T. Washizuka, M. Chinushi, M. Tagawa, H. Kasai, Hiroshi Watanabe, Y. Hosaka, F. Yamashita, H. Furushima, Akira Abe, J. Hayashi, Y. Aizawa
{"title":"Inappropriate discharges by fourth generation implantable cardioverter defibrillators in patients with ventricular arrhythmias.","authors":"T. Washizuka, M. Chinushi, M. Tagawa, H. Kasai, Hiroshi Watanabe, Y. Hosaka, F. Yamashita, H. Furushima, Akira Abe, J. Hayashi, Y. Aizawa","doi":"10.1253/JCJ.65.927","DOIUrl":null,"url":null,"abstract":"The study prospectively investigated the incidence, cause and efficient management of inappropriate discharge by the fourth generation implantable cardioverter-defibrillator (ICD) system in 45 patients (mean age, 57+/-16 years). During the follow-up period of 27+/-17 months, 18 patients (40%) experienced one or more inappropriate therapies: sinus and supraventricular tachycardia (15 patients) and T wave oversensing (3 patients). In the 15 patients, re-programming of the tachycardia detection interval and/or additional treatment with beta-blocking agents were effective. In the 3 patients with T wave oversensing, the arrythmia was associated with an increase in T wave amplitude, change in T wave morphology and decreased R wave amplitude, and re-programming of the sensitivity of the local electrogram or changing the number of intervals to detect ventricular tachycardia decreased the number of inappropriate discharges in all 3 patients. In conclusion, inappropriate therapies are common problems in patients treated with the fourth generation ICD system, but most of them can be resolved using the dual-chamber ICD system. However, in patients with T-wave oversensing, it is difficult to avoid inappropriate discharge completely, even if the dual-chamber ICD system is implanted.","PeriodicalId":14544,"journal":{"name":"Japanese circulation journal","volume":"299302 1","pages":"927-30"},"PeriodicalIF":0.0000,"publicationDate":"2001-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"27","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Japanese circulation journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1253/JCJ.65.927","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 27
Abstract
The study prospectively investigated the incidence, cause and efficient management of inappropriate discharge by the fourth generation implantable cardioverter-defibrillator (ICD) system in 45 patients (mean age, 57+/-16 years). During the follow-up period of 27+/-17 months, 18 patients (40%) experienced one or more inappropriate therapies: sinus and supraventricular tachycardia (15 patients) and T wave oversensing (3 patients). In the 15 patients, re-programming of the tachycardia detection interval and/or additional treatment with beta-blocking agents were effective. In the 3 patients with T wave oversensing, the arrythmia was associated with an increase in T wave amplitude, change in T wave morphology and decreased R wave amplitude, and re-programming of the sensitivity of the local electrogram or changing the number of intervals to detect ventricular tachycardia decreased the number of inappropriate discharges in all 3 patients. In conclusion, inappropriate therapies are common problems in patients treated with the fourth generation ICD system, but most of them can be resolved using the dual-chamber ICD system. However, in patients with T-wave oversensing, it is difficult to avoid inappropriate discharge completely, even if the dual-chamber ICD system is implanted.