M. Hamamoto, Taira Kobayashi, Masamichi Ozawa, Kosuke Yoshimura
{"title":"Pure Cusp Tear of Trifecta Bioprosthesis 2 Years after Aortic Valve Replacement.","authors":"M. Hamamoto, Taira Kobayashi, Masamichi Ozawa, Kosuke Yoshimura","doi":"10.5761/atcs.cr.16-00183","DOIUrl":null,"url":null,"abstract":"Trifecta is a stented bioprosthetic heart valve with a bovine pericardial sheet externally mounted on a titanium stent. This valve is applied only for aortic valve replacement (AVR), providing excellent hemodynamics and extremely low incidence of structural valve deterioration (SVD). A 76-year-old woman presented with dyspnea on effort 24 months after AVR with a 21-mm Trifecta valve. Echocardiography revealed severe aortic regurgitation with prolapse of a cusp of Trifecta valve, which suggested that she developed acute heart failure due to early SVD. In the operation, Trifecta valve had a cusp tear near the commissure with circumferential fibrous pannus ingrowth only at the inflow side. There was neither calcification nor infection. The Trifecta valve was successfully replaced with a new porcine bioprosthesis.","PeriodicalId":93877,"journal":{"name":"Annals of thoracic and cardiovascular surgery : official journal of the Association of Thoracic and Cardiovascular Surgeons of Asia","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2017-06-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"23","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of thoracic and cardiovascular surgery : official journal of the Association of Thoracic and Cardiovascular Surgeons of Asia","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5761/atcs.cr.16-00183","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 23
Abstract
Trifecta is a stented bioprosthetic heart valve with a bovine pericardial sheet externally mounted on a titanium stent. This valve is applied only for aortic valve replacement (AVR), providing excellent hemodynamics and extremely low incidence of structural valve deterioration (SVD). A 76-year-old woman presented with dyspnea on effort 24 months after AVR with a 21-mm Trifecta valve. Echocardiography revealed severe aortic regurgitation with prolapse of a cusp of Trifecta valve, which suggested that she developed acute heart failure due to early SVD. In the operation, Trifecta valve had a cusp tear near the commissure with circumferential fibrous pannus ingrowth only at the inflow side. There was neither calcification nor infection. The Trifecta valve was successfully replaced with a new porcine bioprosthesis.