Fetal Organ Dose Assessment during Chest CT Examination Using Monte Carlo/Gate Simulation

IF 0.4 Q4 PHYSICS, PARTICLES & FIELDS
Y. Benameur, M. Tahiri, M. Mkimel, R. El Baydaoui, B. El Hariri, M.R. Mesradi, A. Hilali, E. Saad
{"title":"Fetal Organ Dose Assessment during Chest CT Examination Using Monte Carlo/Gate Simulation","authors":"Y. Benameur,&nbsp;M. Tahiri,&nbsp;M. Mkimel,&nbsp;R. El Baydaoui,&nbsp;B. El Hariri,&nbsp;M.R. Mesradi,&nbsp;A. Hilali,&nbsp;E. Saad","doi":"10.1134/S1547477123040088","DOIUrl":null,"url":null,"abstract":"<p>A CT scan of a pregnant patient is often a source of distress for the patient and staff. In addition, patients are usually worried about the risk of unfavorable effects on the fetus from radiation. Therefore, assessing radiation dose and related risks to the fetus and pregnant patient is an important aspect of radiation protection. The aim of this study is to estimate the fetal organ and effective dose during a chest CT protocol for a pregnant patient. For this purpose, we model the SOMATOM EMOTION 16 CT (Siemens Medical Solutions, Erlangen, Germany) and the pregnant patient using GATE code. The CT scanner is modelled based on the data provided by the supplier, operating at 80, 110 and 130 kV and 0.5–1.5 for the pitch, regarding the patient, she is modelled with a voxelized pregnant phantom (KATJA, 29 yr old) at the 24th week. Fetal effective doses are estimated to be 0.8, 1.7 and 2.4 mSv for 80, 110 and 130 kV, respectively. When the energy is decreased from 130 to 80 kV, the dose to the fetal organs in the heart and lenses is reduced by 64.3 and 64.6%. Moreover, the fetus organ dose is reduced by 21.5, 19.2, 30.0 and 17.6% for crane, eyes, heart and brain for a pitch ranging from 0.5 to 1.5. Since the cumulative dose to the fetus does not exceed 100 mGy, the doses to the fetus are considered acceptable.</p>","PeriodicalId":730,"journal":{"name":"Physics of Particles and Nuclei Letters","volume":"20 4","pages":"690 - 698"},"PeriodicalIF":0.4000,"publicationDate":"2023-08-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics of Particles and Nuclei Letters","FirstCategoryId":"1085","ListUrlMain":"https://link.springer.com/article/10.1134/S1547477123040088","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PHYSICS, PARTICLES & FIELDS","Score":null,"Total":0}
引用次数: 0

Abstract

A CT scan of a pregnant patient is often a source of distress for the patient and staff. In addition, patients are usually worried about the risk of unfavorable effects on the fetus from radiation. Therefore, assessing radiation dose and related risks to the fetus and pregnant patient is an important aspect of radiation protection. The aim of this study is to estimate the fetal organ and effective dose during a chest CT protocol for a pregnant patient. For this purpose, we model the SOMATOM EMOTION 16 CT (Siemens Medical Solutions, Erlangen, Germany) and the pregnant patient using GATE code. The CT scanner is modelled based on the data provided by the supplier, operating at 80, 110 and 130 kV and 0.5–1.5 for the pitch, regarding the patient, she is modelled with a voxelized pregnant phantom (KATJA, 29 yr old) at the 24th week. Fetal effective doses are estimated to be 0.8, 1.7 and 2.4 mSv for 80, 110 and 130 kV, respectively. When the energy is decreased from 130 to 80 kV, the dose to the fetal organs in the heart and lenses is reduced by 64.3 and 64.6%. Moreover, the fetus organ dose is reduced by 21.5, 19.2, 30.0 and 17.6% for crane, eyes, heart and brain for a pitch ranging from 0.5 to 1.5. Since the cumulative dose to the fetus does not exceed 100 mGy, the doses to the fetus are considered acceptable.

Abstract Image

利用蒙特卡罗/门模拟评估胸部CT检查中的胎儿器官剂量
怀孕病人的CT扫描通常是病人和工作人员痛苦的来源。此外,患者通常担心辐射对胎儿的不利影响。因此,评估辐射剂量及其对胎儿和孕妇的风险是辐射防护的一个重要方面。本研究的目的是估计胎儿器官和有效剂量期间胸部CT协议的孕妇。为此,我们使用GATE代码对SOMATOM EMOTION 16 CT (Siemens Medical Solutions, Erlangen, Germany)和孕妇进行建模。CT扫描仪是根据供应商提供的数据建模的,工作电压为80,110和130千伏,螺距为0.5-1.5。对于患者,她在24周时用体素化怀孕幻影(KATJA, 29岁)建模。胎儿有效剂量估计分别为0.8、1.7和2.4毫西弗,分别为80、110和130千伏。当能量从130 kV降低到80 kV时,对胎儿心脏和晶状体器官的剂量分别减少64.3%和64.6%。在0.5 ~ 1.5的音高范围内,鹤、眼、心、脑的胎儿器官剂量分别降低21.5%、19.2%、30.0%和17.6%。由于对胎儿的累积剂量不超过100毫戈瑞,因此对胎儿的剂量被认为是可接受的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Physics of Particles and Nuclei Letters
Physics of Particles and Nuclei Letters PHYSICS, PARTICLES & FIELDS-
CiteScore
0.80
自引率
20.00%
发文量
108
期刊介绍: The journal Physics of Particles and Nuclei Letters, brief name Particles and Nuclei Letters, publishes the articles with results of the original theoretical, experimental, scientific-technical, methodological and applied research. Subject matter of articles covers: theoretical physics, elementary particle physics, relativistic nuclear physics, nuclear physics and related problems in other branches of physics, neutron physics, condensed matter physics, physics and engineering at low temperatures, physics and engineering of accelerators, physical experimental instruments and methods, physical computation experiments, applied research in these branches of physics and radiology, ecology and nuclear medicine.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信