An On-Demand Oxygen Flow Meter for Enhanced Patient Comfort and Reduced Oxygen Cost in Hospitals

IF 2.2 4区 医学 Q3 RESPIRATORY SYSTEM
E. Derom, E. Meijer, J. V. van Enschot
{"title":"An On-Demand Oxygen Flow Meter for Enhanced Patient Comfort and Reduced Oxygen Cost in Hospitals","authors":"E. Derom, E. Meijer, J. V. van Enschot","doi":"10.1080/15412555.2022.2078695","DOIUrl":null,"url":null,"abstract":"Abstract Background: Hypoxemia is currently treated in hospital wards with oxygen, released continuously by “conventional” flow meters. A new type of hybrid flow meter allows to switch between on-demand and continuous mode. The aim of this observational study was to assess whether this new device reduces oxygen expenditure, is well accepted in a hospital setting and improves patient comfort during oxygen therapy. Methods: Oxygen was administered in hypoxemic patients with conventional or hybrid flow meters to maintain an oxygen saturation of ≥ 92% over a 12-week period. Every two weeks conventional and hybrid flow meters were switched. The overall oxygen delivery to the ward was continuously measured with a data logging device installed in the main oxygen pipeline and corrected for multiple confounding factors. Humidity measurements, for which a sensor placed in front of one of the nostrils, and patient questionnaires, were used to assess patient comfort during continuous and on-demand flow. Results: Overall oxygen delivery decreased by 39% when switching from continuous flow to on-demand therapy after correction for confounding factors. Continuous flows significantly decreased relative humidity more than equivalent on-demand settings and the latter tended to increase comfort. Conclusions: Hybrid flow meters cause a significant reduction in oxygen delivery in a hospital ward, which may lead to financial savings. Using the on-demand technology also lowers the dryness of the upper airways (and may increase patient comfort), while maintaining an adequate oxygenation.","PeriodicalId":10704,"journal":{"name":"COPD: Journal of Chronic Obstructive Pulmonary Disease","volume":"30 1","pages":"274 - 281"},"PeriodicalIF":2.2000,"publicationDate":"2022-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"COPD: Journal of Chronic Obstructive Pulmonary Disease","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/15412555.2022.2078695","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RESPIRATORY SYSTEM","Score":null,"Total":0}
引用次数: 1

Abstract

Abstract Background: Hypoxemia is currently treated in hospital wards with oxygen, released continuously by “conventional” flow meters. A new type of hybrid flow meter allows to switch between on-demand and continuous mode. The aim of this observational study was to assess whether this new device reduces oxygen expenditure, is well accepted in a hospital setting and improves patient comfort during oxygen therapy. Methods: Oxygen was administered in hypoxemic patients with conventional or hybrid flow meters to maintain an oxygen saturation of ≥ 92% over a 12-week period. Every two weeks conventional and hybrid flow meters were switched. The overall oxygen delivery to the ward was continuously measured with a data logging device installed in the main oxygen pipeline and corrected for multiple confounding factors. Humidity measurements, for which a sensor placed in front of one of the nostrils, and patient questionnaires, were used to assess patient comfort during continuous and on-demand flow. Results: Overall oxygen delivery decreased by 39% when switching from continuous flow to on-demand therapy after correction for confounding factors. Continuous flows significantly decreased relative humidity more than equivalent on-demand settings and the latter tended to increase comfort. Conclusions: Hybrid flow meters cause a significant reduction in oxygen delivery in a hospital ward, which may lead to financial savings. Using the on-demand technology also lowers the dryness of the upper airways (and may increase patient comfort), while maintaining an adequate oxygenation.
一种按需氧流量计,提高患者舒适度,降低医院氧气成本
背景:目前在医院病房治疗低氧血症是通过“常规”流量计连续释放氧气。一种新型混合流量计允许按需和连续模式之间切换。这项观察性研究的目的是评估这种新设备是否减少了氧气消耗,在医院环境中被广泛接受,并提高了患者在氧气治疗期间的舒适度。方法:低氧血症患者用常规或混合式流量计给氧,维持血氧饱和度≥92%,持续12周。每两周更换常规流量计和混合流量计。通过安装在主供氧管道上的数据记录设备连续测量病房的总体供氧量,并对多种混杂因素进行校正。湿度测量(传感器放置在一个鼻孔前)和患者问卷,用于评估患者在连续和按需流动期间的舒适度。结果:在校正混杂因素后,从连续输氧切换到按需治疗时,总氧输送减少了39%。连续流动显著降低相对湿度比等效按需设置,后者倾向于增加舒适度。结论:混合流量计显著减少了医院病房的氧气输送,这可能会节省资金。使用按需技术还可以降低上呼吸道的干燥度(并可能增加患者的舒适度),同时保持充足的氧合。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
4.40
自引率
0.00%
发文量
38
审稿时长
6-12 weeks
期刊介绍: From pathophysiology and cell biology to pharmacology and psychosocial impact, COPD: Journal Of Chronic Obstructive Pulmonary Disease publishes a wide range of original research, reviews, case studies, and conference proceedings to promote advances in the pathophysiology, diagnosis, management, and control of lung and airway disease and inflammation - providing a unique forum for the discussion, design, and evaluation of more efficient and effective strategies in patient care.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信