Inappropriate discharges by fourth generation implantable cardioverter defibrillators in patients with ventricular arrhythmias.

T. Washizuka, M. Chinushi, M. Tagawa, H. Kasai, Hiroshi Watanabe, Y. Hosaka, F. Yamashita, H. Furushima, Akira Abe, J. Hayashi, Y. Aizawa
{"title":"Inappropriate discharges by fourth generation implantable cardioverter defibrillators in patients with ventricular arrhythmias.","authors":"T. Washizuka, M. Chinushi, M. Tagawa, H. Kasai, Hiroshi Watanabe, Y. Hosaka, F. Yamashita, H. Furushima, Akira Abe, J. Hayashi, Y. Aizawa","doi":"10.1253/JCJ.65.927","DOIUrl":null,"url":null,"abstract":"The study prospectively investigated the incidence, cause and efficient management of inappropriate discharge by the fourth generation implantable cardioverter-defibrillator (ICD) system in 45 patients (mean age, 57+/-16 years). During the follow-up period of 27+/-17 months, 18 patients (40%) experienced one or more inappropriate therapies: sinus and supraventricular tachycardia (15 patients) and T wave oversensing (3 patients). In the 15 patients, re-programming of the tachycardia detection interval and/or additional treatment with beta-blocking agents were effective. In the 3 patients with T wave oversensing, the arrythmia was associated with an increase in T wave amplitude, change in T wave morphology and decreased R wave amplitude, and re-programming of the sensitivity of the local electrogram or changing the number of intervals to detect ventricular tachycardia decreased the number of inappropriate discharges in all 3 patients. In conclusion, inappropriate therapies are common problems in patients treated with the fourth generation ICD system, but most of them can be resolved using the dual-chamber ICD system. However, in patients with T-wave oversensing, it is difficult to avoid inappropriate discharge completely, even if the dual-chamber ICD system is implanted.","PeriodicalId":14544,"journal":{"name":"Japanese circulation journal","volume":"299302 1","pages":"927-30"},"PeriodicalIF":0.0000,"publicationDate":"2001-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"27","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Japanese circulation journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1253/JCJ.65.927","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 27

Abstract

The study prospectively investigated the incidence, cause and efficient management of inappropriate discharge by the fourth generation implantable cardioverter-defibrillator (ICD) system in 45 patients (mean age, 57+/-16 years). During the follow-up period of 27+/-17 months, 18 patients (40%) experienced one or more inappropriate therapies: sinus and supraventricular tachycardia (15 patients) and T wave oversensing (3 patients). In the 15 patients, re-programming of the tachycardia detection interval and/or additional treatment with beta-blocking agents were effective. In the 3 patients with T wave oversensing, the arrythmia was associated with an increase in T wave amplitude, change in T wave morphology and decreased R wave amplitude, and re-programming of the sensitivity of the local electrogram or changing the number of intervals to detect ventricular tachycardia decreased the number of inappropriate discharges in all 3 patients. In conclusion, inappropriate therapies are common problems in patients treated with the fourth generation ICD system, but most of them can be resolved using the dual-chamber ICD system. However, in patients with T-wave oversensing, it is difficult to avoid inappropriate discharge completely, even if the dual-chamber ICD system is implanted.
第四代植入式心律转复除颤器在室性心律失常患者中的不当出院。
本研究对45例患者(平均年龄57±16岁)采用第四代植入式心律转复除颤器(ICD)系统不当出院的发生率、原因及有效处理进行了前瞻性研究。在27+/-17个月的随访期间,18例(40%)患者出现了一种或多种不适当的治疗:窦性和室上性心动过速(15例)和T波过度敏感(3例)。在15例患者中,重新编程心动过速检测间隔和/或使用β -阻滞剂进行额外治疗是有效的。在3例T波过感患者中,心律失常与T波振幅升高、T波形态改变和R波振幅降低有关,重新编程局部电图的敏感性或改变检测室性心动过速的间隔次数可减少3例患者的不适当放电次数。综上所述,使用第四代ICD系统治疗患者的常见问题是治疗不当,但大多数可以通过双腔ICD系统解决。然而,对于t波过感患者,即使植入双腔ICD系统,也很难完全避免不适当的放电。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信