Inappropriate discharges from an intravenous implantable cardioverter defibrillator due to T-wave oversensing.

T. Washizuka, M. Chinushi, H. Kasai, Hiroshi Watanabe, M. Tagawa, Y. Hosaka, Akira Abe, Y. Aizawa
{"title":"Inappropriate discharges from an intravenous implantable cardioverter defibrillator due to T-wave oversensing.","authors":"T. Washizuka, M. Chinushi, H. Kasai, Hiroshi Watanabe, M. Tagawa, Y. Hosaka, Akira Abe, Y. Aizawa","doi":"10.1253/JCJ.65.685","DOIUrl":null,"url":null,"abstract":"This report describes the clinical management of 2 patients with ventricular fibrillation (VF) who received inappropriate shocks from an implantable cardioverter defibrillator (ICD) due to T-wave oversensing. Cardiac sarcoidosis was confirmed as the underlying heart disease in 1 patient and idiopathic dilated cardiomyopathy in the other. Within 2 months after ICD implantation, both patients received several inappropriate shocks during sinus rhythm. Stored electrograms showed decreased R-wave amplitudes and increased T-wave amplitudes. The ICD sensed both R- and T-waves as ventricular activation, which met the rate criteria for VF treatment. Reprogramming the sensing threshold in association with administration of a drug to slow the heart rate decreased the incidence of the inappropriate shocks in both patients, but these palliative measures did not completely suppress the inappropriate shocks. To avoid T-wave oversensing, the repositioning or adding of a sensing lead is required. The potential risk of T-wave oversensing in ICD patients who have small R-wave amplitudes should be recognized.","PeriodicalId":14544,"journal":{"name":"Japanese circulation journal","volume":"2 1","pages":"685-7"},"PeriodicalIF":0.0000,"publicationDate":"2001-06-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"26","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Japanese circulation journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1253/JCJ.65.685","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 26

Abstract

This report describes the clinical management of 2 patients with ventricular fibrillation (VF) who received inappropriate shocks from an implantable cardioverter defibrillator (ICD) due to T-wave oversensing. Cardiac sarcoidosis was confirmed as the underlying heart disease in 1 patient and idiopathic dilated cardiomyopathy in the other. Within 2 months after ICD implantation, both patients received several inappropriate shocks during sinus rhythm. Stored electrograms showed decreased R-wave amplitudes and increased T-wave amplitudes. The ICD sensed both R- and T-waves as ventricular activation, which met the rate criteria for VF treatment. Reprogramming the sensing threshold in association with administration of a drug to slow the heart rate decreased the incidence of the inappropriate shocks in both patients, but these palliative measures did not completely suppress the inappropriate shocks. To avoid T-wave oversensing, the repositioning or adding of a sensing lead is required. The potential risk of T-wave oversensing in ICD patients who have small R-wave amplitudes should be recognized.
由于t波敏感过度导致静脉内植入式心律转复除颤器的不当出院。
本文报告2例室性颤动(VF)患者的临床处理,这些患者因t波过度敏感而受到植入式心律转复除颤器(ICD)不适当的电击。1例确诊为心脏结节病,另1例确诊为特发性扩张型心肌病。在ICD植入后2个月内,两例患者均在窦性心律期间接受了数次不适当的电击。存储电图显示r波振幅减小,t波振幅增大。ICD检测到R波和t波均为心室激活,符合室性心动过速治疗的速率标准。重新编程感应阈值与给药以减缓心率降低了两名患者不适当电击的发生率,但这些姑息措施并不能完全抑制不适当电击。为了避免t波过感,需要重新定位或增加感测引线。对于r波振幅较小的ICD患者,应认识到t波过感的潜在风险。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信