Sentry Bioconvertible Inferior Vena Cava Filter: Study of Stages of Incorporation in an Experimental Ovine Model.

IF 2.5 Q2 PERIPHERAL VASCULAR DISEASE
International Journal of Vascular Medicine Pub Date : 2018-07-19 eCollection Date: 2018-01-01 DOI:10.1155/2018/6981505
Peter A Gaines, Frank D Kolodgie, Gordon Crowley, Steven Horan, Megan MacDonagh, Emily McLucas, David Rosenthal, Ashley Strong, Michael Sweet, Deepal K Panchal
{"title":"Sentry Bioconvertible Inferior Vena Cava Filter: Study of Stages of Incorporation in an Experimental Ovine Model.","authors":"Peter A Gaines,&nbsp;Frank D Kolodgie,&nbsp;Gordon Crowley,&nbsp;Steven Horan,&nbsp;Megan MacDonagh,&nbsp;Emily McLucas,&nbsp;David Rosenthal,&nbsp;Ashley Strong,&nbsp;Michael Sweet,&nbsp;Deepal K Panchal","doi":"10.1155/2018/6981505","DOIUrl":null,"url":null,"abstract":"<p><p>The Sentry inferior vena cava (IVC) filter is designed to provide temporary protection from pulmonary embolism (PE) and then bioconvert to become incorporated in the vessel wall, leaving a patent IVC lumen. <i>Objective.</i> To evaluate the performance and stages of incorporation of the Sentry IVC filter in an ovine model. <i>Methods.</i> Twenty-four bioconvertible devices and 1 control retrievable filter were implanted in the infrarenal IVC of 25 sheep, with extensive daily monitoring and intensive imaging. Vessels and devices were analyzed at early (≤98 days, <i>n</i> = 10) and late (180 ± 30 days, <i>n</i> = 14 study devices, 1 control) termination and necropsy time-points. <i>Results.</i> Deployment success was 100% with all devices confirmed in filtering configuration, there were no filter-related complications, and bioconversion was 100% at termination with vessels widely patent. By 98 days for all early-incorporation analysis animals, the stabilizing cylindrical part of the Sentry frame was incorporated in the vessel wall, and the filter arms were retracted. By 180 days for all late-incorporation analysis animals, the filter arms as well as frames were stably incorporated. <i>Conclusions.</i> Through 180 days, there were no filter-related complications, and the study devices were all bioconverted and stably incorporated, leaving all IVCs patent.</p>","PeriodicalId":14448,"journal":{"name":"International Journal of Vascular Medicine","volume":"2018 ","pages":"6981505"},"PeriodicalIF":2.5000,"publicationDate":"2018-07-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1155/2018/6981505","citationCount":"5","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Vascular Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1155/2018/6981505","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2018/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 5

Abstract

The Sentry inferior vena cava (IVC) filter is designed to provide temporary protection from pulmonary embolism (PE) and then bioconvert to become incorporated in the vessel wall, leaving a patent IVC lumen. Objective. To evaluate the performance and stages of incorporation of the Sentry IVC filter in an ovine model. Methods. Twenty-four bioconvertible devices and 1 control retrievable filter were implanted in the infrarenal IVC of 25 sheep, with extensive daily monitoring and intensive imaging. Vessels and devices were analyzed at early (≤98 days, n = 10) and late (180 ± 30 days, n = 14 study devices, 1 control) termination and necropsy time-points. Results. Deployment success was 100% with all devices confirmed in filtering configuration, there were no filter-related complications, and bioconversion was 100% at termination with vessels widely patent. By 98 days for all early-incorporation analysis animals, the stabilizing cylindrical part of the Sentry frame was incorporated in the vessel wall, and the filter arms were retracted. By 180 days for all late-incorporation analysis animals, the filter arms as well as frames were stably incorporated. Conclusions. Through 180 days, there were no filter-related complications, and the study devices were all bioconverted and stably incorporated, leaving all IVCs patent.

Abstract Image

Abstract Image

Abstract Image

哨兵生物可转换下腔静脉过滤器:在实验羊模型中掺入阶段的研究。
Sentry下腔静脉(IVC)过滤器旨在提供暂时的保护,防止肺栓塞(PE),然后生物转化为与血管壁结合,留下一个专利的下腔静脉管腔。目标。评估Sentry IVC过滤器在羊模型中的性能和阶段。方法。将24个生物转换装置和1个对照可回收过滤器植入25只羊的肾下IVC,进行广泛的日常监测和强化成像。在早期(≤98天,n = 10)和晚期(180±30天,n = 14个研究装置,1个对照)终止和尸检时间点对血管和器械进行分析。结果。所有设备的过滤配置都得到了确认,部署成功率为100%,没有出现与过滤器相关的并发症,并且在血管广泛专利的情况下,生物转化率为100%。到第98天,所有早期掺入分析动物的Sentry框架的稳定圆柱形部分被纳入血管壁,过滤臂被收回。到180天,所有晚入组分析动物的滤光臂和镜架都稳定入组。结论。通过180天,没有过滤器相关的并发症,研究设备都是生物转化和稳定纳入,留下所有IVCs专利。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
International Journal of Vascular Medicine
International Journal of Vascular Medicine PERIPHERAL VASCULAR DISEASE-
CiteScore
3.50
自引率
0.00%
发文量
7
审稿时长
16 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信