Efficacy of Non-contact BallistocardiographySystem to Determine Heart Rate Variability.

IF 1.8 Q4 NEUROSCIENCES
Annals of Neurosciences Pub Date : 2022-01-01 Epub Date: 2022-02-02 DOI:10.1177/09727531211063426
Gaurav Parchani, Gulshan Kumar, Raghavendra Rao, Kaviraja Udupa, Vibhor Saran
{"title":"Efficacy of Non-contact BallistocardiographySystem to Determine Heart Rate Variability.","authors":"Gaurav Parchani,&nbsp;Gulshan Kumar,&nbsp;Raghavendra Rao,&nbsp;Kaviraja Udupa,&nbsp;Vibhor Saran","doi":"10.1177/09727531211063426","DOIUrl":null,"url":null,"abstract":"Background: Functions of the autonomic nervous system have cardinal importance in day-to-day life. Heart rate variability (HRV) has been shown to estimate the functioning of the autonomic nervous system. Imbalance in the functioning of the autonomic nervous system is seen to be associated with chronic conditions such as chronic kidney disease, cardiovascular diseases, diabetes mellitus, and so on. Purpose: To evaluate the efficacy of a non-contact ballistocardiography (BCG) system to calculate HRV parameters by comparing them to the parameters derived from a standard commercial software that uses an electrocardiogram (ECG). Methods: Current study captured an ECG signal using a three-channel ECG Holter machine, whereas the BCG signal was captured using a BCG sensor sheet consisting of vibroacoustic sensors placed under the mattress of the participants of the study. Results: The study was conducted on 24 subjects for a total of 54 overnight recordings. The proposed method covered 97.92% epochs of the standard deviation of NN intervals (SDNN) and 99.27% epochs of root mean square of successive differences (RMSSD) within 20 ms and 30 ms tolerance, respectively, whereas 98.84% of two-min intervals for low-frequency (LF) to high-frequency (HF) ratio was covered within a tolerance of 1. Kendall’s coefficient of concordance was also calculated, giving a P < .001 for all the three parameters and coefficients 0.66, 0.55, and 0.44 for SDNN, RMSSD, and LF/HF, respectively. Conclusion: The results show that HRV parameters captured using unobtrusive and non-invasive BCG sensors are comparable to HRV calculated using ECG.","PeriodicalId":7921,"journal":{"name":"Annals of Neurosciences","volume":"29 1","pages":"16-20"},"PeriodicalIF":1.8000,"publicationDate":"2022-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/9d/a1/10.1177_09727531211063426.PMC9305910.pdf","citationCount":"5","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Neurosciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/09727531211063426","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2022/2/2 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"NEUROSCIENCES","Score":null,"Total":0}
引用次数: 5

Abstract

Background: Functions of the autonomic nervous system have cardinal importance in day-to-day life. Heart rate variability (HRV) has been shown to estimate the functioning of the autonomic nervous system. Imbalance in the functioning of the autonomic nervous system is seen to be associated with chronic conditions such as chronic kidney disease, cardiovascular diseases, diabetes mellitus, and so on. Purpose: To evaluate the efficacy of a non-contact ballistocardiography (BCG) system to calculate HRV parameters by comparing them to the parameters derived from a standard commercial software that uses an electrocardiogram (ECG). Methods: Current study captured an ECG signal using a three-channel ECG Holter machine, whereas the BCG signal was captured using a BCG sensor sheet consisting of vibroacoustic sensors placed under the mattress of the participants of the study. Results: The study was conducted on 24 subjects for a total of 54 overnight recordings. The proposed method covered 97.92% epochs of the standard deviation of NN intervals (SDNN) and 99.27% epochs of root mean square of successive differences (RMSSD) within 20 ms and 30 ms tolerance, respectively, whereas 98.84% of two-min intervals for low-frequency (LF) to high-frequency (HF) ratio was covered within a tolerance of 1. Kendall’s coefficient of concordance was also calculated, giving a P < .001 for all the three parameters and coefficients 0.66, 0.55, and 0.44 for SDNN, RMSSD, and LF/HF, respectively. Conclusion: The results show that HRV parameters captured using unobtrusive and non-invasive BCG sensors are comparable to HRV calculated using ECG.

Abstract Image

Abstract Image

Abstract Image

非接触式弹道心动图系统测定心率变异性的有效性。
背景:自主神经系统的功能在日常生活中具有至关重要的作用。心率变异性(HRV)已被证明可以估计自主神经系统的功能。自主神经系统功能失衡被认为与慢性疾病如慢性肾病、心血管疾病、糖尿病等有关。目的:通过将非接触式ballicardiography (BCG)系统与使用心电图(ECG)的标准商业软件得出的参数进行比较,评估其计算HRV参数的有效性。方法:目前的研究使用三通道心电图动态心电图机捕获心电信号,而BCG信号是使用由振动声传感器组成的BCG传感器片捕获的,该传感器片放置在研究参与者的床垫下。结果:本研究共对24名受试者进行了54次夜间录音。该方法在20 ms和30 ms的容差范围内分别覆盖了97.92%的神经网络区间标准差(SDNN)和99.27%的连续差均方根(RMSSD)的历元,而在1的容差范围内覆盖了98.84%的低频(LF)高频(HF)比的2分钟间隔。还计算了肯德尔一致性系数,所有三个参数的P < 0.001, SDNN、RMSSD和LF/HF的系数分别为0.66、0.55和0.44。结论:采用无创卡介苗传感器获取的HRV参数与心电图计算的HRV参数相当。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Annals of Neurosciences
Annals of Neurosciences NEUROSCIENCES-
CiteScore
2.40
自引率
0.00%
发文量
39
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信