Comparison of the blood gas analyser Edan i15 Vet with the blood gas analyser Gem Premier 3000 for the analysis of blood gases, electrolytes, hemoglobin and hematocrit in lactating Holstein

IF 0.4 4区 农林科学 Q4 VETERINARY SCIENCES
M. Metin, K. Aksoy, Aytaç Pekmezci, A. Deniz
{"title":"Comparison of the blood gas analyser Edan i15 Vet with the blood gas analyser Gem Premier 3000 for the analysis of blood gases, electrolytes, hemoglobin and hematocrit in lactating Holstein","authors":"M. Metin, K. Aksoy, Aytaç Pekmezci, A. Deniz","doi":"10.21521/mw.6728","DOIUrl":null,"url":null,"abstract":"A point-of-care blood gas analyser, Edan i15 Vet (EDAN), was compared with a benchtop blood gas analyser, GEM Premier 3000 (GEM). GEM and EDAN were used to analyse whole blood from 123 lactating Holsteins within one month of calving for blood gases, electrolytes, hematocrit and hemoglobin. EDAN and GEM showed significant linear correlations for blood gases, electrolytes, Hct and Hb. The 95% confidence intervals (CI) of intercept in Passing-Bablok regressions included zero in pCO2, bicarbonate, TCO2, BE ecf, BE B, hematocrit, and pH, but not for Na+, K+, pO2 and sO2. The CI of the slope included 1.0 for Na+, K+, pCO2, bicarbonate, TCO2, BE ecf, BE B, hematocrit, and pH, but not for pO2, sO2, and hemoglobin. The Bland-Altman plots between EDAN and GEM showed a bias of 1.4% for Na+, 2.4% for K+, –1.6% for pCO2, 3.0% for pO2, –5.3% for bicarbonate, 2.8% for SO2, –7.3% for TCO2, 10.4% for Hct, 21.2% for Hb, –25.1% for BE B and –38.5% for BE ecf. The biases in the analysis of certain estimated parameters were much higher (> 5%) than for measured parameters except for Hct. Parity did not correlate with blood gas parameters but blood pH correlated negatively with K+, pCO2 and positively with pO2, TCO2, sO2, bicarbonate, BE ecf and BE B. The postpartum time correlated positively with pCO2, TCO2, BE and bicarbonate and negatively with Hct and Hb values. Reference values (2.5-97.5% quartiles) were determined for each parameter. Conclusively, EDAN can be used interchangeably with GEM for the analysis of blood pH, K+, pCO2 and HCO3 – as they show acceptable moderate agreement. However, they did not agree for other parameters such as TCO2, BE ecf, BE B, Hb, Hct, sO2, Na+, and pO2, therefore, reference values for each parameter were set for GEM and EDAN.","PeriodicalId":49017,"journal":{"name":"Medycyna Weterynaryjna-Veterinary Medicine-Science and Practice","volume":null,"pages":null},"PeriodicalIF":0.4000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medycyna Weterynaryjna-Veterinary Medicine-Science and Practice","FirstCategoryId":"97","ListUrlMain":"https://doi.org/10.21521/mw.6728","RegionNum":4,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

A point-of-care blood gas analyser, Edan i15 Vet (EDAN), was compared with a benchtop blood gas analyser, GEM Premier 3000 (GEM). GEM and EDAN were used to analyse whole blood from 123 lactating Holsteins within one month of calving for blood gases, electrolytes, hematocrit and hemoglobin. EDAN and GEM showed significant linear correlations for blood gases, electrolytes, Hct and Hb. The 95% confidence intervals (CI) of intercept in Passing-Bablok regressions included zero in pCO2, bicarbonate, TCO2, BE ecf, BE B, hematocrit, and pH, but not for Na+, K+, pO2 and sO2. The CI of the slope included 1.0 for Na+, K+, pCO2, bicarbonate, TCO2, BE ecf, BE B, hematocrit, and pH, but not for pO2, sO2, and hemoglobin. The Bland-Altman plots between EDAN and GEM showed a bias of 1.4% for Na+, 2.4% for K+, –1.6% for pCO2, 3.0% for pO2, –5.3% for bicarbonate, 2.8% for SO2, –7.3% for TCO2, 10.4% for Hct, 21.2% for Hb, –25.1% for BE B and –38.5% for BE ecf. The biases in the analysis of certain estimated parameters were much higher (> 5%) than for measured parameters except for Hct. Parity did not correlate with blood gas parameters but blood pH correlated negatively with K+, pCO2 and positively with pO2, TCO2, sO2, bicarbonate, BE ecf and BE B. The postpartum time correlated positively with pCO2, TCO2, BE and bicarbonate and negatively with Hct and Hb values. Reference values (2.5-97.5% quartiles) were determined for each parameter. Conclusively, EDAN can be used interchangeably with GEM for the analysis of blood pH, K+, pCO2 and HCO3 – as they show acceptable moderate agreement. However, they did not agree for other parameters such as TCO2, BE ecf, BE B, Hb, Hct, sO2, Na+, and pO2, therefore, reference values for each parameter were set for GEM and EDAN.
