Outcomes of LASIK for Myopia or Myopic Astigmatism Correction with the FS200 Femtosecond Laser and EX500 Excimer Laser Platform.

IF 0.4 Q4 OPHTHALMOLOGY
Open Ophthalmology Journal Pub Date : 2018-05-18 eCollection Date: 2018-01-01 DOI:10.2174/1874364101812010063
Muanploy Niparugs, Napaporn Tananuvat, Winai Chaidaroon, Chulaluck Tangmonkongvoragul, Somsanguan Ausayakhun
{"title":"Outcomes of LASIK for Myopia or Myopic Astigmatism Correction with the FS200 Femtosecond Laser and EX500 Excimer Laser Platform.","authors":"Muanploy Niparugs,&nbsp;Napaporn Tananuvat,&nbsp;Winai Chaidaroon,&nbsp;Chulaluck Tangmonkongvoragul,&nbsp;Somsanguan Ausayakhun","doi":"10.2174/1874364101812010063","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the efficacy, predictability, stability and safety of laser in situ keratomileusis (LASIK) using the FS200 femtosecond laser and EX500 excimer laser platform.</p><p><strong>Methods: </strong>The outcomes of 254 eyes of 129 consecutive patients with myopia or myopic astigmatism who underwent full correction femtosecond laser-assisted LASIK at CMU LASIK Center were assessed. Pre-operative and post-operative parameters including manifest refraction, Uncorrected Distance Visual Acuity (UDVA), Best Corrected Distance Visual Acuity (BDVA), corneal topography and tomography were analyzed. The results between low to moderate myopia and high myopia were compared up to 12 months.</p><p><strong>Results: </strong>Mean pre-operative Spherical Equivalent (SE) was -5.15±2.41 Diopters (D) (range -0.50 to -11.50 D) and -0.13±0.28 D, -0.13±0.27 D, -0.13±0.28 D and -0.14±0.30 D at 1, 3, 6, and 12 months, post-operatively. At 12 months, the propor¬tion of eyes achieving UDVA ≥ 20/20 was 90.0% and ≥20/40 was 98.8%. The proportion of eyes achieving post-operative mean SE ±0.5 D, and ±1 D was 91.3%, and 98.5%. No eyes lost more than two lines of BDVA. The low to moderate myopic group had a statistically significant better UDVA at one (<i>p</i>=0.017) and three months (<i>p</i>=0.014) but no difference at six (<i>p</i>=0.061) and 12 months (<i>p</i>=0.091). The mean post-operative SE was better in low to moderate myopic group at every follow-up visit (<i>p</i>=0.001, 0.007, <0.001 and <0.001).</p><p><strong>Conclusion: </strong>One-year clinical results of LASIK with the FS200 femtosecond laser and EX500 excimer laser showed high efficacy, predictability, stability and safety.</p>","PeriodicalId":46347,"journal":{"name":"Open Ophthalmology Journal","volume":null,"pages":null},"PeriodicalIF":0.4000,"publicationDate":"2018-05-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5960747/pdf/","citationCount":"12","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Open Ophthalmology Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2174/1874364101812010063","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2018/1/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 12

Abstract

Purpose: To evaluate the efficacy, predictability, stability and safety of laser in situ keratomileusis (LASIK) using the FS200 femtosecond laser and EX500 excimer laser platform.

Methods: The outcomes of 254 eyes of 129 consecutive patients with myopia or myopic astigmatism who underwent full correction femtosecond laser-assisted LASIK at CMU LASIK Center were assessed. Pre-operative and post-operative parameters including manifest refraction, Uncorrected Distance Visual Acuity (UDVA), Best Corrected Distance Visual Acuity (BDVA), corneal topography and tomography were analyzed. The results between low to moderate myopia and high myopia were compared up to 12 months.

Results: Mean pre-operative Spherical Equivalent (SE) was -5.15±2.41 Diopters (D) (range -0.50 to -11.50 D) and -0.13±0.28 D, -0.13±0.27 D, -0.13±0.28 D and -0.14±0.30 D at 1, 3, 6, and 12 months, post-operatively. At 12 months, the propor¬tion of eyes achieving UDVA ≥ 20/20 was 90.0% and ≥20/40 was 98.8%. The proportion of eyes achieving post-operative mean SE ±0.5 D, and ±1 D was 91.3%, and 98.5%. No eyes lost more than two lines of BDVA. The low to moderate myopic group had a statistically significant better UDVA at one (p=0.017) and three months (p=0.014) but no difference at six (p=0.061) and 12 months (p=0.091). The mean post-operative SE was better in low to moderate myopic group at every follow-up visit (p=0.001, 0.007, <0.001 and <0.001).

Conclusion: One-year clinical results of LASIK with the FS200 femtosecond laser and EX500 excimer laser showed high efficacy, predictability, stability and safety.

Abstract Image

Abstract Image

Abstract Image

FS200飞秒激光和EX500准分子激光平台矫正近视或近视散光的LASIK疗效。
目的:评价FS200飞秒激光与EX500准分子激光平台联合应用激光原位角膜磨成形术(LASIK)的疗效、可预测性、稳定性和安全性。方法:对连续129例254眼的近视或近视散光患者在CMU LASIK中心行飞秒激光辅助全矫正LASIK手术的疗效进行评估。术前、术后参数包括明显屈光、未矫正距离视力(UDVA)、最佳矫正距离视力(BDVA)、角膜地形图及层析成像。低至中度近视和高度近视的结果进行了长达12个月的比较。结果:术前平均球等效(SE)为-5.15±2.41屈光度(D)(范围为-0.50 ~ -11.50 D),术后1、3、6、12个月分别为-0.13±0.28 D、-0.13±0.27 D、-0.13±0.28 D和-0.14±0.30 D。12个月时UDVA≥20/20的眼占90.0%,≥20/40的眼占98.8%。术后平均SE±0.5 D、±1 D的眼睛比例分别为91.3%、98.5%。没有眼睛的BDVA损失超过两行。低至中度近视组在1个月(p=0.017)和3个月(p=0.014)时UDVA较好,在6个月(p=0.061)和12个月(p=0.091)时差异无统计学意义。结论:FS200飞秒激光与EX500准分子激光联合治疗LASIK术后1年临床疗效高、可预测性强、稳定性好、安全性好。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
0.50
自引率
0.00%
发文量
24
期刊介绍: The Open Ophthalmology Journal is an Open Access online journal, which publishes research articles, reviews/mini-reviews, letters and guest edited single topic issues in all important areas of experimental and clinical research in ophthalmology, including use of ophthalmological therapies, devices and surgical techniques. The Open Ophthalmology Journal, a peer-reviewed journal, is an important and reliable source of current information on developments in the field. The emphasis will be on publishing quality papers rapidly and making them freely available to researchers worldwide.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信