{"title":"Method Development and Validation for Tenofovir an Antiretroviral Drug in Plasma by LC-MS/MS Technique","authors":"Sarang Salunke, Sagar Wankhede, S. Medhe, H. Nimje, Samir Ranjan, Vikas Kendre, Payal Bhaskar","doi":"10.29228/jrp.463","DOIUrl":null,"url":null,"abstract":": Bioanalytical method development for Tenofovir (TFR) as an antiretroviral drug by LCMS Technique. A developed Bioanalytical analysis method for TFR can be used routinely in a commercial laboratory. All the solvents used were of HPLC grade. 4000 QTrap along with the Shimadzu LC 20AD LC System used to develop and validate the method. The LLOQ and LOQ for Tenofovir was found were 5ng/mL and 15ng/mL. The method was accurate (within ±15% of control) and precise (coefficient of variation ≤ 15%). Analytes were stable for five freeze/thaw cycles and up to 6 days at room temperature, whereas long-term at − 20°C or at − 80°C. For Precision study using QCs of the drug-85%, 100% and 115% concentration of drug chosen and the levels M1QC (75ng/mL), MQC (300ng/mL) and HQC (600ng/mL) where the % CV were observed of ≤ 15%. In a Precision and Accuracy study (inter day and intraday), the % CV obtained for Tenofovir was observed ≤ 15%. Recovery studies for extracted samples with LQC (15ng/mL), MQC (300ng/mL) and HQC (600ng/mL) were 94.51%, 91.83% and 90.91% respectively. Stability was within 15% deviation. The results of System Suitability Test for TFR and Acyclovir (ACR) are an internal standard with observed %CV ≤ 2.0%. The aim of the study was to develop a method that could be used as an alternative to the existing Tenofovir indirect method. The existing method observes separating the parent drug from the metabolite in LCMS/MS. This method is a good alternative to the indirect methods currently in use.","PeriodicalId":17096,"journal":{"name":"Journal of Research in Pharmacy","volume":"26 1","pages":""},"PeriodicalIF":0.6000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Research in Pharmacy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.29228/jrp.463","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PHARMACOLOGY & PHARMACY","Score":null,"Total":0}
引用次数: 0

Abstract

: Bioanalytical method development for Tenofovir (TFR) as an antiretroviral drug by LCMS Technique. A developed Bioanalytical analysis method for TFR can be used routinely in a commercial laboratory. All the solvents used were of HPLC grade. 4000 QTrap along with the Shimadzu LC 20AD LC System used to develop and validate the method. The LLOQ and LOQ for Tenofovir was found were 5ng/mL and 15ng/mL. The method was accurate (within ±15% of control) and precise (coefficient of variation ≤ 15%). Analytes were stable for five freeze/thaw cycles and up to 6 days at room temperature, whereas long-term at − 20°C or at − 80°C. For Precision study using QCs of the drug-85%, 100% and 115% concentration of drug chosen and the levels M1QC (75ng/mL), MQC (300ng/mL) and HQC (600ng/mL) where the % CV were observed of ≤ 15%. In a Precision and Accuracy study (inter day and intraday), the % CV obtained for Tenofovir was observed ≤ 15%. Recovery studies for extracted samples with LQC (15ng/mL), MQC (300ng/mL) and HQC (600ng/mL) were 94.51%, 91.83% and 90.91% respectively. Stability was within 15% deviation. The results of System Suitability Test for TFR and Acyclovir (ACR) are an internal standard with observed %CV ≤ 2.0%. The aim of the study was to develop a method that could be used as an alternative to the existing Tenofovir indirect method. The existing method observes separating the parent drug from the metabolite in LCMS/MS. This method is a good alternative to the indirect methods currently in use.
方法采用LC-MS/MS技术建立抗逆转录病毒药物替诺福韦的血浆检测方法并进行验证
:利用LCMS技术建立抗逆转录病毒药物替诺福韦(TFR)的生物分析方法。开发的TFR生物分析方法可在商业实验室常规使用。所用溶剂均为HPLC级。4000 QTrap以及岛津LC 20AD LC系统用于开发和验证该方法。替诺福韦的定量限分别为5ng/mL和15ng/mL。方法准确(在对照的±15%范围内),精密度高(变异系数≤15%)。在室温下,分析物在5次冻融循环和6天内是稳定的,而在- 20°C或- 80°C下则长期稳定。精密度研究采用所选药物浓度为85%、100%和115%的质量控制体系,以及百分比CV≤15%的M1QC (75ng/mL)、MQC (300ng/mL)和HQC (600ng/mL)水平。在精密度和准确度研究中(日间和日间),替诺福韦获得的CV %≤15%。LQC (15ng/mL)、MQC (300ng/mL)和HQC (600ng/mL)对提取样品的回收率分别为94.51%、91.83%和90.91%。稳定性偏差在15%以内。TFR和阿昔洛韦(ACR)的系统适宜性试验结果为内标,观察到的%CV≤2.0%。这项研究的目的是开发一种可以替代现有替诺福韦间接方法的方法。现有方法在LCMS/MS中观察到母体药物与代谢物的分离。这种方法是目前使用的间接方法的一种很好的替代方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Research in Pharmacy
Journal of Research in Pharmacy PHARMACOLOGY & PHARMACY-
CiteScore
1.00
自引率
12.50%
发文量
80
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信