{"title":"Calibration of infra-red gas analysers for use with carbon dioxide.","authors":"B J Legg, K J Parkinson","doi":"10.1088/0022-3735/1/10/307","DOIUrl":null,"url":null,"abstract":"An infra-red gas analyser, used for differential measurement of carbon dioxide concentrations, has been calibrated both by a pressure difference method and also by absolute concentration differences. The two calibrations differ, and a correction factor for the pressure method has been obtained.","PeriodicalId":16983,"journal":{"name":"Journal of Scientific Instruments","volume":"1 10","pages":"1003-5"},"PeriodicalIF":0.0000,"publicationDate":"1968-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1088/0022-3735/1/10/307","citationCount":"17","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Scientific Instruments","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1088/0022-3735/1/10/307","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 17
Abstract
An infra-red gas analyser, used for differential measurement of carbon dioxide concentrations, has been calibrated both by a pressure difference method and also by absolute concentration differences. The two calibrations differ, and a correction factor for the pressure method has been obtained.