Agreement among Goldmann Applanation Tonometer, Easyton Transpalpebral Tonometer, Tonopen, and Icare in Patients with Keratoconus.

Beyoglu Eye Journal Pub Date : 2023-09-13 eCollection Date: 2023-01-01 DOI:10.14744/bej.2023.56933
Merve Beyza Yildiz, Alev Ozcelik Kose, Gokhan Celik, Osman Kizilay, Serhat Imamoglu, Elvin Yildiz
{"title":"Agreement among Goldmann Applanation Tonometer, Easyton Transpalpebral Tonometer, Tonopen, and Icare in Patients with Keratoconus.","authors":"Merve Beyza Yildiz,&nbsp;Alev Ozcelik Kose,&nbsp;Gokhan Celik,&nbsp;Osman Kizilay,&nbsp;Serhat Imamoglu,&nbsp;Elvin Yildiz","doi":"10.14744/bej.2023.56933","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>The objective of the study is to evaluate the agreement between Goldmann applanation tonometer (GAT) and Easyton transpalpebral tonometer, Tonopen, and Icare in patients with Keratoconus.</p><p><strong>Methods: </strong>This cross-sectional study included 46 eyes of 26 patients with keratoconus. Intraocular pressure (IOP) is measured using easyton, icare, tonopen, and GAT. Measurements were compared and the influences of corneal topographic variables on IOP measurement were evaluated. Bland-Altman plots were used for assessing agreement between different tonometers.</p><p><strong>Results: </strong>The mean age of the participants was 24.08±6.76 (range, 18-47) years (15 males and 11 females). The highest of the mean IOP values measured with different tonometers was obtained with Easyton (12.33±1.65), followed by Tonopen (11.59±2.17), GAT (10.67±1.52), and Icare (10.04±2.33). The mean IOP value measured with Easyton was significantly higher than that measured with GAT (p<0.001). There was no significant difference between GAT and either Tonopen (p=0.154) or Icare measurements (p=0.732). There was no significant difference between Tonopen and Easyton measurements (p=0.421). Icare measurements were correlated with central corneal thickness and keratometric values. GAT measurements were correlated with only Kmax. Thirty-eight (82.6%) of the differences were within the agreement limits (assumed clinically important deviation of up to ±2 mmHg) of GAT and Tonopen, 73.9% (n=34) were within the agreement limits of GAT and Icare, and 78.3% (n=36) were within the agreement limits of GAT and Easyton.</p><p><strong>Conclusion: </strong>Compared with GAT, the gold standard method, Easyton IOP readings were higher, while both Tonopen and Icare readings were similar to GAT. All three tonometers showed acceptable agreement with the GAT, however, Tonopen showed the greatest agreement.</p>","PeriodicalId":8740,"journal":{"name":"Beyoglu Eye Journal","volume":"8 3","pages":"170-176"},"PeriodicalIF":0.0000,"publicationDate":"2023-09-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/d1/8b/BEJ-8-170.PMC10521132.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Beyoglu Eye Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.14744/bej.2023.56933","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/1/1 0:00:00","PubModel":"eCollection","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: The objective of the study is to evaluate the agreement between Goldmann applanation tonometer (GAT) and Easyton transpalpebral tonometer, Tonopen, and Icare in patients with Keratoconus.

Methods: This cross-sectional study included 46 eyes of 26 patients with keratoconus. Intraocular pressure (IOP) is measured using easyton, icare, tonopen, and GAT. Measurements were compared and the influences of corneal topographic variables on IOP measurement were evaluated. Bland-Altman plots were used for assessing agreement between different tonometers.

Results: The mean age of the participants was 24.08±6.76 (range, 18-47) years (15 males and 11 females). The highest of the mean IOP values measured with different tonometers was obtained with Easyton (12.33±1.65), followed by Tonopen (11.59±2.17), GAT (10.67±1.52), and Icare (10.04±2.33). The mean IOP value measured with Easyton was significantly higher than that measured with GAT (p<0.001). There was no significant difference between GAT and either Tonopen (p=0.154) or Icare measurements (p=0.732). There was no significant difference between Tonopen and Easyton measurements (p=0.421). Icare measurements were correlated with central corneal thickness and keratometric values. GAT measurements were correlated with only Kmax. Thirty-eight (82.6%) of the differences were within the agreement limits (assumed clinically important deviation of up to ±2 mmHg) of GAT and Tonopen, 73.9% (n=34) were within the agreement limits of GAT and Icare, and 78.3% (n=36) were within the agreement limits of GAT and Easyton.

Conclusion: Compared with GAT, the gold standard method, Easyton IOP readings were higher, while both Tonopen and Icare readings were similar to GAT. All three tonometers showed acceptable agreement with the GAT, however, Tonopen showed the greatest agreement.

Abstract Image

Abstract Image

Abstract Image

Goldmann应用眼压计、Easyton经皮眼压计、Tonopen和Icare在圆锥角膜患者中的一致性。
目的:本研究的目的是评估Goldmann压平眼压计(GAT)与Easyton经脊椎眼压计、Tonopen和Icare在圆锥角膜患者中的一致性。方法:本横断面研究包括26例圆锥角膜患者的46眼。使用easyton、icare、tonopen和GAT测量眼压。比较测量结果,并评估角膜地形变量对眼压测量的影响。Bland-Altman图用于评估不同眼压计之间的一致性。结果:参与者的平均年龄为24.08±6.76(范围18-47)岁(15名男性和11名女性)。不同眼压计测得的平均眼压值最高的是Easyton(12.33±1.65),其次是Tonopen(11.59±2.17)、GAT(10.67±1.52),Easyton测得的平均眼压值明显高于GAT测得的眼压值(P结论:与金标法GAT相比,Easyton的眼压值更高,而Tonopen和Icare的眼压值均与GAT相似。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
42
审稿时长
16 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信