[Comparison of film-screen combinations with contrast detail diagram and interactive image analysis. 2: Linear assessment of grey scale ranges with interactive image analysis].
{"title":"[Comparison of film-screen combinations with contrast detail diagram and interactive image analysis. 2: Linear assessment of grey scale ranges with interactive image analysis].","authors":"G Stamm, G Eichbaum, G Hagemann","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The following three screen-film combinations were compared: a) a combination of anticrossover film and UV-light emitting screens, b) a combination of blue-light emitting screens and film, and c) a conventional green fluorescing screen-film combination. Radiographs of a specially designed plexiglass phantom (0.2 x 0.2 x 0.12 m3) with bar patterns of lead and plaster and of air, respectively were obtained using the following parameters: 12 pulse generator, 0.6 mm focus size, 4.7 mm aluminum pre-filter, a grid with 40 lines/cm (12:1) and a focus-detector distance of 1.15 m. Image analysis was performed using an IBAS system and a Zeiss Kontron computer. Display conditions were the following: display distance 0.12 m, a vario film objective 35/70 (Zeiss), a video camera tube with a PbO photocathode, 625 lines (Siemens Heimann), an IBAS image matrix of 512 x 512 pixels with a resolution of 7 lines/mm, the projected matrix area was 5000 microns2. Grey scale ranges were measured on a line perpendicular to the grouped bar patterns. The difference between the maximum and minimum density value served as signal. The spatial resolution of the detector system was measured when the signal value was three times higher than the standard deviation of the means of multiple density measurements. The results showed considerable advantages of the two new screen-film combinations as compared to the conventional screen-film combination. The result was contradictory to the findings with pure visual assessment of thresholds (part I) that had found no differences. The authors concluded that (automatic) interactive image analysis algorithms serve as an objective measure and are specifically advantageous when small differences in image quality are to be evaluated.</p>","PeriodicalId":76986,"journal":{"name":"Aktuelle Radiologie","volume":"7 5","pages":"284-7"},"PeriodicalIF":0.0000,"publicationDate":"1997-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Aktuelle Radiologie","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
The following three screen-film combinations were compared: a) a combination of anticrossover film and UV-light emitting screens, b) a combination of blue-light emitting screens and film, and c) a conventional green fluorescing screen-film combination. Radiographs of a specially designed plexiglass phantom (0.2 x 0.2 x 0.12 m3) with bar patterns of lead and plaster and of air, respectively were obtained using the following parameters: 12 pulse generator, 0.6 mm focus size, 4.7 mm aluminum pre-filter, a grid with 40 lines/cm (12:1) and a focus-detector distance of 1.15 m. Image analysis was performed using an IBAS system and a Zeiss Kontron computer. Display conditions were the following: display distance 0.12 m, a vario film objective 35/70 (Zeiss), a video camera tube with a PbO photocathode, 625 lines (Siemens Heimann), an IBAS image matrix of 512 x 512 pixels with a resolution of 7 lines/mm, the projected matrix area was 5000 microns2. Grey scale ranges were measured on a line perpendicular to the grouped bar patterns. The difference between the maximum and minimum density value served as signal. The spatial resolution of the detector system was measured when the signal value was three times higher than the standard deviation of the means of multiple density measurements. The results showed considerable advantages of the two new screen-film combinations as compared to the conventional screen-film combination. The result was contradictory to the findings with pure visual assessment of thresholds (part I) that had found no differences. The authors concluded that (automatic) interactive image analysis algorithms serve as an objective measure and are specifically advantageous when small differences in image quality are to be evaluated.