Value of Mini Electrodes for Mapping Myocardial Arrhythmogenic Substrate—The Influence of Tip-to-Tissue Angulation and Irrigation Flow on Signal Quality

Karen Bickel, T. Lewalter, J. Fischer, C. Baumgartner, P. Hoppmann, K. Tiemann, C. Jilek
{"title":"Value of Mini Electrodes for Mapping Myocardial Arrhythmogenic Substrate—The Influence of Tip-to-Tissue Angulation and Irrigation Flow on Signal Quality","authors":"Karen Bickel, T. Lewalter, J. Fischer, C. Baumgartner, P. Hoppmann, K. Tiemann, C. Jilek","doi":"10.3390/jvd1010002","DOIUrl":null,"url":null,"abstract":"Background: The use of mini electrodes with a small surface and narrow electrode-to-electrode spacing is believed to lead to a higher electrical resolution. Until now, the effects of tip-to-tissue contact, angulation, and irrigation on signal quality and morphology are unknown. Methods: The beating heart of an open-chest pig was examined while controlling the angulation and contact between the catheter tip and myocardial tissue, as well as the irrigation of the catheter tip. The mini electrodes were mounted onto commercially available 8 mm non-irrigated and 4 mm irrigated tip catheters. Different electrode interconnections, angulations, contact forces, and irrigation flow were analyzed and compared to signals recorded from conventional electrodes. Results: A total of 63 electrode samples of 21 defined, stable settings, each lasting 30 s, were analyzed. (1) Tissue contact of mini electrodes was given as soon as the conventional tip electrode showed tissue contact. (2) Angulation of the tip-to-tissue contact showed a trend towards changes in the integral of signals derived from mini electrodes, and no significant changes were seen in signals derived from conventional or mini electrodes. (3) Irrigation flow surrounding the mini electrodes did not influence signals derived from mini electrodes, whereas conventional electrodes showed signals with a longer duration under higher irrigation. Conclusion: Mini electrodes are robust to contact force and irrigation flow regarding signal quality, whereas signals of conventional electrodes are affected by irrigation flow, leading to substantial changes in signal duration and kurtosis. Signals of mini electrodes are sensitive to local electrical changes because of a high local resolution.","PeriodicalId":74009,"journal":{"name":"Journal of vascular diseases","volume":"1 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-08-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of vascular diseases","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3390/jvd1010002","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The use of mini electrodes with a small surface and narrow electrode-to-electrode spacing is believed to lead to a higher electrical resolution. Until now, the effects of tip-to-tissue contact, angulation, and irrigation on signal quality and morphology are unknown. Methods: The beating heart of an open-chest pig was examined while controlling the angulation and contact between the catheter tip and myocardial tissue, as well as the irrigation of the catheter tip. The mini electrodes were mounted onto commercially available 8 mm non-irrigated and 4 mm irrigated tip catheters. Different electrode interconnections, angulations, contact forces, and irrigation flow were analyzed and compared to signals recorded from conventional electrodes. Results: A total of 63 electrode samples of 21 defined, stable settings, each lasting 30 s, were analyzed. (1) Tissue contact of mini electrodes was given as soon as the conventional tip electrode showed tissue contact. (2) Angulation of the tip-to-tissue contact showed a trend towards changes in the integral of signals derived from mini electrodes, and no significant changes were seen in signals derived from conventional or mini electrodes. (3) Irrigation flow surrounding the mini electrodes did not influence signals derived from mini electrodes, whereas conventional electrodes showed signals with a longer duration under higher irrigation. Conclusion: Mini electrodes are robust to contact force and irrigation flow regarding signal quality, whereas signals of conventional electrodes are affected by irrigation flow, leading to substantial changes in signal duration and kurtosis. Signals of mini electrodes are sensitive to local electrical changes because of a high local resolution.
微型电极对心肌致心律失常底物定位的价值——尖端与组织成角和灌流对信号质量的影响
背景:使用具有小表面和窄电极间距的微型电极被认为可以导致更高的电分辨率。到目前为止,尖端与组织接触、成角和灌溉对信号质量和形态的影响尚不清楚。方法:在控制导管端部与心肌组织成角、接触及导管端部冲洗的情况下,观察开胸猪跳动的心脏。微型电极安装在市售的8毫米非冲洗和4毫米冲洗尖端导管上。分析了不同电极的互连、角度、接触力和灌流量,并将其与传统电极记录的信号进行了比较。结果:共分析了21个稳定设置的63个电极样品,每个设置持续30秒。(1)传统尖端电极出现组织接触后,立即进行微型电极的组织接触。(2)尖端与组织接触的角度在微型电极的信号积分中呈现变化趋势,而传统电极和微型电极的信号积分没有明显变化。(3)微电极周围的灌水流量不影响微电极产生的信号,而传统电极在高灌水条件下信号持续时间更长。结论:在信号质量方面,微型电极对接触力和灌洗流量具有鲁棒性,而常规电极的信号受到灌洗流量的影响,导致信号持续时间和峰度发生实质性变化。微型电极的局部分辨率高,对局部电变化敏感。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信