Insertion method for minimizing fine needle deflection in bowel insertion based on experimental analysis

R. Tsumura, K. Shitashima, H. Iwata
{"title":"Insertion method for minimizing fine needle deflection in bowel insertion based on experimental analysis","authors":"R. Tsumura, K. Shitashima, H. Iwata","doi":"10.1109/IROS.2017.8202156","DOIUrl":null,"url":null,"abstract":"Accurate insertion of fine needles is difficult due to needle deflection. Needle deflection in the lower abdomen is particularly complex, as the needle has to pass through various tissues. As the area of the bowel is dominant in lower abdominal insertion, it is important to analyze the deflection during bowel insertion and to control the needle to minimize deflection. Few studies have focused on bowel insertion. We performed a fundamental deflection analysis of needle insertion in the bowel. Moreover, we have proposed an insertion method for minimizing the needle deflection based on our analysis. First, we performed needle insertion at various insertion positions and insertion angles into the hollow-shaped bowel, and determined the trend of the needle deflection during bowel insertion. The results revealed that the needle deflection was increased due to an increase in insertion angle, and therefore insertion angle should be minimized as much as possible. However, during actual bowel needle insertion, there are many situations in which the ideal path cannot be selected due to the arrangement of the bowel loops. We proposed an insertion method that can eliminate the total needle deflections during bowel insertion by controlling the needle tip direction at the breaching of each bowel wall. The results suggest that the needle deflection can be minimized by selecting the insertion path in which the sum of each insertion angle is zero. We verified the results of this insertion method in multiple bowels in in vivo experiments, and showed that it has the potential to be used in clinical practice.","PeriodicalId":6658,"journal":{"name":"2017 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS)","volume":"1 1","pages":"187-192"},"PeriodicalIF":0.0000,"publicationDate":"2017-12-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"6","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2017 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IROS.2017.8202156","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 6

Abstract

Accurate insertion of fine needles is difficult due to needle deflection. Needle deflection in the lower abdomen is particularly complex, as the needle has to pass through various tissues. As the area of the bowel is dominant in lower abdominal insertion, it is important to analyze the deflection during bowel insertion and to control the needle to minimize deflection. Few studies have focused on bowel insertion. We performed a fundamental deflection analysis of needle insertion in the bowel. Moreover, we have proposed an insertion method for minimizing the needle deflection based on our analysis. First, we performed needle insertion at various insertion positions and insertion angles into the hollow-shaped bowel, and determined the trend of the needle deflection during bowel insertion. The results revealed that the needle deflection was increased due to an increase in insertion angle, and therefore insertion angle should be minimized as much as possible. However, during actual bowel needle insertion, there are many situations in which the ideal path cannot be selected due to the arrangement of the bowel loops. We proposed an insertion method that can eliminate the total needle deflections during bowel insertion by controlling the needle tip direction at the breaching of each bowel wall. The results suggest that the needle deflection can be minimized by selecting the insertion path in which the sum of each insertion angle is zero. We verified the results of this insertion method in multiple bowels in in vivo experiments, and showed that it has the potential to be used in clinical practice.
基于实验分析的肠插入中最小化细针偏转的插入方法
由于针的偏转,细针难以准确插入。下腹的针偏转特别复杂,因为针必须穿过各种组织。由于下腹部穿刺以肠区为主,因此分析肠插入时的偏转情况并控制针头以减少偏转是很重要的。很少有研究关注肠插入。我们对针头插入肠道进行了基本的偏转分析。此外,在分析的基础上,我们提出了一种最小化针挠度的插入方法。首先,我们以不同的插入位置和角度将针插入空心肠内,并确定针在插入肠时的偏转趋势。结果表明,针的偏转由于插入角的增大而增大,因此应尽量减小插入角。然而,在实际的肠针插入过程中,由于肠袢的排列,很多情况下无法选择理想路径。我们提出了一种插入方法,通过控制每个肠壁的针尖方向,可以消除插入过程中针的总偏转。结果表明,选择各插入角之和为零的插入路径可以使针的偏转最小。我们在多个肠道的体内实验中验证了这种插入方法的结果,并表明它具有应用于临床实践的潜力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信