{"title":"Catheter-Mounted CMOS Front-Ends for Broadband Intravascular Ultrasonic Imaging.","authors":"Steve J A Majerus, Soumyajit Mandal, Aaron Fleischman","doi":"10.1109/mwscas.2017.8052938","DOIUrl":null,"url":null,"abstract":"<p><p>Intravascular ultrasonic (IVUS) imaging catheters currently use ceramic piezoelectric transducers to form radial images of blood vessel walls. Further improvements in image quality may be enabled through Capacitive and Piezoelectric Micromachined Ultrasonic Transducers (CMUTs and PMUTs). Polymer PMUTs offer many benefits in imaging quality, however, the low acoustic sensitivity and high electrical impedance of polymer PMUTs prevents unbuffered use with 50-Ω IVUS cables. Here, we present the design and optimization of an integrated CMOS front-end specifically designed to be integrated on a 0.8-mm imaging catheter. A series-duplexer topology with active limiter was selected for simplicity of integration, and compatibility with conventional high-voltage IVUS pulser-receiver systems. Noise optimization of the front-end revealed that an impedance-matching optimum exists, which balances the system parasitic capacitances relative to the transducer complex impedance. Optimized design equations compared favorably to simulation results in predicting front-end bandwidth and transducer-referred SNR. Large-signal transient simulation showed that the front-end can withstand 100-V pulses while recovering in 240 ns for echo reception. The integrated front-end occupied 0.74 × 1.8 mm die area (including large solder-bump bond pads), with an estimated transducer-referred noise floor of 6.5 <math> <mrow> <mrow> <msub><mrow><mi>n</mi> <mi>V</mi></mrow> <mrow><mi>R</mi> <mi>M</mi> <mi>S</mi></mrow> </msub> </mrow> <mo>/</mo> <mrow><msqrt><mi>H</mi> <mi>z</mi></msqrt> </mrow> </mrow> </math> over a 100-MHz imaging bandwidth.</p>","PeriodicalId":89019,"journal":{"name":"The ... Midwest Symposium on Circuits and Systems conference proceedings : MWSCAS. Midwest Symposium on Circuits and Systems","volume":"2017 ","pages":"373-376"},"PeriodicalIF":0.0000,"publicationDate":"2017-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8211115/pdf/nihms-1690187.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The ... Midwest Symposium on Circuits and Systems conference proceedings : MWSCAS. Midwest Symposium on Circuits and Systems","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/mwscas.2017.8052938","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2017/10/2 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Intravascular ultrasonic (IVUS) imaging catheters currently use ceramic piezoelectric transducers to form radial images of blood vessel walls. Further improvements in image quality may be enabled through Capacitive and Piezoelectric Micromachined Ultrasonic Transducers (CMUTs and PMUTs). Polymer PMUTs offer many benefits in imaging quality, however, the low acoustic sensitivity and high electrical impedance of polymer PMUTs prevents unbuffered use with 50-Ω IVUS cables. Here, we present the design and optimization of an integrated CMOS front-end specifically designed to be integrated on a 0.8-mm imaging catheter. A series-duplexer topology with active limiter was selected for simplicity of integration, and compatibility with conventional high-voltage IVUS pulser-receiver systems. Noise optimization of the front-end revealed that an impedance-matching optimum exists, which balances the system parasitic capacitances relative to the transducer complex impedance. Optimized design equations compared favorably to simulation results in predicting front-end bandwidth and transducer-referred SNR. Large-signal transient simulation showed that the front-end can withstand 100-V pulses while recovering in 240 ns for echo reception. The integrated front-end occupied 0.74 × 1.8 mm die area (including large solder-bump bond pads), with an estimated transducer-referred noise floor of 6.5 over a 100-MHz imaging bandwidth.
血管内超声波(IVUS)成像导管目前使用陶瓷压电换能器来形成血管壁的径向图像。电容式和压电式微机械超声换能器(CMUT 和 PMUT)可进一步提高成像质量。聚合物 PMUT 在成像质量方面有很多优点,但由于聚合物 PMUT 的声灵敏度低、电阻抗高,因此无法与 50Ω IVUS 电缆一起使用。在此,我们介绍了集成 CMOS 前端的设计和优化,该前端专为集成在 0.8 毫米成像导管上而设计。我们选择了带有源限幅器的串联双工拓扑结构,以简化集成,并与传统的高压 IVUS 脉冲接收器系统兼容。前端噪声优化显示存在阻抗匹配最佳值,该值平衡了系统寄生电容与传感器复阻抗之间的关系。在预测前端带宽和换能器参考信噪比方面,优化后的设计方程与仿真结果相差无几。大信号瞬态仿真显示,前端可承受 100 V 脉冲,并在 240 ns 内恢复以接收回音。集成前端占用 0.74 × 1.8 mm 的芯片面积(包括大焊块接合焊盘),在 100-MHz 成像带宽上,估计换能器参考噪声本底为 6.5 n V R M S / H z。