Position-Dependent Dysfunction of Deep Brain Stimulation in Tourette Syndrome: Diagnostic Clues

IF 2.5 Q2 CLINICAL NEUROLOGY
M. Wolf, Olaf Majewski, K. Müller-Vahl, C. Blahak, D. Schulte, J. Krauss
{"title":"Position-Dependent Dysfunction of Deep Brain Stimulation in Tourette Syndrome: Diagnostic Clues","authors":"M. Wolf, Olaf Majewski, K. Müller-Vahl, C. Blahak, D. Schulte, J. Krauss","doi":"10.7916/tohm.v0.713","DOIUrl":null,"url":null,"abstract":"Background Detection of defective deep brain stimulation (DBS) contacts/electrodes is sometimes challenging. Case Report We report a patient with Tourette syndrome (TS), who presented with abrupt tic increase and mild generalized headache 9 years after DBS implantation. On the suspicion of a hardware defect, a fracture of the DBS electrode and extension lead was ruled out by radiography and standard implantable pulse generator readouts. Further investigation revealed position-dependent modifiable therapeutic impedances, suggesting an impaired contact of the extension lead/adaptor. After replacement normal impedances were recorded, and the patient fully recovered. Discussion In DBS dysfunction with inconspicuous hardware check, position-dependent defects might be suspected.","PeriodicalId":23317,"journal":{"name":"Tremor and Other Hyperkinetic Movements","volume":" ","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2019-10-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Tremor and Other Hyperkinetic Movements","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.7916/tohm.v0.713","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 1

Abstract

Background Detection of defective deep brain stimulation (DBS) contacts/electrodes is sometimes challenging. Case Report We report a patient with Tourette syndrome (TS), who presented with abrupt tic increase and mild generalized headache 9 years after DBS implantation. On the suspicion of a hardware defect, a fracture of the DBS electrode and extension lead was ruled out by radiography and standard implantable pulse generator readouts. Further investigation revealed position-dependent modifiable therapeutic impedances, suggesting an impaired contact of the extension lead/adaptor. After replacement normal impedances were recorded, and the patient fully recovered. Discussion In DBS dysfunction with inconspicuous hardware check, position-dependent defects might be suspected.
抽动秽语综合征深部脑刺激的位置依赖性功能障碍:诊断线索
背景检测有缺陷的深部脑刺激(DBS)接触/电极有时具有挑战性。病例报告我们报告了一名患有抽动秽语综合征(TS)的患者,他在DBS植入9年后出现抽搐突然加重和轻度全身性头痛。在怀疑硬件缺陷的情况下,通过射线照相和标准植入式脉冲发生器读数排除了DBS电极和延长导线断裂的可能性。进一步的研究揭示了位置依赖性可改变的治疗阻抗,表明延长导线/适配器的接触受损。更换后,记录正常阻抗,患者完全康复。讨论在DBS功能障碍伴有不明显硬件检查的情况下,可能怀疑存在位置依赖性缺陷。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
4.00
自引率
4.50%
发文量
31
审稿时长
6 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信