High-Throughput Ultra Performance Liquid Chromatography-Tandem Mass Spectrometry Method Validation for the Estimation of Atorvastatin and Active Metabolites in Human Plasma.
{"title":"High-Throughput Ultra Performance Liquid Chromatography-Tandem Mass Spectrometry Method Validation for the Estimation of Atorvastatin and Active Metabolites in Human Plasma.","authors":"Nikhil Agrawal, Amit Mittal","doi":"10.1089/adt.2022.113","DOIUrl":null,"url":null,"abstract":"<p><p>A highly selective, sensitive, rugged, and rapid ultra high-performance liquid chromatography-tandem mass spectrometry method (UPLC-MS/MS) is optimized and validated for reliable quantification of atorvastatin (ATR) and its active metabolites, 2-hydroxy atorvastatin (2-ATR) and 4-hydroxy atorvastatin (4-ATR) in human plasma using atorvastatin-D5 (ATR-D5), 2-hydroxy atorvastatin-D5 (2-ATR-D5), and 4-hydroxy atorvastatin-D5 (4-ATR-D5) as deuterium-labeled internal standards (ISTDs), respectively. Isocratic mode chromatographic separation was used with a reverse-phase C<sub>18</sub> Symmetry Shield (150 × 4.6 mm, 5.0 μm) column and a mobile phase of acetonitrile:2 mM ammonium formate (pH-3.0) [65:35%v/v] at a flow rate of 0.7 mL/min. Electrospray ionization technique with positive ion mode polarity was applied to achieve the best signal intensity and stable response. Solid-phase extraction by direct elution method was applied to extract the drugs from the plasma sample. The calibration curve range was validated from a concentration range of 0.500-250 ng/mL for ATR and 2-ATR and 0.200-20 ng/mL for 4-ATR. The within-batch and between-batch precision and accuracy were found to be consistent and reproducible for all the analytes across the validation. Extraction recoveries were >80% for all analytes and ISTDs. All peaks of analytes and the respective ISTDs were eluted within 5.2 min. In this validated method, selective multivariate analytical approaches were utilized such as best fit linearity range for different strength formulations, preventive measures for <i>in vivo</i> and <i>ex vivo</i> autodegradation of metabolites, and shorter analysis time. This validated method can be useful for challenging quantification of ATR and its active metabolites for therapeutic drug monitoring and in high-throughput clinical study sample analysis.</p>","PeriodicalId":8586,"journal":{"name":"Assay and drug development technologies","volume":"21 3","pages":"110-125"},"PeriodicalIF":1.6000,"publicationDate":"2023-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Assay and drug development technologies","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1089/adt.2022.113","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0
Abstract
A highly selective, sensitive, rugged, and rapid ultra high-performance liquid chromatography-tandem mass spectrometry method (UPLC-MS/MS) is optimized and validated for reliable quantification of atorvastatin (ATR) and its active metabolites, 2-hydroxy atorvastatin (2-ATR) and 4-hydroxy atorvastatin (4-ATR) in human plasma using atorvastatin-D5 (ATR-D5), 2-hydroxy atorvastatin-D5 (2-ATR-D5), and 4-hydroxy atorvastatin-D5 (4-ATR-D5) as deuterium-labeled internal standards (ISTDs), respectively. Isocratic mode chromatographic separation was used with a reverse-phase C18 Symmetry Shield (150 × 4.6 mm, 5.0 μm) column and a mobile phase of acetonitrile:2 mM ammonium formate (pH-3.0) [65:35%v/v] at a flow rate of 0.7 mL/min. Electrospray ionization technique with positive ion mode polarity was applied to achieve the best signal intensity and stable response. Solid-phase extraction by direct elution method was applied to extract the drugs from the plasma sample. The calibration curve range was validated from a concentration range of 0.500-250 ng/mL for ATR and 2-ATR and 0.200-20 ng/mL for 4-ATR. The within-batch and between-batch precision and accuracy were found to be consistent and reproducible for all the analytes across the validation. Extraction recoveries were >80% for all analytes and ISTDs. All peaks of analytes and the respective ISTDs were eluted within 5.2 min. In this validated method, selective multivariate analytical approaches were utilized such as best fit linearity range for different strength formulations, preventive measures for in vivo and ex vivo autodegradation of metabolites, and shorter analysis time. This validated method can be useful for challenging quantification of ATR and its active metabolites for therapeutic drug monitoring and in high-throughput clinical study sample analysis.
期刊介绍:
ASSAY and Drug Development Technologies provides access to novel techniques and robust tools that enable critical advances in early-stage screening. This research published in the Journal leads to important therapeutics and platforms for drug discovery and development. This reputable peer-reviewed journal features original papers application-oriented technology reviews, topical issues on novel and burgeoning areas of research, and reports in methodology and technology application.
ASSAY and Drug Development Technologies coverage includes:
-Assay design, target development, and high-throughput technologies-
Hit to Lead optimization and medicinal chemistry through preclinical candidate selection-
Lab automation, sample management, bioinformatics, data mining, virtual screening, and data analysis-
Approaches to assays configured for gene families, inherited, and infectious diseases-
Assays and strategies for adapting model organisms to drug discovery-
The use of stem cells as models of disease-
Translation of phenotypic outputs to target identification-
Exploration and mechanistic studies of the technical basis for assay and screening artifacts