The agreement of point-of-care and standard laboratory electrolyte and glucose analysis in critically ill patients in a sub-Saharan tertiary teaching hospital.

O O Adekola, N K Irurhe, I A Meka, N O Akanmu, I D Menkiti, I C Udenze, N A Awolola
{"title":"The agreement of point-of-care and standard laboratory electrolyte and glucose analysis in critically ill patients in a sub-Saharan tertiary teaching hospital.","authors":"O O Adekola,&nbsp;N K Irurhe,&nbsp;I A Meka,&nbsp;N O Akanmu,&nbsp;I D Menkiti,&nbsp;I C Udenze,&nbsp;N A Awolola","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The critically ill patient undergoes rapid changes in the internal milieu requiring quick intervention. Point of care testing has been shown to be valuable in the early diagnosis and management of such patients.</p><p><strong>Objective: </strong>This study determined the agreement between I-STAT Abbot point of care testing with standard laboratory testing in the analysis of electrolytes and glucose concentrations in critically ill patients.</p><p><strong>Methods: </strong>The study was performed in a Sub-Saharan Tertiary Teaching Hospital in critically ill patients. Electrolyte and glucose analysis were measured with I-STAT Abbot Analyzer unit with parallel blood specimens (n = 30) tested in the laboratory on an ion-selective electrode, SFRI analyzer ISE 6000.</p><p><strong>Results: </strong>There was no significant difference in mean sodium, potassium, chloride and glucose between I-STAT POCT and standard laboratory measurements. The agreement between POCT and laboratory glucose was good p(c) = 0.967, mean difference of 0.79 and 95% limit of agreement from -3.83 to +5.107 mmol/L, p = 0.733. Bicarbonate was moderate (p) = 0.637, mean difference of 1.95 and 95% limit of agreement from -4.294 to +0.394 mmol/L, p = 0.101. There was moderate agreement for sodium (p(c)) = 0.32, mean difference of 5.8 and 95% limit of agreement from -0.378 to +11.98 mmol/L, p = 0.064. Agreement for potassium was moderate (p(c)) = 0.439, mean difference of 0.15 and limit of agreement from -0.401 to +0.701 mmol/L, p = 0.588. There was, however, a significant difference in mean chloride, and BUN values; chloride (p(c)) = 0.0796, mean difference of 13.8 and 95% limit of agreement from -7.55 to + 20.015 mmol/L. Blood urea nitrogen (p(c)) = 0.064, mean difference of 18.55 and 95% limit of agreement from -30.126 to +6.974 mmol/L.</p><p><strong>Conclusion: </strong>The mean sodium, potassium, glucose and bicarbonate were comparable with moderate to good agreement between I-STAT POCT and ISE 6000 Analyzer. Though, the mean BUN and chloride levels between the analytical methods differ significantly.</p>","PeriodicalId":19202,"journal":{"name":"Nigerian quarterly journal of hospital medicine","volume":"23 1","pages":"58-64"},"PeriodicalIF":0.0000,"publicationDate":"2013-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nigerian quarterly journal of hospital medicine","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The critically ill patient undergoes rapid changes in the internal milieu requiring quick intervention. Point of care testing has been shown to be valuable in the early diagnosis and management of such patients.

Objective: This study determined the agreement between I-STAT Abbot point of care testing with standard laboratory testing in the analysis of electrolytes and glucose concentrations in critically ill patients.

Methods: The study was performed in a Sub-Saharan Tertiary Teaching Hospital in critically ill patients. Electrolyte and glucose analysis were measured with I-STAT Abbot Analyzer unit with parallel blood specimens (n = 30) tested in the laboratory on an ion-selective electrode, SFRI analyzer ISE 6000.

Results: There was no significant difference in mean sodium, potassium, chloride and glucose between I-STAT POCT and standard laboratory measurements. The agreement between POCT and laboratory glucose was good p(c) = 0.967, mean difference of 0.79 and 95% limit of agreement from -3.83 to +5.107 mmol/L, p = 0.733. Bicarbonate was moderate (p) = 0.637, mean difference of 1.95 and 95% limit of agreement from -4.294 to +0.394 mmol/L, p = 0.101. There was moderate agreement for sodium (p(c)) = 0.32, mean difference of 5.8 and 95% limit of agreement from -0.378 to +11.98 mmol/L, p = 0.064. Agreement for potassium was moderate (p(c)) = 0.439, mean difference of 0.15 and limit of agreement from -0.401 to +0.701 mmol/L, p = 0.588. There was, however, a significant difference in mean chloride, and BUN values; chloride (p(c)) = 0.0796, mean difference of 13.8 and 95% limit of agreement from -7.55 to + 20.015 mmol/L. Blood urea nitrogen (p(c)) = 0.064, mean difference of 18.55 and 95% limit of agreement from -30.126 to +6.974 mmol/L.

Conclusion: The mean sodium, potassium, glucose and bicarbonate were comparable with moderate to good agreement between I-STAT POCT and ISE 6000 Analyzer. Though, the mean BUN and chloride levels between the analytical methods differ significantly.

撒哈拉以南某三级教学医院危重病人即时护理与标准实验室电解质和葡萄糖分析的一致性
背景:危重患者体内环境变化迅速,需要快速干预。护理点检测已被证明对这类患者的早期诊断和管理是有价值的。目的:本研究确定I-STAT Abbot护理点测试与标准实验室测试在重症患者电解质和葡萄糖浓度分析中的一致性。方法:在撒哈拉以南地区某三级教学医院对危重病人进行研究。电解质和葡萄糖分析用I-STAT Abbot分析仪测量,平行血样(n = 30)在实验室用离子选择电极SFRI分析仪ISE 6000检测。结果:I-STAT POCT与标准实验室测量值之间的平均钠、钾、氯和葡萄糖无显著差异。POCT与实验室葡萄糖的一致性较好,p(c) = 0.967,平均差为0.79,95%的一致性限为-3.83 ~ +5.107 mmol/L, p = 0.733。碳酸氢盐为中度(p) = 0.637,平均差异为1.95,95%一致性限为-4.294 ~ +0.394 mmol/L, p = 0.101。钠(p(c))有中等一致性(p(c)) = 0.32,平均差异为5.8,95%一致性限为-0.378 ~ +11.98 mmol/L, p = 0.064。钾的一致性为中等(p(c)) = 0.439,平均差值为0.15,一致性限为-0.401 ~ +0.701 mmol/L, p = 0.588。然而,平均氯化物值和BUN值有显著差异;氯化物(p(c)) = 0.0796,平均差为13.8,95%的一致性限为-7.55 ~ + 20.015 mmol/L。血尿素氮(p(c)) = 0.064,平均差值为18.55,95%一致性限为-30.126 ~ +6.974 mmol/L。结论:I-STAT POCT和ISE 6000分析仪之间的平均钠、钾、葡萄糖和碳酸氢盐具有中等到良好的一致性。然而,分析方法之间的平均BUN和氯化物水平差异很大。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信