Quantifying Stent Wall Shear Stress for Assessing Stent and Flow Diverter Performance for Treating Intracranial Aneurysms

T. Suess, S. Gent
{"title":"Quantifying Stent Wall Shear Stress for Assessing Stent and Flow Diverter Performance for Treating Intracranial Aneurysms","authors":"T. Suess, S. Gent","doi":"10.1115/1.4062107","DOIUrl":null,"url":null,"abstract":"\n Analysis of flow diverting intracranial aneurysm repair devices has traditionally focused on reducing intrasaccular blood flow velocity and aneurysm wall shear stress (WSS) as the primary metrics for improved perceived device performance. However, the interpretation of this data has been debated, particularly with regards to the specific biological benefits of high or low aneurysm WSS. Therefore, this research proposes an additional parameter of WSS at the stent struts that could provide valuable insight regarding the device's potential to promote occlusion at the aneurysm neck by indicating locations of increased platelet activation and microparticle shedding. Fluid flow effects were evaluated for two flow diverters (PipelineTM and FREDTM) and three stents (EnterpriseTM, AtlasTM, and LVISTM) using computational fluid dynamics (CFD) models developed from two patient-derived CTA datasets with aneurysms in the middle cerebral artery (MCA) and basilar artery (BA), respectively. The device WSS data provides an additional metric for evaluating the ability of the device to constrain the blood flow within the main vessel, as well as indicating potential locations of the initiation of aneurysm occlusion. It is hypothesized that high device WSS close to the aneurysm neck creates a higher likelihood of thrombus formation and aneurysm occlusion due to platelet activation and microparticle shedding, while high device WSS proximal or distal on the device would indicate higher likelihood of undesirable daughter vessel occlusion. Conversely, low-to-moderate device WSS throughout the device length could be interpreted as a reduced likelihood of complete occlusion of the aneurysm over time, resulting in lesser device performance.","PeriodicalId":73734,"journal":{"name":"Journal of engineering and science in medical diagnostics and therapy","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2023-03-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of engineering and science in medical diagnostics and therapy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1115/1.4062107","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Analysis of flow diverting intracranial aneurysm repair devices has traditionally focused on reducing intrasaccular blood flow velocity and aneurysm wall shear stress (WSS) as the primary metrics for improved perceived device performance. However, the interpretation of this data has been debated, particularly with regards to the specific biological benefits of high or low aneurysm WSS. Therefore, this research proposes an additional parameter of WSS at the stent struts that could provide valuable insight regarding the device's potential to promote occlusion at the aneurysm neck by indicating locations of increased platelet activation and microparticle shedding. Fluid flow effects were evaluated for two flow diverters (PipelineTM and FREDTM) and three stents (EnterpriseTM, AtlasTM, and LVISTM) using computational fluid dynamics (CFD) models developed from two patient-derived CTA datasets with aneurysms in the middle cerebral artery (MCA) and basilar artery (BA), respectively. The device WSS data provides an additional metric for evaluating the ability of the device to constrain the blood flow within the main vessel, as well as indicating potential locations of the initiation of aneurysm occlusion. It is hypothesized that high device WSS close to the aneurysm neck creates a higher likelihood of thrombus formation and aneurysm occlusion due to platelet activation and microparticle shedding, while high device WSS proximal or distal on the device would indicate higher likelihood of undesirable daughter vessel occlusion. Conversely, low-to-moderate device WSS throughout the device length could be interpreted as a reduced likelihood of complete occlusion of the aneurysm over time, resulting in lesser device performance.
量化支架壁剪切应力以评估支架和血流分流器治疗颅内动脉瘤的性能
传统上,对分流颅内动脉瘤修复装置的分析主要集中在降低囊内血流速度和动脉瘤壁剪切应力(WSS)上,这是改善感知装置性能的主要指标。然而,对这些数据的解释一直存在争议,特别是关于高或低动脉瘤WSS的特定生物学益处。因此,本研究提出了支架支撑处WSS的附加参数,该参数可以通过指示血小板活化增加和微粒脱落的位置,为该装置促进动脉瘤颈部闭塞的潜力提供有价值的见解。使用计算流体动力学(CFD)模型评估两种分流器(PipelineTM和FREDTM)和三种支架(EnterpriseTM, AtlasTM和LVISTM)的流体流动效果,这些模型分别来自两个大脑中动脉(MCA)和基底动脉(BA)动脉瘤患者的CTA数据集。设备WSS数据为评估设备限制主血管内血流的能力以及指示动脉瘤闭塞起始的潜在位置提供了额外的指标。假设靠近动脉瘤颈部的高装置WSS由于血小板活化和微粒脱落导致血栓形成和动脉瘤闭塞的可能性更高,而装置近端或远端的高装置WSS则表明更有可能出现不良的子血管闭塞。相反,在整个装置长度中,低至中等的装置WSS可以解释为随着时间的推移,动脉瘤完全闭塞的可能性降低,从而导致较低的装置性能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信