{"title":"An automated fast neutron computed tomography instrument with on-line focusing for non-destructive evaluation.","authors":"M G Bisbee, I Oksuz, M P VanZile, N J Cherepy, L R Cao","doi":"10.1063/5.0091032","DOIUrl":null,"url":null,"abstract":"<p><p>A fast neutron tomography imaging instrument has been designed, built, and tested at The Ohio State University 500 kW Research Reactor on a fast neutron beamline with a peak neutron flux ≈5.4 × 10<sup>7</sup> n·cm<sup>-2</sup>·s<sup>-1</sup> at 1.6 MeV median neutron energy. The instrument and beamline are also configurable for thermal neutron imaging. The imaging apparatus is composed of a lens coupled, water-cooled Electron Multiplying Charge Coupled Device camera, a front-surface mirror, and a high light yield plastic Polyvinyl toluene scintillator. The instrument sits on a mobile cart. A total of 5 motion-control stages are built into the system for XYZ and rotational degrees of freedom for sample positioning; the fifth stage fine tunes the focal distance between the camera and the scintillator to achieve on-line focusing. A Python code with a user-friendly graphical user interface controls the fully automated image acquisition, not requiring user interaction, yet facilitating tracking of the image acquisition. A complete fast neutron computed tomography dataset with 360 projections requires less than 3 h, with 30 s per projection. On-line focusing is accomplished with a commercial, off-the-shelf, dielectrically actuated liquid lens. Finally, tomographic reconstructions are visualized using the Livermore Tomography Tools software package. The effective pixel size (width and height) is ≈0.1058 mm, yielding a minimum voxel size of 0.1058 × 0.1058 × 0.1058 mm<sup>3</sup>, and produces a spatial resolution of 231 μm when calculated from knife-edge measurements.</p>","PeriodicalId":519534,"journal":{"name":"The Review of scientific instruments","volume":" ","pages":"113702"},"PeriodicalIF":0.0000,"publicationDate":"2022-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Review of scientific instruments","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1063/5.0091032","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1
Abstract
A fast neutron tomography imaging instrument has been designed, built, and tested at The Ohio State University 500 kW Research Reactor on a fast neutron beamline with a peak neutron flux ≈5.4 × 107 n·cm-2·s-1 at 1.6 MeV median neutron energy. The instrument and beamline are also configurable for thermal neutron imaging. The imaging apparatus is composed of a lens coupled, water-cooled Electron Multiplying Charge Coupled Device camera, a front-surface mirror, and a high light yield plastic Polyvinyl toluene scintillator. The instrument sits on a mobile cart. A total of 5 motion-control stages are built into the system for XYZ and rotational degrees of freedom for sample positioning; the fifth stage fine tunes the focal distance between the camera and the scintillator to achieve on-line focusing. A Python code with a user-friendly graphical user interface controls the fully automated image acquisition, not requiring user interaction, yet facilitating tracking of the image acquisition. A complete fast neutron computed tomography dataset with 360 projections requires less than 3 h, with 30 s per projection. On-line focusing is accomplished with a commercial, off-the-shelf, dielectrically actuated liquid lens. Finally, tomographic reconstructions are visualized using the Livermore Tomography Tools software package. The effective pixel size (width and height) is ≈0.1058 mm, yielding a minimum voxel size of 0.1058 × 0.1058 × 0.1058 mm3, and produces a spatial resolution of 231 μm when calculated from knife-edge measurements.