{"title":"Development and Validation of Waste Decontamination Cycle in a Biosafety Level 3 Laboratory.","authors":"Ernesto Cascardo, Silvina Goenaga, Sebastián Fossa, Alejandro Bottale, Silvana Levis, Laura Riera","doi":"10.1177/1535676020933714","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Steam sterilization has been used for decades to effectively kill microbial contaminants in a variety of medical and commercial settings. One of the most critical aspects of safe operations in biosafety level 3 biocontainment laboratories (BSL-3) is the effective inactivation of biological select agents in the waste generated in these environments. The Instituto Nacional de Enfermedades Virales Humanas \"Dr. Julio I. Maiztegui\" (INEVH, Pergamino, Argentina) is an institute that offers epidemiological surveillance, production of biological reagents, and production of biologicals for human use and studies of reservoirs and vectors. Some of the activities need to be done in a BSL-3 that provides biocontainment, ensuring that the materials are decontaminated before they leave the facility. The objective of this study was to design and validate a decontamination procedure for biological waste from the BSL-3 facility that guarantees steam sterilization processes.</p><p><strong>Methods: </strong>The amount and the distribution of biological waste into the autoclave and other physical parameters were defined and evaluated by calculating lethalities.</p><p><strong>Results: </strong>We evaluated autoclave basic factory programmed cycles, and it was concluded that the sterilization autoclave cycle was not efficient for decontamination of waste. A new simulated load distribution had to be defined.</p><p><strong>Discussion: </strong>The results demonstrated that autoclave factory default settings can be inadequate for sterilizing highly infectious waste, depending of types of waste, such as animal carcass and animal bed waste.</p><p><strong>Conclusion: </strong>These results of the validation process can set the standard to the design of waste management protocols to ensure effective treatment of highly infectious biological waste.</p>","PeriodicalId":520561,"journal":{"name":"Applied biosafety : journal of the American Biological Safety Association","volume":" ","pages":"225-231"},"PeriodicalIF":0.7000,"publicationDate":"2020-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1177/1535676020933714","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Applied biosafety : journal of the American Biological Safety Association","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/1535676020933714","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: Steam sterilization has been used for decades to effectively kill microbial contaminants in a variety of medical and commercial settings. One of the most critical aspects of safe operations in biosafety level 3 biocontainment laboratories (BSL-3) is the effective inactivation of biological select agents in the waste generated in these environments. The Instituto Nacional de Enfermedades Virales Humanas "Dr. Julio I. Maiztegui" (INEVH, Pergamino, Argentina) is an institute that offers epidemiological surveillance, production of biological reagents, and production of biologicals for human use and studies of reservoirs and vectors. Some of the activities need to be done in a BSL-3 that provides biocontainment, ensuring that the materials are decontaminated before they leave the facility. The objective of this study was to design and validate a decontamination procedure for biological waste from the BSL-3 facility that guarantees steam sterilization processes.
Methods: The amount and the distribution of biological waste into the autoclave and other physical parameters were defined and evaluated by calculating lethalities.
Results: We evaluated autoclave basic factory programmed cycles, and it was concluded that the sterilization autoclave cycle was not efficient for decontamination of waste. A new simulated load distribution had to be defined.
Discussion: The results demonstrated that autoclave factory default settings can be inadequate for sterilizing highly infectious waste, depending of types of waste, such as animal carcass and animal bed waste.
Conclusion: These results of the validation process can set the standard to the design of waste management protocols to ensure effective treatment of highly infectious biological waste.
蒸汽灭菌已经使用了几十年,在各种医疗和商业环境中有效地杀死微生物污染物。生物安全3级生物控制实验室(BSL-3)安全操作的最关键方面之一是使这些环境中产生的废物中的生物选择剂有效失活。国家人类病毒研究所“Julio I. Maiztegui博士”(阿根廷佩加米诺省)是一个提供流行病学监测、生产生物试剂和生产供人使用的生物制剂以及研究宿主和媒介的研究所。一些活动需要在提供生物防护的BSL-3中进行,确保材料在离开设施之前得到净化。本研究的目的是设计和验证BSL-3设施的生物废物净化程序,以保证蒸汽灭菌过程。方法:定义生物废物进入高压灭菌器的数量和分布以及其他物理参数,并通过计算致死性来评估。结果:我们评估了高压灭菌器的基本工厂程序循环,结论是灭菌高压灭菌器循环对废物的去污效率不高。必须定义一个新的模拟负载分布。讨论:结果表明,高压灭菌器工厂默认设置可能不足以灭菌高传染性废物,这取决于废物的类型,如动物尸体和动物床废物。结论:验证过程的结果可为废弃物管理方案的设计提供标准,确保高传染性生物废弃物的有效处理。