Test and Development of a Specialized Pipeline for Ventilator Calibration.

Junming Liu, Wenxiu Zheng, Ziyan Wang
{"title":"Test and Development of a Specialized Pipeline for Ventilator Calibration.","authors":"Junming Liu,&nbsp;Wenxiu Zheng,&nbsp;Ziyan Wang","doi":"10.1615/CritRevBiomedEng.2023048365","DOIUrl":null,"url":null,"abstract":"<p><p>A specialized Y-type pipeline is developed to replace medical disposable Y-type pipelines for ventilator calibration and to improve accuracy, providing a reference for improvement. According to the calibration specifications of ventilators, tests are performed on specialized pipelines and more to compare their data. Taking the tidal volume 400-mL detection point as an example, the calibration errors made by medical disposable pipelines and specialized pipelines are 6.2% and -0.8%, respectively, and the errors at other detection points are roughly the same. The accuracy of the specialized Y-type pipeline has significantly improved, reducing the impact of the pipeline on tidal volume calibration from more than 6% to less than 1%. The use of specialized pipelines can significantly improve the accuracy and qualification rate of calibration, reducing the consumption of disposable Y-type pipelines and thereby significantly reducing costs and increasing efficiency.</p>","PeriodicalId":94308,"journal":{"name":"Critical reviews in biomedical engineering","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Critical reviews in biomedical engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1615/CritRevBiomedEng.2023048365","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

A specialized Y-type pipeline is developed to replace medical disposable Y-type pipelines for ventilator calibration and to improve accuracy, providing a reference for improvement. According to the calibration specifications of ventilators, tests are performed on specialized pipelines and more to compare their data. Taking the tidal volume 400-mL detection point as an example, the calibration errors made by medical disposable pipelines and specialized pipelines are 6.2% and -0.8%, respectively, and the errors at other detection points are roughly the same. The accuracy of the specialized Y-type pipeline has significantly improved, reducing the impact of the pipeline on tidal volume calibration from more than 6% to less than 1%. The use of specialized pipelines can significantly improve the accuracy and qualification rate of calibration, reducing the consumption of disposable Y-type pipelines and thereby significantly reducing costs and increasing efficiency.

呼吸机校准专用管路的测试与开发。
开发了一种专用的Y型管道,以取代医用一次性Y型管道进行呼吸机校准,并提高准确性,为改进提供参考。根据呼吸机的校准规范,对专用管道和更多管道进行测试,以比较其数据。以潮气量400mL检测点为例,医用一次性管道和专用管道的校准误差分别为6.2%和-0.8%,其他检测点的误差大致相同。专用Y型管道的精度显著提高,将管道对潮气量校准的影响从6%以上降低到1%以下。使用专用管道可以显著提高校准的准确性和合格率,减少一次性Y型管道的消耗,从而显著降低成本和提高效率。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信