{"title":"Thermal Testing of Titanium Implants and the Surrounding Ex-Vivo Tissue Irradiated With 9.3um CO2 Laser.","authors":"S. Froum, Roni Cantor-Balan, C. Kerbage, S. Froum","doi":"10.1097/ID.0000000000000923","DOIUrl":null,"url":null,"abstract":"PURPOSE\nTo measure the temperature rise and surface damage of titanium dental implants and the surrounding tissue in a pig jaw during 9.3-μm carbon dioxide (CO2) laser irradiation at various durations of time.\n\n\nMATERIALS AND METHODS\nThermal analysis tests were performed on 12 implants with the same surface. Twelve implants mounted alone or in pig jaws were laser-irradiated with a 9.3-μm CO2 laser using 3 different power settings. The temperature of the implant body and the proximal tissues was measured with a J-Type Thermocouple after being laser-irradiated with 3 different power setting for 30, 60 seconds, and 2 minutes. Scanning electron microscope (SEM) and digital microscope images were also taken of the all the implants before and after laser irradiation to detect the presence or absence of surface damage.\n\n\nRESULTS\nTemperature analysis showed that in all cases the implant and the proximal tissue temperatures remained around the start temperatures of the implant and tissues with fluctuations of ±3°C but never reached the upper threshold of 44°C, the temperature at which thermal injury to bone has been reported. Digital and SEM images that were taken of the implants showed an absence of surface damage at the cutting speed of 20% (0.7 W); however, cutting speeds of 30% to 100% (1.0-4.2 W) did yield surface damage.\n\n\nCONCLUSIONS\nLaser irradiation of titanium implant surfaces using a 9.3-μm carbon dioxide laser with an average power of 0.7 W showed no increase in thermal temperature of the implant body and tissue temperatures as well as no evidence of implant surface damage.","PeriodicalId":13309,"journal":{"name":"Implant Dentistry","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2019-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1097/ID.0000000000000923","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Implant Dentistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/ID.0000000000000923","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"Dentistry","Score":null,"Total":0}
引用次数: 1
Abstract
PURPOSE
To measure the temperature rise and surface damage of titanium dental implants and the surrounding tissue in a pig jaw during 9.3-μm carbon dioxide (CO2) laser irradiation at various durations of time.
MATERIALS AND METHODS
Thermal analysis tests were performed on 12 implants with the same surface. Twelve implants mounted alone or in pig jaws were laser-irradiated with a 9.3-μm CO2 laser using 3 different power settings. The temperature of the implant body and the proximal tissues was measured with a J-Type Thermocouple after being laser-irradiated with 3 different power setting for 30, 60 seconds, and 2 minutes. Scanning electron microscope (SEM) and digital microscope images were also taken of the all the implants before and after laser irradiation to detect the presence or absence of surface damage.
RESULTS
Temperature analysis showed that in all cases the implant and the proximal tissue temperatures remained around the start temperatures of the implant and tissues with fluctuations of ±3°C but never reached the upper threshold of 44°C, the temperature at which thermal injury to bone has been reported. Digital and SEM images that were taken of the implants showed an absence of surface damage at the cutting speed of 20% (0.7 W); however, cutting speeds of 30% to 100% (1.0-4.2 W) did yield surface damage.
CONCLUSIONS
Laser irradiation of titanium implant surfaces using a 9.3-μm carbon dioxide laser with an average power of 0.7 W showed no increase in thermal temperature of the implant body and tissue temperatures as well as no evidence of implant surface damage.
期刊介绍:
Cessation. Implant Dentistry, an interdisciplinary forum for general practitioners, specialists, educators, and researchers, publishes relevant clinical, educational, and research articles that document current concepts of oral implantology in sections on biomaterials, clinical reports, oral and maxillofacial surgery, oral pathology, periodontics, prosthodontics, and research. The journal includes guest editorials, letters to the editor, book reviews, abstracts of current literature, and news of sponsoring societies.