Anti-rotation pins for the compress implant do not increase risk of mechanical failure or impair osseointegration

IF 0.5 4区 医学 Q4 ORTHOPEDICS
S. Campbell, R. Steffner, Andrea K. Finlay, D. Mohler, R. Avedian
{"title":"Anti-rotation pins for the compress implant do not increase risk of mechanical failure or impair osseointegration","authors":"S. Campbell, R. Steffner, Andrea K. Finlay, D. Mohler, R. Avedian","doi":"10.21037/AOJ.2019.08.02","DOIUrl":null,"url":null,"abstract":"Background: The use of compliant fixation for endoprosthetic implants is gaining popularity. Previous work has shown that anti-rotation pins improve rotational stability at the bone-implant interface, but there is concern that these pins lead to increased risk of mechanical failure. We asked: (I) are anti-rotation pins used with the Compress implant associated with mechanical failure? (II) Are anti-rotation pins associated with less effective osseointegration? Methods: We performed retrospective review of cases using a Compress implant from 2004–2016. Mechanical failure rates and bone growth at the bone-implant interface were compared between pin and no-pin cohorts. Regression models were used to examine patient and surgical factors associated with mechanical failure. \n Results: Anti-rotation pins were not associated with mechanical failure (P=1.0, odds ratio 1.17, 95% confidence interval: 0.12–15.40). Anti-rotation pins were not associated with impaired osseointegration at any time point (P=1.0 at 3–6 months, P=0.33 at 6-9 months, P=0.34 at 9–12 months, P=0.40 at 12–24 months, P=0.28 at 24–48 months, P=1.0 at >48 months). No patient or surgical variables were predictors of mechanical failure. \n Conclusions: Anti-rotation pins were not associated with mechanical failure or impaired osseointegration.","PeriodicalId":44459,"journal":{"name":"Annals of Joint","volume":" ","pages":""},"PeriodicalIF":0.5000,"publicationDate":"2019-08-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Joint","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.21037/AOJ.2019.08.02","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The use of compliant fixation for endoprosthetic implants is gaining popularity. Previous work has shown that anti-rotation pins improve rotational stability at the bone-implant interface, but there is concern that these pins lead to increased risk of mechanical failure. We asked: (I) are anti-rotation pins used with the Compress implant associated with mechanical failure? (II) Are anti-rotation pins associated with less effective osseointegration? Methods: We performed retrospective review of cases using a Compress implant from 2004–2016. Mechanical failure rates and bone growth at the bone-implant interface were compared between pin and no-pin cohorts. Regression models were used to examine patient and surgical factors associated with mechanical failure. Results: Anti-rotation pins were not associated with mechanical failure (P=1.0, odds ratio 1.17, 95% confidence interval: 0.12–15.40). Anti-rotation pins were not associated with impaired osseointegration at any time point (P=1.0 at 3–6 months, P=0.33 at 6-9 months, P=0.34 at 9–12 months, P=0.40 at 12–24 months, P=0.28 at 24–48 months, P=1.0 at >48 months). No patient or surgical variables were predictors of mechanical failure. Conclusions: Anti-rotation pins were not associated with mechanical failure or impaired osseointegration.
用于压缩种植体的防旋转销钉不会增加机械故障或损害骨整合的风险
背景:顺应性固定在内修复植入物中的应用越来越受欢迎。先前的研究表明,防旋转销可以提高骨植入物界面的旋转稳定性,但人们担心这些销会增加机械故障的风险。我们问:(I)Compress植入物使用的防旋转销是否与机械故障有关?(II) 抗旋转销是否与无效的骨整合有关?方法:我们对2004年至2016年使用压缩植入物的病例进行了回顾性审查。比较有针组和无针组在骨-植入物界面处的机械失效率和骨生长。回归模型用于检查与机械故障相关的患者和手术因素。结果:抗旋转钉与机械故障无关(P=1.0,比值比1.17,95%置信区间:0.12-15.40)。抗旋转钉在任何时间点都与骨整合受损无关(3-6个月P=1.0,6-9个月P=0.33,9-12个月P=0.34,12-24个月P=0.40,24-48个月P=0.28,>48个月P=0.00)。没有患者或手术变量是机械故障的预测因素。结论:抗旋转钉与机械故障或骨整合受损无关。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Annals of Joint
Annals of Joint ORTHOPEDICS-
CiteScore
1.10
自引率
-25.00%
发文量
17
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信