Transfer Accuracy of Three Indirect Bonding Trays: An In Vitro Study with 3D Scanned Models.

IF 0.8 Q4 DENTISTRY, ORAL SURGERY & MEDICINE
Hilal Gündoğ, Ayça Arman Özçırpıcı, Hande Pamukçu
{"title":"Transfer Accuracy of Three Indirect Bonding Trays: An <i>In Vitro</i> Study with 3D Scanned Models.","authors":"Hilal Gündoğ,&nbsp;Ayça Arman Özçırpıcı,&nbsp;Hande Pamukçu","doi":"10.4274/TurkJOrthod.2022.2022.34","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The goal of the current study is to compare the transfer accuracy of two different conventional indirect bonding trays with 3D-printed trays.</p><p><strong>Methods: </strong>Twenty-two patients' upper dental models were duplicated, scanned and brackets were bonded digitally. Different indirect bonding trays (double vacuum formed, transparent silicone and 3D-printed) were prepared according to three groups. These trays were used for the transfer of the brackets to the patients' models, then models with brackets were scanned. GOM Inspect software was used for the superimposition of virtual bracket setups and models with brackets. A total of 788 brackets and tubes were analyzed. Transfer accuracies were determined according to the clinical limit of 0.5 mm for linear and 2° for angular measurements.</p><p><strong>Results: </strong>3D-printed trays had significantly lower linear deviation values than other trays for all planes (p<0.05). 3D-printed trays have significantly lower torque and tip deviation values than other groups (p<0.05). Transfer deviations were within the clinically acceptable limit for all transfer trays in horizontal, vertical and transverse planes. Deviation values of the molars were higher than those of the other tooth groups for all trays in the horizontal and vertical planes (p<0.05). Brackets were generally deviated toward the buccal direction in all tray groups.</p><p><strong>Conclusion: </strong>The transfer accuracy of 3D-printed transfer trays was more successful than the double vacuum formed and transparent silicone trays in the indirect bonding technique procedure. Deviations in the molar group were greater than those in the other tooth groups for all transfer trays.</p>","PeriodicalId":37013,"journal":{"name":"Turkish Journal of Orthodontics","volume":"36 1","pages":"1-9"},"PeriodicalIF":0.8000,"publicationDate":"2023-03-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/b5/cb/tjo-36-1.PMC10140660.pdf","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Turkish Journal of Orthodontics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4274/TurkJOrthod.2022.2022.34","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 1

Abstract

Objective: The goal of the current study is to compare the transfer accuracy of two different conventional indirect bonding trays with 3D-printed trays.

Methods: Twenty-two patients' upper dental models were duplicated, scanned and brackets were bonded digitally. Different indirect bonding trays (double vacuum formed, transparent silicone and 3D-printed) were prepared according to three groups. These trays were used for the transfer of the brackets to the patients' models, then models with brackets were scanned. GOM Inspect software was used for the superimposition of virtual bracket setups and models with brackets. A total of 788 brackets and tubes were analyzed. Transfer accuracies were determined according to the clinical limit of 0.5 mm for linear and 2° for angular measurements.

Results: 3D-printed trays had significantly lower linear deviation values than other trays for all planes (p<0.05). 3D-printed trays have significantly lower torque and tip deviation values than other groups (p<0.05). Transfer deviations were within the clinically acceptable limit for all transfer trays in horizontal, vertical and transverse planes. Deviation values of the molars were higher than those of the other tooth groups for all trays in the horizontal and vertical planes (p<0.05). Brackets were generally deviated toward the buccal direction in all tray groups.

Conclusion: The transfer accuracy of 3D-printed transfer trays was more successful than the double vacuum formed and transparent silicone trays in the indirect bonding technique procedure. Deviations in the molar group were greater than those in the other tooth groups for all transfer trays.

Abstract Image

Abstract Image

Abstract Image

三种间接连接托盘的传递精度:三维扫描模型的体外研究。
目的:本研究的目的是比较两种不同的传统间接粘接托盘与3d打印托盘的传递精度。方法:对22例患者进行上牙模型复制、扫描、托槽数字化粘接。根据三组制备了不同的间接粘接托盘(双真空成型、透明硅胶和3d打印)。这些托盘用于将托架转移到患者的模型中,然后扫描带有托架的模型。使用GOM Inspect软件对虚拟支架装置和带有支架的模型进行叠加。共分析了788个托管。传递精度根据临床极限0.5 mm线性和2°角测量确定。结果:3d打印托盘在各平面的线性偏差值均明显低于其他托盘(p)。结论:在间接粘接工艺过程中,3d打印托盘的转移精度优于双真空成型透明硅胶托盘。所有转移托盘中磨牙组的偏差均大于其他牙组。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Turkish Journal of Orthodontics
Turkish Journal of Orthodontics Dentistry-Orthodontics
CiteScore
2.10
自引率
9.10%
发文量
34
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信