{"title":"Effect of Saturation Pulse Duration and Power on pH-weighted Amide Proton Transfer Imaging: A Phantom Study.","authors":"Chiaki Tokunaga, Tatsuhiro Wada, Osamu Togao, Yasuo Yamashita, Kouji Kobayashi, Toyoyuki Kato","doi":"10.2463/mrms.mp.2021-0109","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Amide proton transfer (APT) imaging may detect changes in tissues' pH based on the chemical exchange saturation transfer (CEST) phenomenon, and thus it may be useful for identifying the penumbra in ischemic stroke patients. We investigated the effect of saturation pulse duration and power on the APT effect in phantoms with different pH values.</p><p><strong>Methods: </strong>Five samples were prepared from a 1:10 solution of egg-white albumin in phosphate-buffered saline at pH 6.53-7.65. The APT signal intensity (SI) was defined as asymmetry of the magnetization transfer ratio at 3.5 ppm. We measured the APT SIs in the egg-white albumin samples of different pH values with saturation pulse durations of 0.5, 1.0, 2.0, and 3.0 sec and saturation pulse powers of 0.5, 1.5, and 2.5 μT. The relative change in the APT SI in relation to the saturation duration and power at different pH values was defined as follows: (APT SI <sub>each saturation pulse</sub> - APT SI <sub>shortest or weakest pulse</sub>)/APT SI<sub>shortest or weakest pulse</sub>. The dependence of the APT SI on pH and the relative change in the APT SI were calculated as the slope of the linear regression.</p><p><strong>Results: </strong>The lower the pH, the larger the relative change in the APT SI, due to the change in saturation pulse duration and power. The APT SI was highly correlated with the pH at all saturation pulse durations and powers.</p><p><strong>Conclusion: </strong>The influence of saturation duration and power on the APT effect was greater at lower pH than higher pH. The combination of saturation pulse ≥ 1.0 s and power ≥ 1.5 μT was useful for the sensitive detection of changes in APT effects in the egg-white albumin samples with different pH values.</p>","PeriodicalId":18119,"journal":{"name":"Magnetic Resonance in Medical Sciences","volume":null,"pages":null},"PeriodicalIF":2.5000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/2d/59/mrms-22-487.PMC10552662.pdf","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic Resonance in Medical Sciences","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2463/mrms.mp.2021-0109","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2022/9/1 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 1
Abstract
Purpose: Amide proton transfer (APT) imaging may detect changes in tissues' pH based on the chemical exchange saturation transfer (CEST) phenomenon, and thus it may be useful for identifying the penumbra in ischemic stroke patients. We investigated the effect of saturation pulse duration and power on the APT effect in phantoms with different pH values.
Methods: Five samples were prepared from a 1:10 solution of egg-white albumin in phosphate-buffered saline at pH 6.53-7.65. The APT signal intensity (SI) was defined as asymmetry of the magnetization transfer ratio at 3.5 ppm. We measured the APT SIs in the egg-white albumin samples of different pH values with saturation pulse durations of 0.5, 1.0, 2.0, and 3.0 sec and saturation pulse powers of 0.5, 1.5, and 2.5 μT. The relative change in the APT SI in relation to the saturation duration and power at different pH values was defined as follows: (APT SI each saturation pulse - APT SI shortest or weakest pulse)/APT SIshortest or weakest pulse. The dependence of the APT SI on pH and the relative change in the APT SI were calculated as the slope of the linear regression.
Results: The lower the pH, the larger the relative change in the APT SI, due to the change in saturation pulse duration and power. The APT SI was highly correlated with the pH at all saturation pulse durations and powers.
Conclusion: The influence of saturation duration and power on the APT effect was greater at lower pH than higher pH. The combination of saturation pulse ≥ 1.0 s and power ≥ 1.5 μT was useful for the sensitive detection of changes in APT effects in the egg-white albumin samples with different pH values.
期刊介绍:
Magnetic Resonance in Medical Sciences (MRMS or Magn
Reson Med Sci) is an international journal pursuing the
publication of original articles contributing to the progress
of magnetic resonance in the field of biomedical sciences
including technical developments and clinical applications.
MRMS is an official journal of the Japanese Society for
Magnetic Resonance in Medicine (JSMRM).