[Cell Dyn 3000: technical performance and applications in hematologic pathology].

H Merle-Béral
{"title":"[Cell Dyn 3000: technical performance and applications in hematologic pathology].","authors":"H Merle-Béral","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The Abbott Cell-Dyn (CD) 3000 is an automatic analyser, designed to give cell blood counts (CBC) and white blood differential (WBCD) by 4 angels diffraction analysis. Evaluation was performed by comparison between results obtained by CD and those obtained by Technicon H1 and by microscopic examination. Technical performances as reproducibility, linearity, carryover were quite well acceptable. On normal samples, neutrophils (NE), lymphocytes (LY) and eosinophils (EO) were found to correlate with optical method with coefficient (R) values higher than 0.9. Weaker R coefficients were found for monocytes (MO) and basophils (BA), however it didn't involve clinical consequences. Significant threshold for immature granulocytes (IG) and variant lymphocytes (VL) flags were built. By using those, a false positive rate of 7% was shown. In hematological diseases, no false negative was detected because all samples were flagged. However, no acute lymphoblastic leukemia (ALL) was studied, so the detection of lymphoblast, as known as a real difficulty for analyser, was not yet evaluated. Blasts in acute myeloblastic leukemia (AML) and hairy cells were both noticed by CD. Moreover, for chronic lymphocytic leukemia (CLL), three groups were described, according to the flags released. Knowing its performances and its limits, CD is a good analyser usable in an all round or a hematological laboratory with high number of WBCD per day.</p>","PeriodicalId":19366,"journal":{"name":"Nouvelle revue francaise d'hematologie","volume":"36 Suppl 2 ","pages":"S111-3"},"PeriodicalIF":0.0000,"publicationDate":"1994-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nouvelle revue francaise d'hematologie","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The Abbott Cell-Dyn (CD) 3000 is an automatic analyser, designed to give cell blood counts (CBC) and white blood differential (WBCD) by 4 angels diffraction analysis. Evaluation was performed by comparison between results obtained by CD and those obtained by Technicon H1 and by microscopic examination. Technical performances as reproducibility, linearity, carryover were quite well acceptable. On normal samples, neutrophils (NE), lymphocytes (LY) and eosinophils (EO) were found to correlate with optical method with coefficient (R) values higher than 0.9. Weaker R coefficients were found for monocytes (MO) and basophils (BA), however it didn't involve clinical consequences. Significant threshold for immature granulocytes (IG) and variant lymphocytes (VL) flags were built. By using those, a false positive rate of 7% was shown. In hematological diseases, no false negative was detected because all samples were flagged. However, no acute lymphoblastic leukemia (ALL) was studied, so the detection of lymphoblast, as known as a real difficulty for analyser, was not yet evaluated. Blasts in acute myeloblastic leukemia (AML) and hairy cells were both noticed by CD. Moreover, for chronic lymphocytic leukemia (CLL), three groups were described, according to the flags released. Knowing its performances and its limits, CD is a good analyser usable in an all round or a hematological laboratory with high number of WBCD per day.

Cell Dyn 3000:技术性能及在血液病理学中的应用
雅培cell - dyn (CD) 3000是一款自动分析仪,设计用于通过4天使衍射分析提供血细胞计数(CBC)和白细胞差(WBCD)。通过比较CD和Technicon H1获得的结果以及显微镜检查进行评估。再现性、线性、结转等技术性能均可接受。在正常样品中,中性粒细胞(NE)、淋巴细胞(LY)和嗜酸性粒细胞(EO)与光学方法相关,系数(R)值大于0.9。单核细胞(MO)和嗜碱性细胞(BA)的R系数较弱,但不涉及临床后果。建立了未成熟粒细胞(IG)和变异淋巴细胞(VL)标志的显著阈值。通过使用这些,假阳性率为7%。在血液学疾病中,没有检测到假阴性,因为所有样本都被标记。然而,没有急性淋巴细胞白血病(ALL)的研究,因此淋巴母细胞的检测是分析仪的真正难点,尚未进行评估。急性髓母细胞白血病(AML)和毛细胞的母细胞都被CD发现。此外,慢性淋巴细胞白血病(CLL)的母细胞被描述为三组。了解了它的性能和局限性,CD是一种很好的分析仪,可用于每天有大量WBCD的全方位或血液学实验室。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信