Comparative analysis on intaglio surface trueness, wear volume loss of antagonist, and fracture resistance of full-contour monolithic zirconia crown for single-visit dentistry under simulated mastication.

IF 2.7 3区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE
Journal of Advanced Prosthodontics Pub Date : 2022-06-01 Epub Date: 2022-06-27 DOI:10.4047/jap.2022.14.3.173
Yong-Kyu Kim, Hyung-In Yoon, Dae-Joon Kim, Jung-Suk Han
{"title":"Comparative analysis on intaglio surface trueness, wear volume loss of antagonist, and fracture resistance of full-contour monolithic zirconia crown for single-visit dentistry under simulated mastication.","authors":"Yong-Kyu Kim,&nbsp;Hyung-In Yoon,&nbsp;Dae-Joon Kim,&nbsp;Jung-Suk Han","doi":"10.4047/jap.2022.14.3.173","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This analysis aimed to evaluate the intaglio surface trueness, antagonist's wear volume loss, and fracture resistance of full-contour crowns of (Y, Nb)-stabilized fully-sintered zirconia (FSZ), 4 mol% or 5 mol% yttria-stabilized partially sintered zirconia (4YZ or 5YZ) with high-speed sintering.</p><p><strong>Materials and methods: </strong>A total of 42 zirconia crowns were separated into three groups: FSZ, 4YZ, and 5YZ (n = 14). The intaglio surface trueness of the crowns was evaluated at the inner surface, occlusal, margin, and axial areas and reported as root-mean-square, positive and negative average deviation. Half of the specimens were aged for 120,000 cycles in the chewing simulator, and the wear volume loss of antagonist was measured. Before and after chewing, the fracture load was measured for each group. The trueness values were analyzed with Welch's ANOVA, and the wear volume loss with the Kruskal-Wallis tests. Effect of the zirconia type and aging on fracture resistance of crowns was tested using two-way ANOVA.</p><p><strong>Results: </strong>The intaglio surface trueness measured at four different areas of the crown was less than 50 µm, regardless of the type of zirconia. No significant <i>P</i> in wear volume loss of antagonists were detected among the groups (<i>P</i> > .05). Both the type of zirconia and aging showed statistically significant effects on fracture resistance (<i>P</i> < .05).</p><p><strong>Conclusion: </strong>The full-contour crowns of FSZ as well as 4YZ or 5YZ with high-speed sintering were clinically acceptable, in terms of intaglio surface trueness, antagonist's wear volume loss, and fracture resistance after simulated mastication.</p>","PeriodicalId":51291,"journal":{"name":"Journal of Advanced Prosthodontics","volume":null,"pages":null},"PeriodicalIF":2.7000,"publicationDate":"2022-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/6c/48/jap-14-173.PMC9259349.pdf","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Advanced Prosthodontics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.4047/jap.2022.14.3.173","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2022/6/27 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 3

Abstract

Purpose: This analysis aimed to evaluate the intaglio surface trueness, antagonist's wear volume loss, and fracture resistance of full-contour crowns of (Y, Nb)-stabilized fully-sintered zirconia (FSZ), 4 mol% or 5 mol% yttria-stabilized partially sintered zirconia (4YZ or 5YZ) with high-speed sintering.

Materials and methods: A total of 42 zirconia crowns were separated into three groups: FSZ, 4YZ, and 5YZ (n = 14). The intaglio surface trueness of the crowns was evaluated at the inner surface, occlusal, margin, and axial areas and reported as root-mean-square, positive and negative average deviation. Half of the specimens were aged for 120,000 cycles in the chewing simulator, and the wear volume loss of antagonist was measured. Before and after chewing, the fracture load was measured for each group. The trueness values were analyzed with Welch's ANOVA, and the wear volume loss with the Kruskal-Wallis tests. Effect of the zirconia type and aging on fracture resistance of crowns was tested using two-way ANOVA.

Results: The intaglio surface trueness measured at four different areas of the crown was less than 50 µm, regardless of the type of zirconia. No significant P in wear volume loss of antagonists were detected among the groups (P > .05). Both the type of zirconia and aging showed statistically significant effects on fracture resistance (P < .05).

Conclusion: The full-contour crowns of FSZ as well as 4YZ or 5YZ with high-speed sintering were clinically acceptable, in terms of intaglio surface trueness, antagonist's wear volume loss, and fracture resistance after simulated mastication.

Abstract Image

Abstract Image

Abstract Image

模拟咀嚼下单次牙科全轮廓单片氧化锆冠凹刻表面真实度、拮抗剂磨损体积损失及抗断裂性能对比分析
目的:评价高速烧结(Y, Nb)稳定全烧结氧化锆(FSZ)、4YZ或5YZ部分氧化锆(4YZ或5YZ)全轮廓冠的凹痕表面真实度、拮抗剂磨损体积损失和抗断裂性能。材料与方法:将42个氧化锆冠分为FSZ、4YZ、5YZ三组(n = 14)。在牙冠的内表面、咬合、边缘和轴向区域评估凹刻表面的真实性,并以均方根、正负平均偏差报告。一半的试件在咀嚼模拟器中老化12万次,并测量拮抗剂的磨损体积损失。各组咀嚼前后分别测量断裂负荷。用Welch方差分析分析真度值,用Kruskal-Wallis检验分析磨损体积损失。采用双因素方差分析检验氧化锆类型和老化对冠抗断裂性能的影响。结果:无论氧化锆类型如何,在冠的四个不同区域测量的凹版表面真实度均小于50µm。各组间拮抗剂磨损体积损失P值差异不显著(P > 0.05)。氧化锆类型和老化对抗断裂能力的影响均有统计学意义(P < 0.05)。结论:FSZ全轮廓冠与4YZ、5YZ高速烧结冠在模拟咀嚼后凹痕表面真实度、拮抗剂磨损体积损失、抗断性等方面均为临床可接受的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Advanced Prosthodontics
Journal of Advanced Prosthodontics DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
4.70
自引率
3.80%
发文量
25
期刊介绍: This journal aims to convey scientific and clinical progress in the field of prosthodontics and its related areas to many dental communities concerned with esthetic and functional restorations, occlusion, implants, prostheses, and biomaterials related to prosthodontics. This journal publishes • Original research data of high scientific merit in the field of diagnosis, function, esthetics and stomatognathic physiology related to prosthodontic rehabilitation, physiology and mechanics of occlusion, mechanical and biologic aspects of prosthodontic materials including dental implants. • Review articles by experts on controversies and new developments in prosthodontics. • Case reports if they provide or document new fundamental knowledge.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信