/1807-3107bor-2023.vol37.0059","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"Dentistry","Score":null,"Total":0}
引用次数: 0

Abstract

The aim of this study was to compare the quantification of hard-tissue debris by using micro-computed tomography (micro-CT) and nano-focus computed tomography (nano-CT) after root canal instrumentation. Ten mandibular molars containing an isthmus in the mesial root were scanned in a SkyScan 1172 micro-CT device with a voxel size of 12.8 µm and in a NanoTom nano-CT device with 5.5 µm. The mesial root canals were irrigated with 5 mL of saline solution at the orifice level, instrumented with Reciproc R25 files and a second scanning was performed by micro-CT and nano-CT devices for post-instrumentation images. DataViewer software was used for registering the pre- and post-operative micro-CT and nano-CT images. The root canal and the debris were segmented for quantitative analysis of the volume of the canal and volume of debris using CTAn software. Statistical analysis was performed using the T test for comparison between volume of the canal after instrumentation and volume of debris in both image modalities. The level of significance was set at p < 0.05. Nano-CT images showed higher values of debris when compared with micro-CT (p < 0.05) after root canal instrumentation. No difference was observed between the volume of the root canal after instrumentation in the two imaging methods used (p > 0.05). Nano-CT technology can be recommended as a more precise method for quantitative analysis of hard-tissue debris. Moreover, in Endodontic research it is a promising method, as it is capable of providing higher spatial and contrast resolution, faster scanning and higher image quality.

纳米聚焦计算机断层扫描如何影响根管系统内碎片的量化?
本研究的目的是比较根管预备后使用微计算机断层扫描(micro-CT)和纳米焦点计算机断层扫描(nano-CT)对硬组织碎片的定量。采用体素尺寸为12.8µm的SkyScan 1172微型ct设备和5.5µm的NanoTom纳米ct设备对10颗下颌磨牙进行扫描。在孔位用5ml生理盐水溶液冲洗近中根管,用Reciproc R25锉进行器械检查,并用micro-CT和nano-CT设备进行第二次扫描,以获得器械检查后的图像。应用DataViewer软件对术前、术后微ct和纳米ct图像进行配准。使用CTAn软件对根管和碎片进行分割,定量分析根管体积和碎片体积。使用T检验进行统计分析,比较两种图像模式下预备后的根管体积和碎片体积。p < 0.05为显著性水平。根管预备后,纳米ct图像显示碎片值高于微ct (p < 0.05)。两种成像方法根管内固定后根管体积比较差异无统计学意义(p > 0.05)。纳米ct技术可以作为一种更精确的硬组织碎片定量分析方法。此外,在牙髓研究中,它具有更高的空间分辨率和对比度分辨率,更快的扫描速度和更高的图像质量,是一种很有前途的方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Brazilian Oral Research
Brazilian Oral Research DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
3.70
自引率
4.00%
发文量
107
审稿时长
12 weeks
文献相关原料
公司名称 产品信息 采购帮参考价格
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信