N. Y. Jagath B. Nikapitiya, Mun Mun Nahar, Hyejin Moon
{"title":"Accurate, consistent, and fast droplet splitting and dispensing in electrowetting on dielectric digital microfluidics","authors":"N. Y. Jagath B. Nikapitiya, Mun Mun Nahar, Hyejin Moon","doi":"10.1186/s40486-017-0058-6","DOIUrl":null,"url":null,"abstract":"<p>This letter reports two novel electrode design considerations to satisfy two very important aspects of EWOD operation—(1) Highly <i>consistent</i> volume of generated droplets and (2) Highly improved <i>accuracy</i> in the generated droplet volume. Considering the design principles investigated two novel designs were proposed; L-junction electrode design to offer high throughput droplet generation and Y-junction electrode design to split a droplet very fast while maintaining equal volume of each part. Devices of novel designs were fabricated and tested, and the results are compared with those of conventional approach. It is demonstrated that inaccuracy and inconsistency of droplet volume dispensed in the device with novel electrode designs are as low as 0.17 and 0.10%, respectively, while those of conventional approach are 25 and 0.76%, respectively. The dispensing frequency is enhanced from 4 to 9?Hz by using the novel design.</p>","PeriodicalId":704,"journal":{"name":"Micro and Nano Systems Letters","volume":"5 1","pages":""},"PeriodicalIF":4.7000,"publicationDate":"2017-06-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1186/s40486-017-0058-6","citationCount":"31","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Micro and Nano Systems Letters","FirstCategoryId":"1085","ListUrlMain":"https://link.springer.com/article/10.1186/s40486-017-0058-6","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"NANOSCIENCE & NANOTECHNOLOGY","Score":null,"Total":0}
引用次数: 31
Abstract
This letter reports two novel electrode design considerations to satisfy two very important aspects of EWOD operation—(1) Highly consistent volume of generated droplets and (2) Highly improved accuracy in the generated droplet volume. Considering the design principles investigated two novel designs were proposed; L-junction electrode design to offer high throughput droplet generation and Y-junction electrode design to split a droplet very fast while maintaining equal volume of each part. Devices of novel designs were fabricated and tested, and the results are compared with those of conventional approach. It is demonstrated that inaccuracy and inconsistency of droplet volume dispensed in the device with novel electrode designs are as low as 0.17 and 0.10%, respectively, while those of conventional approach are 25 and 0.76%, respectively. The dispensing frequency is enhanced from 4 to 9?Hz by using the novel design.