White Blood Cell Counting in Severely Leukopenic Samples on a Sysmex XN Hematology Analyzer: Analytical Performance Evaluation of Whole Blood Mode and Low White Blood Cell Mode.
Jongho Yi, Hanah Kim, Gun-Hyuk Lee, Hyejung Lee, SeungWan Kim, Mina Hur
{"title":"White Blood Cell Counting in Severely Leukopenic Samples on a Sysmex XN Hematology Analyzer: Analytical Performance Evaluation of Whole Blood Mode and Low White Blood Cell Mode.","authors":"Jongho Yi, Hanah Kim, Gun-Hyuk Lee, Hyejung Lee, SeungWan Kim, Mina Hur","doi":"10.3343/alm.2026.0225","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Sysmex XN-9100 (XN) measures WBC counts in whole blood (WB) mode and automatically switches to low WBC (LW) mode for leukopenic samples (WBC <0.50×10<sup>9</sup>/L). However, the analytical performance of LW mode in such samples has not been evaluated according to standard guidelines. We evaluated the analytical performance of the two modes in samples with WBC <0.50×10<sup>9</sup>/L.</p><p><strong>Methods: </strong>Precision of WBC quantification in the two modes was evaluated using four samples according to the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) criterion (CV% 5.6). Linearity was evaluated at five concentrations. The two modes were compared in 650 leukopenic samples. Two samples showing discrepancies between the two modes were reviewed.</p><p><strong>Results: </strong>WBC counting precision met the EFLM criterion for both modes, except in a sample with WBC <0.2×10<sup>9</sup>/L. Linearity was acceptable and within the allowable deviation of ±4.8% for both modes. Very high correlation (<i>r</i>=0.99), minimal mean difference (0.00×10<sup>9</sup>/L), and almost perfect agreement (<i>κ</i>=0.90) were noted. Discrepant samples were interpreted as clinically consistent in WB and LW modes in one case each.</p><p><strong>Conclusions: </strong>This is the first study to evaluate the analytical performance of WBC counting in WB and LW modes on XN in samples with WBC <0.50×10<sup>9</sup>/L according to guidelines. The two modes demonstrated reliable analytical performance. As precision decreased at WBC <0.2×10<sup>9</sup>/L and discrepancies between the two modes were rare, careful attention is recommended.</p>","PeriodicalId":8421,"journal":{"name":"Annals of Laboratory Medicine","volume":" ","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2026-10-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Laboratory Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3343/alm.2026.0225","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Sysmex XN-9100 (XN) measures WBC counts in whole blood (WB) mode and automatically switches to low WBC (LW) mode for leukopenic samples (WBC <0.50×109/L). However, the analytical performance of LW mode in such samples has not been evaluated according to standard guidelines. We evaluated the analytical performance of the two modes in samples with WBC <0.50×109/L.
Methods: Precision of WBC quantification in the two modes was evaluated using four samples according to the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) criterion (CV% 5.6). Linearity was evaluated at five concentrations. The two modes were compared in 650 leukopenic samples. Two samples showing discrepancies between the two modes were reviewed.
Results: WBC counting precision met the EFLM criterion for both modes, except in a sample with WBC <0.2×109/L. Linearity was acceptable and within the allowable deviation of ±4.8% for both modes. Very high correlation (r=0.99), minimal mean difference (0.00×109/L), and almost perfect agreement (κ=0.90) were noted. Discrepant samples were interpreted as clinically consistent in WB and LW modes in one case each.
Conclusions: This is the first study to evaluate the analytical performance of WBC counting in WB and LW modes on XN in samples with WBC <0.50×109/L according to guidelines. The two modes demonstrated reliable analytical performance. As precision decreased at WBC <0.2×109/L and discrepancies between the two modes were rare, careful attention is recommended.