Accuracy of flow and sensors of indirect calorimetry for neonates and infants: using mass spectrometry and a pneumotachogram.

Teruyoshi Amagai, Takeshi Mouri, Haruo Ohkawa, Isao Nishi
{"title":"Accuracy of flow and sensors of indirect calorimetry for neonates and infants: using mass spectrometry and a pneumotachogram.","authors":"Teruyoshi Amagai,&nbsp;Takeshi Mouri,&nbsp;Haruo Ohkawa,&nbsp;Isao Nishi","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>1. The accuracy of measurements of flow rate and concentrations of O2 and CO2 in expiratory gas by indirect calorimetry (IC) using paediatric and adult circuits were assessed by mass spectrometry and a pneumotachogram, which have been proven as the most reliable instruments for these purposes. 2. In the paediatric circuit, all measurements were demonstrated to be reliable: error rates for flow rate, O2 concentration and CO2 concentration were +3.13, +2.66 and -6.63%, respectively. All were within the 10%, which is acceptable as a biological error. 3. However, in the adult circuit of IC, all measurements were reliable except for measurements of CO2 concentration: error rates for flow rate, O2 concentration and CO2 concentration were +2.82, +1.64 and +11.42%, respectively. 4. A fluctuation phenomenon of expiratory gas concentration was observed only in IC. Mass spectrometry did not show this phenomenon. This phenomenon seems to be derived from the IC machine itself.</p>","PeriodicalId":75710,"journal":{"name":"Clinical and experimental pharmacology & physiology. Supplement","volume":" 29","pages":"S7-8"},"PeriodicalIF":0.0000,"publicationDate":"2002-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical and experimental pharmacology & physiology. Supplement","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

1. The accuracy of measurements of flow rate and concentrations of O2 and CO2 in expiratory gas by indirect calorimetry (IC) using paediatric and adult circuits were assessed by mass spectrometry and a pneumotachogram, which have been proven as the most reliable instruments for these purposes. 2. In the paediatric circuit, all measurements were demonstrated to be reliable: error rates for flow rate, O2 concentration and CO2 concentration were +3.13, +2.66 and -6.63%, respectively. All were within the 10%, which is acceptable as a biological error. 3. However, in the adult circuit of IC, all measurements were reliable except for measurements of CO2 concentration: error rates for flow rate, O2 concentration and CO2 concentration were +2.82, +1.64 and +11.42%, respectively. 4. A fluctuation phenomenon of expiratory gas concentration was observed only in IC. Mass spectrometry did not show this phenomenon. This phenomenon seems to be derived from the IC machine itself.

新生儿和婴儿间接量热仪的流量和传感器的准确性:使用质谱法和气测图。
1. 通过质谱法和气测图评估了使用儿童和成人电路间接量热法(IC)测量呼气气体中O2和CO2的流速和浓度的准确性,这已被证明是用于这些目的的最可靠的仪器。2. 在儿科回路中,所有测量结果均被证明是可靠的:流量、O2浓度和CO2浓度的错误率分别为+3.13、+2.66和-6.63%。所有的结果都在10%以内,这是可以接受的生物学误差。3.然而,在IC的成人电路中,除CO2浓度测量外,所有测量都是可靠的:流量、O2浓度和CO2浓度的误差率分别为+2.82、+1.64和+11.42%。4. 仅在IC中观察到呼气气体浓度的波动现象,质谱法未见此现象。这种现象似乎源于IC机本身。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书