Technical ultrasonic quality assurance in Danish Radiology Departments

Karen Brage , Kristina T.T. Pank , Sisse Hansen , Lis K. Sondergaard , Mark F. McEntee , Malene Roland V. Pedersen
{"title":"Technical ultrasonic quality assurance in Danish Radiology Departments","authors":"Karen Brage ,&nbsp;Kristina T.T. Pank ,&nbsp;Sisse Hansen ,&nbsp;Lis K. Sondergaard ,&nbsp;Mark F. McEntee ,&nbsp;Malene Roland V. Pedersen","doi":"10.1016/j.wfumbo.2023.100005","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>In general, there is a lack of technical quality assurance (TQA) in ultrasound, which threatens patients. This study aimed to map the level of ultrasound quality assurance in Danish Radiology Departments, assess the intra- and interrater reliability of the in-air method for evaluating transducer quality, and determine the prevalence of faulty transducers.</p></div><div><h3>Methods</h3><p>From October to November 2021, an anonymous survey focused on the extent of TQA was e-mailed to clinicians (via managers) responsible for ultrasound TQA. The six-item survey included both closed and open-ended questions. In addition, a sample of 63 ultrasound b-mode reverberation images was collected from four Danish Radiological Departments, later rated by two blinded radiographers on a dichotomous scale.</p></div><div><h3>Results</h3><p>A total of 46 participants responded to the survey. In the survey, 15 respondents reported not having a TQA procedure, while six thought a procedure existed but were uncertain. Various methods were reported for the respondents who answered that TQA was performed (n = 13).</p><p>Half of the respondents (n = 23, 50%) reported experiencing transducer malfunctions, and seven said they did not find the lack of regular testing problematic.</p><p>The interrater reliability was κ = 0.75 (95% CI.: 0.61–0.89), and the inter-rater reliability was κ = 0.84 (95% CI.: 0.70–0.98). Based on the agreement between the two raters, 20 out of 50 (40%) transducers were flawed.</p></div><div><h3>Conclusion</h3><p>This study demonstrates the general deficiency of TQA in Danish Radiological Departments, with 21 clinicians responsible for ultrasound TQA responding either not having a TQA procedure or being uncertain if one exists. Furthermore, the study reveals that the in-air method for assessing transducer quality demonstrates moderate to almost perfect reliability and that nearly half of the transducer tested exhibits visual faults.</p></div>","PeriodicalId":101281,"journal":{"name":"WFUMB Ultrasound Open","volume":"1 1","pages":"Article 100005"},"PeriodicalIF":0.0000,"publicationDate":"2023-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"WFUMB Ultrasound Open","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2949668323000058","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Introduction

In general, there is a lack of technical quality assurance (TQA) in ultrasound, which threatens patients. This study aimed to map the level of ultrasound quality assurance in Danish Radiology Departments, assess the intra- and interrater reliability of the in-air method for evaluating transducer quality, and determine the prevalence of faulty transducers.

Methods

From October to November 2021, an anonymous survey focused on the extent of TQA was e-mailed to clinicians (via managers) responsible for ultrasound TQA. The six-item survey included both closed and open-ended questions. In addition, a sample of 63 ultrasound b-mode reverberation images was collected from four Danish Radiological Departments, later rated by two blinded radiographers on a dichotomous scale.

Results

A total of 46 participants responded to the survey. In the survey, 15 respondents reported not having a TQA procedure, while six thought a procedure existed but were uncertain. Various methods were reported for the respondents who answered that TQA was performed (n = 13).

Half of the respondents (n = 23, 50%) reported experiencing transducer malfunctions, and seven said they did not find the lack of regular testing problematic.

The interrater reliability was κ = 0.75 (95% CI.: 0.61–0.89), and the inter-rater reliability was κ = 0.84 (95% CI.: 0.70–0.98). Based on the agreement between the two raters, 20 out of 50 (40%) transducers were flawed.

Conclusion

This study demonstrates the general deficiency of TQA in Danish Radiological Departments, with 21 clinicians responsible for ultrasound TQA responding either not having a TQA procedure or being uncertain if one exists. Furthermore, the study reveals that the in-air method for assessing transducer quality demonstrates moderate to almost perfect reliability and that nearly half of the transducer tested exhibits visual faults.

丹麦放射科超声技术质量保证
一般来说,超声缺乏技术质量保证(TQA),这对患者构成了威胁。本研究旨在绘制丹麦放射科超声质量保证水平图,评估评估换能器质量的空气中方法的内部和内部可靠性,并确定故障换能器的普遍性。方法2021年10月至11月,一项关于TQA程度的匿名调查通过电子邮件发送给负责超声TQA的临床医生(通过经理)。这项六项调查包括封闭式和开放式问题。此外,从四个丹麦放射部门收集了63张超声b模式混响图像样本,随后由两名盲法放射技师以二分制进行评分。结果共有46名参与者对调查做出了回应。在调查中,15名受访者表示没有TQA程序,而6名受访者认为存在程序,但不确定。报告了回答进行TQA的受访者的各种方法(n=13)。一半的受访者(n=23,50%)报告说遇到了传感器故障,7人表示他们没有发现缺乏定期测试的问题。评分者间的可靠性为κ=0.75(95%可信区间:0.61–0.89),评分者间可靠性为κ=0.84(95%置信区间:0.70–0.98)。根据两位评分者之间的一致性,50个传感器中有20个(40%)存在缺陷。结论本研究证明了丹麦放射科TQA的普遍不足,21名负责超声TQA的临床医生要么没有TQA程序,要么不确定是否存在TQA程序。此外,研究表明,用于评估换能器质量的空气中方法显示出中等到几乎完美的可靠性,并且近一半的测试换能器显示出视觉故障。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信