White Blood Cell Counting in Severely Leukopenic Samples on a Sysmex XN Hematology Analyzer: Analytical Performance Evaluation of Whole Blood Mode and Low White Blood Cell Mode.

IF 4.4 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY
Jongho Yi, Hanah Kim, Gun-Hyuk Lee, Hyejung Lee, SeungWan Kim, Mina Hur
{"title":"White Blood Cell Counting in Severely Leukopenic Samples on a Sysmex XN Hematology Analyzer: Analytical Performance Evaluation of Whole Blood Mode and Low White Blood Cell Mode.","authors":"Jongho Yi, Hanah Kim, Gun-Hyuk Lee, Hyejung Lee, SeungWan Kim, Mina Hur","doi":"10.3343/alm.2026.0225","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Sysmex XN-9100 (XN) measures WBC counts in whole blood (WB) mode and automatically switches to low WBC (LW) mode for leukopenic samples (WBC <0.50×10<sup>9</sup>/L). However, the analytical performance of LW mode in such samples has not been evaluated according to standard guidelines. We evaluated the analytical performance of the two modes in samples with WBC <0.50×10<sup>9</sup>/L.</p><p><strong>Methods: </strong>Precision of WBC quantification in the two modes was evaluated using four samples according to the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) criterion (CV% 5.6). Linearity was evaluated at five concentrations. The two modes were compared in 650 leukopenic samples. Two samples showing discrepancies between the two modes were reviewed.</p><p><strong>Results: </strong>WBC counting precision met the EFLM criterion for both modes, except in a sample with WBC <0.2×10<sup>9</sup>/L. Linearity was acceptable and within the allowable deviation of ±4.8% for both modes. Very high correlation (<i>r</i>=0.99), minimal mean difference (0.00×10<sup>9</sup>/L), and almost perfect agreement (<i>κ</i>=0.90) were noted. Discrepant samples were interpreted as clinically consistent in WB and LW modes in one case each.</p><p><strong>Conclusions: </strong>This is the first study to evaluate the analytical performance of WBC counting in WB and LW modes on XN in samples with WBC <0.50×10<sup>9</sup>/L according to guidelines. The two modes demonstrated reliable analytical performance. As precision decreased at WBC <0.2×10<sup>9</sup>/L and discrepancies between the two modes were rare, careful attention is recommended.</p>","PeriodicalId":8421,"journal":{"name":"Annals of Laboratory Medicine","volume":" ","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2026-10-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Laboratory Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3343/alm.2026.0225","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Sysmex XN-9100 (XN) measures WBC counts in whole blood (WB) mode and automatically switches to low WBC (LW) mode for leukopenic samples (WBC <0.50×109/L). However, the analytical performance of LW mode in such samples has not been evaluated according to standard guidelines. We evaluated the analytical performance of the two modes in samples with WBC <0.50×109/L.

Methods: Precision of WBC quantification in the two modes was evaluated using four samples according to the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) criterion (CV% 5.6). Linearity was evaluated at five concentrations. The two modes were compared in 650 leukopenic samples. Two samples showing discrepancies between the two modes were reviewed.

Results: WBC counting precision met the EFLM criterion for both modes, except in a sample with WBC <0.2×109/L. Linearity was acceptable and within the allowable deviation of ±4.8% for both modes. Very high correlation (r=0.99), minimal mean difference (0.00×109/L), and almost perfect agreement (κ=0.90) were noted. Discrepant samples were interpreted as clinically consistent in WB and LW modes in one case each.

Conclusions: This is the first study to evaluate the analytical performance of WBC counting in WB and LW modes on XN in samples with WBC <0.50×109/L according to guidelines. The two modes demonstrated reliable analytical performance. As precision decreased at WBC <0.2×109/L and discrepancies between the two modes were rare, careful attention is recommended.

Sysmex XN血液学分析仪对严重白细胞减少样品的白细胞计数:全血模式和低白细胞模式的分析性能评价。
背景:Sysmex XN-9100 (XN)在全血(WB)模式下测量白细胞计数,并在白细胞减少样本(WBC 9/L)时自动切换到低白细胞(LW)模式。然而,LW模式在此类样品中的分析性能尚未根据标准指南进行评估。我们在WBC为9/L的样品中评估了这两种模式的分析性能。方法:根据欧洲临床化学与检验医学联合会(EFLM)标准(CV% 5.6),对4个样品进行两种方法的白细胞定量精密度评价。在五种浓度下进行线性评价。在650例白细胞减少样本中比较了两种模式。两个样品显示两种模式之间的差异进行了审查。结果:除WBC为9/L的样品外,两种模式的计数精度均满足EFLM标准。两种模式的线性度均可接受,且允许偏差在±4.8%以内。非常高的相关性(r=0.99),最小的平均差异(0.00×109/L),几乎完全一致(κ=0.90)。差异样本在WB和LW模式中各有一例被解释为临床一致。结论:本研究首次根据指南评价WBC为9/L的样品中WB和LW模式的WBC计数对XN的分析性能。两种模式均具有可靠的分析性能。由于WBC 9/L的精度降低,两种模式之间的差异很少,建议仔细注意。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Annals of Laboratory Medicine
Annals of Laboratory Medicine MEDICAL LABORATORY TECHNOLOGY-
CiteScore
8.30
自引率
12.20%
发文量
100
审稿时长
6-12 weeks
期刊介绍: Annals of Laboratory Medicine is the official journal of Korean Society for Laboratory Medicine. The journal title has been recently changed from the Korean Journal of Laboratory Medicine (ISSN, 1598-6535) from the January issue of 2012. The JCR 2017 Impact factor of Ann Lab Med was 1.916.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书