Influence of cyclic variation of right ventricular volume on left ventricular mechanical parameters measured with conductance catheter.

Y. Tanoue, Shigeki Morita, M. Hisahara, Ryuji Tominaga, Yoshito Kawachi, Hisataka Yasui
{"title":"Influence of cyclic variation of right ventricular volume on left ventricular mechanical parameters measured with conductance catheter.","authors":"Y. Tanoue, Shigeki Morita, M. Hisahara, Ryuji Tominaga, Yoshito Kawachi, Hisataka Yasui","doi":"10.1253/JCJ.65.749","DOIUrl":null,"url":null,"abstract":"The conductance catheter is widely used for the continuous measurement of the left ventricular (LV) pressure-volume loops. Cyclical change of the right ventricular (RV) volume may alter the parallel conductance volume, thereby affecting the LV mechanical parameters. Using 8 open-chest adult mongrel dogs, multiple LV pressure-volume loops were obtained by 2 methods: first with a vena cava occlusion (VCO) method, which involved RV volume alteration, and second with a right-heart-bypass (RHB) preparation, which decompressed the right ventricle completely. The slope of the end-systolic pressure-volume relation (Ees), the end-systolic volume associated with the end-systolic pressure of 100 mmHg (V100,es), stiffness constant (beta), and the end-diastolic volume associated with the end-diastolic pressure of 9 mmHg (V9,ed) were calculated from each loop. There was minimal influence from RV volume alteration on systolic-phase indices [Ees (VCO method, 6.37 +/- 1.91 mmHg/ml; RHB preparation, 6.60 +/- 1.66mmHg/ml; p=0.356), and V100,es (VCO method, 18.4 +/- 9.3ml; RHB preparation, 17.8 +/- 9.0 ml; p=0.681)], but there was a significant influence on diastolic-phase indices [beta (VCO method, 0.0599 +/- 0.0152; RHB preparation, 0.0839 +/- 0.0150; p=0.007), and V9,ed (VCO method, 35.6 +/- 11.3 ml; RHB preparation, 31.9 +/- 12.3 ml; p=0.001)]. The increase in the RV volume in the diastolic phase increased the parallel conductance volume, causing overestimation of the LV diastolic volume measured by the conductance catheter.","PeriodicalId":14544,"journal":{"name":"Japanese circulation journal","volume":"11 1","pages":"749-52"},"PeriodicalIF":0.0000,"publicationDate":"2001-07-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"6","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Japanese circulation journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1253/JCJ.65.749","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 6

Abstract

The conductance catheter is widely used for the continuous measurement of the left ventricular (LV) pressure-volume loops. Cyclical change of the right ventricular (RV) volume may alter the parallel conductance volume, thereby affecting the LV mechanical parameters. Using 8 open-chest adult mongrel dogs, multiple LV pressure-volume loops were obtained by 2 methods: first with a vena cava occlusion (VCO) method, which involved RV volume alteration, and second with a right-heart-bypass (RHB) preparation, which decompressed the right ventricle completely. The slope of the end-systolic pressure-volume relation (Ees), the end-systolic volume associated with the end-systolic pressure of 100 mmHg (V100,es), stiffness constant (beta), and the end-diastolic volume associated with the end-diastolic pressure of 9 mmHg (V9,ed) were calculated from each loop. There was minimal influence from RV volume alteration on systolic-phase indices [Ees (VCO method, 6.37 +/- 1.91 mmHg/ml; RHB preparation, 6.60 +/- 1.66mmHg/ml; p=0.356), and V100,es (VCO method, 18.4 +/- 9.3ml; RHB preparation, 17.8 +/- 9.0 ml; p=0.681)], but there was a significant influence on diastolic-phase indices [beta (VCO method, 0.0599 +/- 0.0152; RHB preparation, 0.0839 +/- 0.0150; p=0.007), and V9,ed (VCO method, 35.6 +/- 11.3 ml; RHB preparation, 31.9 +/- 12.3 ml; p=0.001)]. The increase in the RV volume in the diastolic phase increased the parallel conductance volume, causing overestimation of the LV diastolic volume measured by the conductance catheter.
右心室容积循环变化对电导导管测量左心室力学参数的影响。
导流导管广泛应用于左室压力-容积循环的连续测量。右心室容积的周期性变化可能改变平行电导容积,从而影响左室力学参数。8只开胸成年杂种犬,采用两种方法获得多个左室压力-容积环路:一种是腔静脉闭塞法(VCO),这涉及到右室容积的改变;另一种是右心旁路(RHB)准备,这是完全减压的右心室。从每个循环中计算收缩期末期压力-容积关系的斜率(Ees)、收缩期末期压力为100 mmHg时的收缩期末期容积(V100,es)、刚度常数(beta)和舒张末期压力为9 mmHg时的舒张末期容积(V9,ed)。左心室容积改变对收缩期指标的影响最小[Ees (VCO法),6.37 +/- 1.91 mmHg/ml;RHB制剂,6.60 +/- 1.66mmHg/ml;p=0.356), V100,es (VCO法,18.4±9.3ml;RHB制剂,17.8 +/- 9.0 ml;p=0.681)],但对舒张期指标有显著影响[β (VCO法,0.0599 +/- 0.0152;RHB制剂,0.0839 +/- 0.0150;p=0.007), V9,ed (VCO法,35.6±11.3 ml;RHB制剂,31.9 +/- 12.3 ml;p = 0.001)。舒张期左室容积增大,平行电导容积增大,导致电导导管测量的左室舒张容积高估。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信