Influence of Operator Experience on Scanning Time and Accuracy with Two Different Intraoral Scanners - A Prospective Clinical Trial.

IF 0.8 Q4 DENTISTRY, ORAL SURGERY & MEDICINE
Anjali Anna Thomas, Ravindra Kumar Jain
{"title":"Influence of Operator Experience on Scanning Time and Accuracy with Two Different Intraoral Scanners - A Prospective Clinical Trial.","authors":"Anjali Anna Thomas,&nbsp;Ravindra Kumar Jain","doi":"10.4274/TurkJOrthod.2022.2021.0220","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Operator experience and scanner type may influence the time taken and obtained accuracy of intraoral scanning. This study aimed to evaluate the influence of operator experience on the scanning time and correlate the accuracy of the scans taken with two different intraoral scanners (TRIOS 3, 3Shape and i500, Medit).</p><p><strong>Methods: </strong>In this trial, a total of 20 subjects who required intraoral scanning for orthodontic treatment were included. Intraoral scanning was done with two different scanners, TRIOS 3 and i500. One operator each with high (group 1), medium (group 2) and low (group 3) levels of experience performed intra-oral scanning with two different intraoral scanners. A One-Way ANOVA test was performed to assess the intergroup difference in scanning time and Kendall's tau's correlation test to determine the correlation between the experience of the operator and accuracy among the three groups using the two scanners. Also Independent samples t-test were performed to assess the intragroup differences in scanning time with two different scanners.</p><p><strong>Results: </strong>The scanning time was influenced by the type of intraoral scanner and operator experience (p<0.05). No significant correlation between operator experience and scanning accuracy in the three groups was noted (p>0.05). Statistically significant intragroup differences in scanning time between the two scanners were noted (p<0.05).</p><p><strong>Conclusion: </strong>Less experienced operators took more time to scan a subject. Accuracy of scanning among three groups using two scanners was not influenced by the experience of the operator. Scanning with i500 IOS took more time than TRIOS.</p>","PeriodicalId":37013,"journal":{"name":"Turkish Journal of Orthodontics","volume":"36 1","pages":"10-14"},"PeriodicalIF":0.8000,"publicationDate":"2023-03-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/e6/06/tjo-36-10.PMC10140657.pdf","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Turkish Journal of Orthodontics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4274/TurkJOrthod.2022.2021.0220","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 2

Abstract

Objective: Operator experience and scanner type may influence the time taken and obtained accuracy of intraoral scanning. This study aimed to evaluate the influence of operator experience on the scanning time and correlate the accuracy of the scans taken with two different intraoral scanners (TRIOS 3, 3Shape and i500, Medit).

Methods: In this trial, a total of 20 subjects who required intraoral scanning for orthodontic treatment were included. Intraoral scanning was done with two different scanners, TRIOS 3 and i500. One operator each with high (group 1), medium (group 2) and low (group 3) levels of experience performed intra-oral scanning with two different intraoral scanners. A One-Way ANOVA test was performed to assess the intergroup difference in scanning time and Kendall's tau's correlation test to determine the correlation between the experience of the operator and accuracy among the three groups using the two scanners. Also Independent samples t-test were performed to assess the intragroup differences in scanning time with two different scanners.

Results: The scanning time was influenced by the type of intraoral scanner and operator experience (p<0.05). No significant correlation between operator experience and scanning accuracy in the three groups was noted (p>0.05). Statistically significant intragroup differences in scanning time between the two scanners were noted (p<0.05).

Conclusion: Less experienced operators took more time to scan a subject. Accuracy of scanning among three groups using two scanners was not influenced by the experience of the operator. Scanning with i500 IOS took more time than TRIOS.

Abstract Image

Abstract Image

Abstract Image

操作者经验对两种不同口内扫描仪扫描时间和准确性的影响——一项前瞻性临床试验。
目的:操作人员经验和扫描仪类型对口腔内扫描所需时间和获得的准确性有影响。本研究旨在评估操作人员经验对扫描时间的影响,并将两种不同的口腔内扫描仪(TRIOS 3, 3Shape和i500, Medit)扫描的准确性联系起来。方法:本试验共纳入20例需要口腔内扫描进行正畸治疗的受试者。使用TRIOS 3和i500两种不同的扫描仪进行口内扫描。高(1组)、中(2组)和低(3组)经验的操作者分别使用两种不同的口腔内扫描仪进行口腔内扫描。采用单因素方差分析(One-Way ANOVA)检验评估扫描时间的组间差异,并采用Kendall's tau相关检验确定三组使用两种扫描仪的操作者经验与准确率之间的相关性。此外,采用独立样本t检验来评估两种不同扫描仪扫描时间的组内差异。结果:扫描时间受口腔内扫描仪类型和操作人员经验的影响(p0.05)。两种扫描仪在扫描时间上的组内差异具有统计学意义(p结论:经验较少的操作员花费更多的时间扫描一个受试者。三组使用两台扫描仪的扫描精度不受操作者经验的影响。i500的扫描时间比TRIOS要长。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Turkish Journal of Orthodontics
Turkish Journal of Orthodontics Dentistry-Orthodontics
CiteScore
2.10
自引率
9.10%
发文量
34
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信