血气分析仪Edan i15 Vet与血气分析仪Gem Premier 3000的比较,用于分析哺乳期荷斯坦犬的血气、电解质、血红蛋白和红细胞压积
将现场血气分析仪Edan i15 Vet (Edan)与台式血气分析仪GEM Premier 3000 (GEM)进行比较。采用GEM和EDAN对123头泌乳期荷斯坦奶牛1个月内的全血进行血气、电解质、红细胞压积和血红蛋白分析。EDAN和GEM与血气、电解质、Hct和Hb呈显著的线性相关。passin - bablok回归的95%置信区间(CI)包括pCO2、碳酸氢盐、TCO2、BE ecf、BE B、红细胞压积和pH为零,但Na+、K+、pO2和sO2不为零。Na+、K+、pCO2、碳酸氢盐、TCO2、BE ecf、BE B、红细胞压积和pH的斜率CI为1.0,但pO2、sO2和血红蛋白的斜率CI不为1.0。EDAN和GEM之间的Bland-Altman图显示Na+的偏差为1.4%,K+的偏差为2.4%,pCO2的偏差为-1.6%,pO2的偏差为3.0%,碳酸氢盐的偏差为-5.3%,SO2的偏差为2.8%,TCO2的偏差为-7.3%,Hct的偏差为10.4%,Hb的偏差为21.2%,BE B的偏差为-25.1%,BE ecf的偏差为-38.5%。除Hct外,某些估计参数的分析偏差比测量参数高得多(约5%)。胎次与血气参数无相关性,而血pH值与K+、pCO2呈负相关,与pO2、TCO2、sO2、碳酸氢盐、BE、ecf、BE b呈正相关。产后时间与pCO2、TCO2、BE、碳酸氢盐呈正相关,与Hct、Hb值呈负相关。为每个参数确定参考值(2.5-97.5%四分位数)。最后,EDAN可以与GEM互换用于血液pH、K+、pCO2和HCO3 -的分析,因为它们显示出可接受的适度一致性。但对于TCO2、BE ecf、BE B、Hb、Hct、sO2、Na+、pO2等参数,双方意见不一致,因此对GEM和EDAN设置了各参数的参考值。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
0.80
自引率
0.00%
发文量
73
审稿时长
4-8 weeks
期刊介绍: "Medycyna Weterynaryjna" publishes various types of articles which are grouped in the following editorial categories: reviews, original studies, scientific and professional problems, the history of veterinary medicine, posthumous memoirs, as well as chronicles that briefly relate scientific advances and developments in the veterinary profession and medicine. The most important are the first two categories, which are published with short summaries in English. Moreover, from 2001 the editors of "Medycyna Weterynaryjna", bearing in mind market demands, has also started publishing entire works in English. Since 2008 the periodical has appeared in an electronic version. The following are available in this version: summaries of studies published from 1999 to 2005, full versions of all the studies published in the years 2006-2011 (in pdf files), and full versions of the English studies published in the current year (pdf). Only summaries of the remaining studies from the current year are available. In accordance with the principles accepted by the editors, the full versions of these texts will not be made available until next year. All articles are evaluated twice by leading Polish scientists and professionals before they are considered for publication. For years now "Medycyna Weterynaryjna" has maintained a high standard thanks to this system. The review articles are actually succinct monographs dealing with specific scientific and professional problems that are based on the most recent findings. Original works have a particular value, since they present research carried out in Polish and international scientific centers.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信