Improvement of Measurement Precisions for Uranium Isotopes at Ultra Trace Levels by Modification of the Sample Introduction System in MC-ICP-MS

IF 0.4 Q4 SPECTROSCOPY
Ranhee Park, S. Lim, Sun-ho Han, Min Young Lee, Jinkyu Park, Chi-Gyu Lee, K. Song
{"title":"Improvement of Measurement Precisions for Uranium Isotopes at Ultra Trace Levels by Modification of the Sample Introduction System in MC-ICP-MS","authors":"Ranhee Park, S. Lim, Sun-ho Han, Min Young Lee, Jinkyu Park, Chi-Gyu Lee, K. Song","doi":"10.5478/MSL.2016.7.2.50","DOIUrl":null,"url":null,"abstract":": Multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) is currently used in our laboratory for isotopic and quantitative analyses of nuclear materials at ultra-trace levels in environmental swipe samples, which is a very useful for monitoring undeclared nuclear activities. In this study, to improve measurement precisions of uranium isotopes at ultra-trace levels, we adopted a desolvating nebulizer system (Aridus-II, CETAC., USA), which can improve signal sensitivity and reduce formation of uranium hydride. A peristaltic pump was combined with Aridus-II in the sample introduction system of MC-ICP-MS to reduce long-term signal fluctuations by maintaining a constant flow rate of the sample solution. The signal sensitivity in the presence of Aridus-II was improved more than 10-fold and the formation ratio of UH/U decreased by 16- to 17-fold compared to a normal spray chamber. Long-term signal fluctuations were significantly reduced by using the peristaltic pump. Detailed optimizations and evaluations with uranium standards are also discussed in this paper.","PeriodicalId":18238,"journal":{"name":"Mass Spectrometry Letters","volume":"6 1","pages":"50-54"},"PeriodicalIF":0.4000,"publicationDate":"2016-06-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Mass Spectrometry Letters","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5478/MSL.2016.7.2.50","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"SPECTROSCOPY","Score":null,"Total":0}
引用次数: 3

Abstract

: Multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) is currently used in our laboratory for isotopic and quantitative analyses of nuclear materials at ultra-trace levels in environmental swipe samples, which is a very useful for monitoring undeclared nuclear activities. In this study, to improve measurement precisions of uranium isotopes at ultra-trace levels, we adopted a desolvating nebulizer system (Aridus-II, CETAC., USA), which can improve signal sensitivity and reduce formation of uranium hydride. A peristaltic pump was combined with Aridus-II in the sample introduction system of MC-ICP-MS to reduce long-term signal fluctuations by maintaining a constant flow rate of the sample solution. The signal sensitivity in the presence of Aridus-II was improved more than 10-fold and the formation ratio of UH/U decreased by 16- to 17-fold compared to a normal spray chamber. Long-term signal fluctuations were significantly reduced by using the peristaltic pump. Detailed optimizations and evaluations with uranium standards are also discussed in this paper.
MC-ICP-MS样品导入系统改进对超痕量铀同位素测量精度的影响
多收集器电感耦合等离子体质谱(MC-ICP-MS)目前在我们的实验室中用于环境滑动样品中超痕量水平核材料的同位素和定量分析,这对于监测未申报的核活动非常有用。为了提高超痕量铀同位素的测量精度,本研究采用了一种脱溶雾化系统(Aridus-II, CETAC)。(美国),可以提高信号灵敏度,减少氢化铀的形成。在MC-ICP-MS的进样系统中,将蠕动泵与Aridus-II相结合,通过保持样品溶液的恒定流速来减少长期的信号波动。与普通喷雾室相比,Aridus-II存在时的信号灵敏度提高了10倍以上,UH/U的形成比降低了16 ~ 17倍。使用蠕动泵可显著降低长期信号波动。本文还详细讨论了用铀标准进行优化和评价的方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
0.90
自引率
20.00%
发文量
0
审稿时长
6 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信