José F Alfonso, Luis Fernández-Vega-Cueto, Belén Alfonso-Bartolozzi, Robert Montés-Micó, Luis Fernández-Vega
{"title":"Five-Year Follow-up of Correction of Myopia: Posterior Chamber Phakic Intraocular Lens With a Central Port Design.","authors":"José F Alfonso, Luis Fernández-Vega-Cueto, Belén Alfonso-Bartolozzi, Robert Montés-Micó, Luis Fernández-Vega","doi":"10.3928/1081597X-20190118-01","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To assess the long-term correction of moderate to high myopia using a posterior chamber phakic intraocular lens with a central port design.</p><p><strong>Methods: </strong>Uncorrected (UDVA) and corrected (CDVA) distance visual acuities, refraction, intraocular pressure (IOP), endothelial cell density (ECD), vault, and adverse events over a 5-year period were evaluated retrospectively.</p><p><strong>Results: </strong>A total of 147 eyes (83 patients) were evaluated. Mean postoperative UDVA and CDVA were 0.05 ± 0.11 and 0.02 ± 0.08 logMAR at 1 year and 0.13 ± 0.18 and 0.02 ± 0.09 logMAR at 5 years, respectively. More than 95% of eyes achieved CDVA of 20/25 or better at both follow-up periods. CDVA was unchanged or improved from one to three or more lines in all eyes. Preoperatively, mean spherical equivalent (SE) was -9.20 ± 3.02 diopters (D). At 1 year, the mean SE was -0.17 ± 0.26 D, with 91.53% of eyes within ±0.50 D and 100% of eyes within ±1.00 D of the target. At 5 years, the mean SE was -0.44 ± 0.47 D, with 67.4% of eyes within ±0.50 D and 90.1% of eyes within ±1.00 D of the target. Mean IOP was 12.74 ± 1.65 and 13.0 ± 2.03 mm Hg, at 1 and 5 years, respectively. No significant rise in IOP (> 20 mm Hg) occurred during the follow-up period. Mean ECD was 2,696 ± 358 and 2,645 ± 359 cells/mm<sup>2</sup> at 1 and 5 years, respectively, representing a non-significant loss of 0.43% from preoperative values (P = .304). Mean vault changed significantly from 398 ± 187 µm at 1 year to 340 ± 163 µm at 5 years (P < .001). No intraoperative or postoperative complications or adverse events occurred during the follow-up period.</p><p><strong>Conclusions: </strong>The good long-term outcomes found in this study support the use of this lens with a central port design for the correction of moderate to high myopia. [J Refract Surg. 2019;35(3):169-176.].</p>","PeriodicalId":16951,"journal":{"name":"Journal of refractive surgery","volume":" ","pages":"169-176"},"PeriodicalIF":3.0000,"publicationDate":"2019-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"51","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3928/1081597X-20190118-01","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 51
Abstract
Purpose: To assess the long-term correction of moderate to high myopia using a posterior chamber phakic intraocular lens with a central port design.
Methods: Uncorrected (UDVA) and corrected (CDVA) distance visual acuities, refraction, intraocular pressure (IOP), endothelial cell density (ECD), vault, and adverse events over a 5-year period were evaluated retrospectively.
Results: A total of 147 eyes (83 patients) were evaluated. Mean postoperative UDVA and CDVA were 0.05 ± 0.11 and 0.02 ± 0.08 logMAR at 1 year and 0.13 ± 0.18 and 0.02 ± 0.09 logMAR at 5 years, respectively. More than 95% of eyes achieved CDVA of 20/25 or better at both follow-up periods. CDVA was unchanged or improved from one to three or more lines in all eyes. Preoperatively, mean spherical equivalent (SE) was -9.20 ± 3.02 diopters (D). At 1 year, the mean SE was -0.17 ± 0.26 D, with 91.53% of eyes within ±0.50 D and 100% of eyes within ±1.00 D of the target. At 5 years, the mean SE was -0.44 ± 0.47 D, with 67.4% of eyes within ±0.50 D and 90.1% of eyes within ±1.00 D of the target. Mean IOP was 12.74 ± 1.65 and 13.0 ± 2.03 mm Hg, at 1 and 5 years, respectively. No significant rise in IOP (> 20 mm Hg) occurred during the follow-up period. Mean ECD was 2,696 ± 358 and 2,645 ± 359 cells/mm2 at 1 and 5 years, respectively, representing a non-significant loss of 0.43% from preoperative values (P = .304). Mean vault changed significantly from 398 ± 187 µm at 1 year to 340 ± 163 µm at 5 years (P < .001). No intraoperative or postoperative complications or adverse events occurred during the follow-up period.
Conclusions: The good long-term outcomes found in this study support the use of this lens with a central port design for the correction of moderate to high myopia. [J Refract Surg. 2019;35(3):169-176.].
期刊介绍:
The Journal of Refractive Surgery, the official journal of the International Society of Refractive Surgery, a partner of the American Academy of Ophthalmology, has been a monthly peer-reviewed forum for original research, review, and evaluation of refractive and lens-based surgical procedures for more than 30 years. Practical, clinically valuable articles provide readers with the most up-to-date information regarding advances in the field of refractive surgery. Begin to explore the Journal and all of its great benefits such as:
• Columns including “Translational Science,” “Surgical Techniques,” and “Biomechanics”
• Supplemental videos and materials available for many articles
• Access to current articles, as well as several years of archived content
• Articles posted online just 2 months after acceptance.