{"title":"Proposal of Brazilian National Program for Quality Control in the small animal PET scanners","authors":"R. Gontijo","doi":"10.15392/2319-0612.2022.1978","DOIUrl":null,"url":null,"abstract":"In Brazil, currently there are six preclinical molecular imaging centers and seven different small animal PET systems in use. However, there is still no national specific legislation for research laboratories focused on preclinical molecular images, unlike clinical nuclear medicine centers. A standardization of quality control protocols is needed to harmonize the use of small animal PET scanners in the research field. Thus, the aim of this work was to propose a National Program for Quality Control (NPQC) for the preclinical PET imaging systems in Brazil. This work was based on the quality control tests performed in Molecular Image Laboratory LIM/CDTN based on NEMA NU 4-2008 as described and presented in doctoral thesis of GONTIJO, 2019. The proposal of the NQCP was based on the CNEN NN 3.05 Publication (2013) and includes fundamental tests to be adopted, if possible, in laboratory practice to corroborate the experimental data. The Quality Control Program implemented in LIM/CDTN is an innovative and unprecedented proposal in the scope of preclinical molecular imaging services in Brazil. Therefore, the LIM/CDTN Program was carried out as a pilot to evidence the applicability and viability of the NQCP using small animal PET scanners. Summarizing, this work presents a viable set of Quality Control tests and their periodicities like a proposal for standardization to harmonize the use of this imaging technology in the field of research. Therefore, a National Quality Control Program (NQCP) applied to small animal PET scanners.","PeriodicalId":9203,"journal":{"name":"Brazilian Journal of Radiation Sciences","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-12-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Brazilian Journal of Radiation Sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.15392/2319-0612.2022.1978","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1
Abstract
In Brazil, currently there are six preclinical molecular imaging centers and seven different small animal PET systems in use. However, there is still no national specific legislation for research laboratories focused on preclinical molecular images, unlike clinical nuclear medicine centers. A standardization of quality control protocols is needed to harmonize the use of small animal PET scanners in the research field. Thus, the aim of this work was to propose a National Program for Quality Control (NPQC) for the preclinical PET imaging systems in Brazil. This work was based on the quality control tests performed in Molecular Image Laboratory LIM/CDTN based on NEMA NU 4-2008 as described and presented in doctoral thesis of GONTIJO, 2019. The proposal of the NQCP was based on the CNEN NN 3.05 Publication (2013) and includes fundamental tests to be adopted, if possible, in laboratory practice to corroborate the experimental data. The Quality Control Program implemented in LIM/CDTN is an innovative and unprecedented proposal in the scope of preclinical molecular imaging services in Brazil. Therefore, the LIM/CDTN Program was carried out as a pilot to evidence the applicability and viability of the NQCP using small animal PET scanners. Summarizing, this work presents a viable set of Quality Control tests and their periodicities like a proposal for standardization to harmonize the use of this imaging technology in the field of research. Therefore, a National Quality Control Program (NQCP) applied to small animal PET scanners.
在巴西,目前有6个临床前分子成像中心和7种不同的小动物PET系统在使用中。然而,与临床核医学中心不同,仍然没有针对临床前分子图像研究实验室的国家具体立法。为了协调小动物PET扫描仪在研究领域的使用,需要一个标准化的质量控制协议。因此,这项工作的目的是提出巴西临床前PET成像系统的国家质量控制计划(NPQC)。本工作基于GONTIJO 2019博士论文中描述并提出的基于NEMA NU 4-2008的分子图像实验室LIM/CDTN进行的质量控制测试。NQCP的建议是基于CNEN NN 3.05出版物(2013),并包括在实验室实践中采用的基本测试,如果可能的话,以证实实验数据。LIM/CDTN实施的质量控制计划是巴西临床前分子成像服务范围内的一项创新和前所未有的提议。因此,LIM/CDTN计划作为试点进行,以证明NQCP使用小动物PET扫描仪的适用性和可行性。总之,这项工作提出了一套可行的质量控制测试及其周期,如标准化建议,以协调该成像技术在研究领域的使用。因此,一个国家质量控制程序(NQCP)适用于小动物PET扫描仪。