{"title":"A technique for anchor plug salvage in the setting of a cold welded taper adapter following compress implant failure","authors":"S. Campbell, Allison K. Roe, R. Avedian","doi":"10.21037/aoj.2019.08.01","DOIUrl":null,"url":null,"abstract":"The Compress implant is used to achieve compliant fixation for endoprosthetic reconstruction during limb salvage. Failures, when they occur, are typically at the bone-implant interface. We describe a case of implant failure at a modular junction near a cold-welded taper adapter, blocking access to the spindle nut. Salvage of the anchor plug was performed in this case to avoid re-cutting the remaining bone.","PeriodicalId":44459,"journal":{"name":"Annals of Joint","volume":" ","pages":""},"PeriodicalIF":0.5000,"publicationDate":"2019-08-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.21037/aoj.2019.08.01","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Joint","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.21037/aoj.2019.08.01","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
The Compress implant is used to achieve compliant fixation for endoprosthetic reconstruction during limb salvage. Failures, when they occur, are typically at the bone-implant interface. We describe a case of implant failure at a modular junction near a cold-welded taper adapter, blocking access to the spindle nut. Salvage of the anchor plug was performed in this case to avoid re-cutting the remaining bone.